Accuracy of commercial blood gas analyzers for monitoring ionized calcium at low concentrations

被引:9
|
作者
D'Orazio, Paul [1 ]
Visnick, Helen [1 ]
Balasubramanian, Shankar [1 ]
机构
[1] Instrumentat Lab Co, 180 Hartwell Rd, Bedford, MA USA
关键词
Ionized calcium; Blood gas analyzer; Citrate anticoagulation; Sodium interference; Carryover; REGIONAL CITRATE ANTICOAGULATION; MEMBRANE ELECTRODES; SELECTIVE ELECTRODE; FORMATION-CONSTANTS; SERUM; CRRT; MAGNESIUM; DESIGN;
D O I
10.1016/j.cca.2016.07.010
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Background: Variable ionized calcium measurements in post filter blood samples during continuous renal replacement therapy (renal dialysis) using regional citrate anticoagulation (RCA) have been reported using commercial blood gas analyzers, resulting in analyzer-dependent differences in decisions regarding adjustment of citrate dose. Methods: We evaluated accuracy for measurement of iCa at low concentrations by 4 commercial blood gas analyzers using primary reference solutions formulated down to 0.15 mmol/l iCa. Results: Of the 4 analyzers tested, GEM Premier 4000 demonstrates acceptable accuracy for iCa measurement with a median deviation of -6.7% (-0.01 mmol/l) at 0.15 mmol/l, while other analyzers tested show increasingly positive biases from +40% (+0.06 mmol/l) to +60% (+0.09 mmol/l) relative to target. These relative differences are consistent with discordant results reported for measurement of iCa in blood during RCA. Interference from sodium with measured iCa and carryover from system rinse solution to sample are likely contributors to variability. Conclusions: We conclude the GEM Premier 4000 shows acceptable accuracy for measuring iCa at low concentrations required to control citrate dose during RCA. The method presented here may be used to test accuracy of any blood gas analyzer prior to use in clinical applications requiring measurement of iCa at low concentrations. (C) 2016 Elsevier B.V. All rights reserved.
引用
收藏
页码:34 / 40
页数:7
相关论文
共 50 条
  • [21] Heparin Lowers Ionized Calcium (iCa) Measurements By Blood Gas Analyzer† 206
    S P Jonna
    G J Hauser
    J H Heretzog
    H J Dalton
    [J]. Pediatric Research, 1998, 43 (Suppl 4) : 38 - 38
  • [22] The effect of attenuating dietary phosphate restriction on blood ionized calcium concentrations in cats with chronic kidney disease and ionized hypercalcemia
    Rebecca, F. Geddes
    van den Broek, D. Hendrik N.
    Chang, Yu-Mei
    Biourge, Vincent
    Elliott, Jonathan
    Jepson, Rosanne E.
    [J]. JOURNAL OF VETERINARY INTERNAL MEDICINE, 2021, 35 (02) : 997 - 1007
  • [23] Neonatal hemoglobin affects the accuracy of whole blood bilirubin measurement on GEM Premier 4000 blood gas analyzers
    Huang, Yun
    Dean, Robert
    Dubbelman, Yvonne
    Vincent, Anne
    Khurshid, Faiza
    [J]. PRACTICAL LABORATORY MEDICINE, 2021, 25
  • [24] Evaluation of two benchtop blood gas analyzers for measurement of electrolyte concentrations in venous blood samples from dogs
    Schmitz, Katrina L.
    Jeffery, Unity
    Heinz, Justin A.
    Rutter, Christine R.
    [J]. AMERICAN JOURNAL OF VETERINARY RESEARCH, 2021, 82 (02) : 105 - 109
  • [25] INTENSIVE MONITORING OF POST FILTER IONIZED CALCIUM CONCENTRATIONS DURING CVVHD WITH REGIONAL CITRATE ANTICOAGULATION
    Khadzhynov, Dmytro
    Halleck, Fabian
    Lehner, Lukas
    Staeck, Oliver
    Slowinski, Torsten
    [J]. NEPHROLOGY DIALYSIS TRANSPLANTATION, 2018, 33
  • [26] Discrepant post-filter ionized calcium concentrations by 2 common gas analyzers in continuous renal replacement therapy using regional citrate anticoagulation: another piece of the puzzle
    Larcher, Romaric
    Kuster, Nils
    Bargnoux, Anne-Sophie
    Klouche, Kada
    Pieroni, Laurence
    Cristol, Jean-Paul
    [J]. KIDNEY INTERNATIONAL, 2021, 99 (01) : 268 - 269
  • [27] Predictive accuracy of serum total calcium for both critically high and critically low ionized calcium in critical illness
    Hu, Zhi-De
    Huang, Yuan-Lan
    Wang, Mei-Ying
    Hu, Ge-Ji-Le
    Han, Yan-Qiu
    [J]. JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2018, 32 (09)
  • [28] Low levels of total and ionized calcium in blood of COVID-19 patients
    Cappellini, Fabrizio
    Brivio, Rinaldo
    Casati, Marco
    Cavallero, Annalisa
    Contro, Ernesto
    Brambilla, Paolo
    [J]. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2020, 58 (09) : E171 - E173
  • [29] MONITORING BLOOD IONIZED CALCIUM IN CHILDREN UNDERGOING OPEN-HEART (BYPASS) SURGERY
    BELTON, NR
    ADAMS, A
    [J]. JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1981, 19 (08): : 611 - 611
  • [30] Evaluation of an electrolyte analyzer for measurement of ionized calcium and magnesium concentrations in blood, plasma, and serum of dogs
    Unterer, S
    Lutz, H
    Gerber, B
    Glaus, TM
    Hässig, M
    Reusch, CE
    [J]. AMERICAN JOURNAL OF VETERINARY RESEARCH, 2004, 65 (02) : 183 - 187