Ion-recombination correction for different ionization chambers in high dose rate flattening-filter-free photon beams

被引:32
|
作者
Lang, Stephanie [1 ]
Hrbacek, Jan [1 ]
Leong, Aidan [1 ]
Kloeck, Stephan [1 ]
机构
[1] Univ Zurich Hosp, Dept Radiat Oncol, CH-8901 Zurich, Switzerland
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2012年 / 57卷 / 09期
关键词
GENERAL COLLECTION EFFICIENCY; LIQUID ISOOCTANE; SATURATION CORRECTION; PLANE-PARALLEL; PULSED PHOTON; ELECTRON; TETRAMETHYLSILANE; YIELD;
D O I
10.1088/0031-9155/57/9/2819
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recently, there has been an increased interest in flattening-filter-free (FFF) linear accelerators. Removal of the filter results in available dose rates up to 24 Gy min(-1) (for nominal energy 10 MV in depth of maximum dose, a source-surface distance of 100 cm and a field size of 10 x 10 cm(2)). To guarantee accurate relative and reference dosimetry for the FFF beams, we investigated the charge collection efficiency of multiple air-vented and one liquid ionization chamber for dose rates up to 31.9 Gy min(-1). For flattened beams, the ion-collection efficiency of all air-vented ionization chambers (except for the PinPoint chamber) was above 0.995. By removing the flattening filter, we found a reduction in collection efficiency of approximately 0.5-0.9% for a 10 MV beam. For FFF beams, the Markus chamber showed the largest collection efficiency of 0.994. The observed collection efficiencies were dependent on dose per pulse, but independent of the pulse repetition frequency. Using the liquid ionization chamber, the ion-collection efficiency for flattened beams was above 0.990 for all dose rates. However, this chamber showed a low collection efficiency of 0.940 for the FFF 10 MV beam at a dose rate of 31.9 Gy min(-1). All investigated air-vented ionization chambers can be reliably used for relative dosimetry of FFF beams. The order of correction for reference dosimetry is given in the manuscript. Due to their increased saturation in high dose rate FFF beams, liquid ionization chambers appear to be unsuitable for dosimetry within these contexts.
引用
收藏
页数:9
相关论文
共 50 条
  • [41] Determination of ion recombination correction factors for a liquid ionization chamber in megavoltage photon beams
    Choi, Sang Hyoun
    Kim, Kum-Bae
    Ji, Young Hoon
    Kim, Chan Hyeong
    Kim, Seonghoon
    Huh, Hyun Do
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2015, 66 (09) : 1439 - 1447
  • [42] Determination of ion recombination correction factors for a liquid ionization chamber in megavoltage photon beams
    Sang Hyoun Choi
    Kum-Bae Kim
    Young Hoon Ji
    Chan Hyeong Kim
    Seonghoon Kim
    Hyun Do Huh
    Journal of the Korean Physical Society, 2015, 66 : 1439 - 1447
  • [43] Ion Recombination Correction Factors (Pion) for Varian TrueBeam High Dose Rate Therapy Beams
    Kry, S.
    Popple, R.
    Molineu, A.
    Followill, D.
    MEDICAL PHYSICS, 2012, 39 (06) : 3790 - 3790
  • [44] The perturbation correction factors for cylindrical ionization chambers in high-energy photon beams
    Yoshiyama F.
    Araki F.
    Ono T.
    Radiological Physics and Technology, 2010, 3 (2) : 159 - 164
  • [45] Fidelity of dose delivery at high dose rate of volumetric modulated arc therapy in a truebeam linac with flattening filter free beams
    Kalantzis, Georgios
    Qian, Jianguo
    Han, Bin
    Luxton, Gary
    JOURNAL OF MEDICAL PHYSICS, 2012, 37 (04) : 193 - 199
  • [46] Investigations of Dose Delivery Accuracy of High Dose Rate Flattening Filter Free (FFF) Beams in RapidArc™ and IMRT Deliveries on TrueBeam™
    Luxton, G.
    Mok, E.
    Qian, J.
    Fahimian, B.
    Kielar, K.
    Hsu, A.
    Xing, L.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [47] Correction for volume recombination in liquid ionization chambers at high dose-per-pulse
    Gotz, Malte
    Ka, Leonhard
    Tolli, Heikki
    Pawelke, Joerg
    MEDICAL PHYSICS, 2019, 46 (08) : 3692 - 3699
  • [48] Influence of the Applied Voltage On the Ion Recombination Correction Factor Using the Two-Voltage Technique in Flattening Filter-Free Beams
    Yada, R.
    Hayashi, N.
    Akasaka, H.
    Mukumoto, N.
    Sasaki, R.
    MEDICAL PHYSICS, 2017, 44 (06) : 3220 - 3220
  • [49] Dosimetric impact of failing to apply correction factors to ion recombination in percentage depth dose measurements and the volume-averaging effect in flattening filter-free beams
    Martin-Martin, Guadalupe
    Walter, Stefan
    Guibelalde, Eduardo
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 77 : 176 - 180
  • [50] Dose calculation accuracy in the build-up region of flattening filter-free photon beams
    De Puysseleyr, A.
    Lechner, W.
    Georg, D.
    De Wagter, C.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S71 - S72