Ion-recombination correction for different ionization chambers in high dose rate flattening-filter-free photon beams

被引:32
|
作者
Lang, Stephanie [1 ]
Hrbacek, Jan [1 ]
Leong, Aidan [1 ]
Kloeck, Stephan [1 ]
机构
[1] Univ Zurich Hosp, Dept Radiat Oncol, CH-8901 Zurich, Switzerland
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2012年 / 57卷 / 09期
关键词
GENERAL COLLECTION EFFICIENCY; LIQUID ISOOCTANE; SATURATION CORRECTION; PLANE-PARALLEL; PULSED PHOTON; ELECTRON; TETRAMETHYLSILANE; YIELD;
D O I
10.1088/0031-9155/57/9/2819
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recently, there has been an increased interest in flattening-filter-free (FFF) linear accelerators. Removal of the filter results in available dose rates up to 24 Gy min(-1) (for nominal energy 10 MV in depth of maximum dose, a source-surface distance of 100 cm and a field size of 10 x 10 cm(2)). To guarantee accurate relative and reference dosimetry for the FFF beams, we investigated the charge collection efficiency of multiple air-vented and one liquid ionization chamber for dose rates up to 31.9 Gy min(-1). For flattened beams, the ion-collection efficiency of all air-vented ionization chambers (except for the PinPoint chamber) was above 0.995. By removing the flattening filter, we found a reduction in collection efficiency of approximately 0.5-0.9% for a 10 MV beam. For FFF beams, the Markus chamber showed the largest collection efficiency of 0.994. The observed collection efficiencies were dependent on dose per pulse, but independent of the pulse repetition frequency. Using the liquid ionization chamber, the ion-collection efficiency for flattened beams was above 0.990 for all dose rates. However, this chamber showed a low collection efficiency of 0.940 for the FFF 10 MV beam at a dose rate of 31.9 Gy min(-1). All investigated air-vented ionization chambers can be reliably used for relative dosimetry of FFF beams. The order of correction for reference dosimetry is given in the manuscript. Due to their increased saturation in high dose rate FFF beams, liquid ionization chambers appear to be unsuitable for dosimetry within these contexts.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Assessment of ion recombination correction and polarity effects for specific ionization chambers in flattening-filter-free photon beams
    Martin-Martin, Guadalupe
    Borja Aguilar, Pedro
    Barbes, Benigno
    Guibelalde, Eduardo
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 67 : 176 - 184
  • [2] Ion recombination and polarity corrections for small-volume ionization chambers in high-dose-rate, flattening-filter-free pulsed photon beams
    Hyun, Megan A.
    Miller, Jessica R.
    Micka, John A.
    DeWerd, Larry A.
    [J]. MEDICAL PHYSICS, 2017, 44 (02) : 618 - 627
  • [3] Use and Choice of Ionization Chambers for the Commissioning of Flattened and Flattening-Filter-Free Photon Beams: Determination of Recombination Correction Factor (ks)
    Stucchi, C.
    Bresolin, A.
    Mongioj, V.
    Carrara, M.
    Pignoli, E.
    Bonfantini, F.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 371 - 371
  • [4] Study of the ionization chamber response to flattening-filter-free photon beams
    Bresolin, A.
    Bonfantini, F.
    Stucchi, C. G.
    Mongioj, V.
    Carrara, M.
    Pignoli, E.
    [J]. RADIATION MEASUREMENTS, 2017, 97 : 47 - 53
  • [5] Ion recombination corrections of ionization chambers in flattening filter-free photon radiation
    Wang, Yuenan
    Easterling, Stephen B.
    Ting, Joseph Y.
    [J]. JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (05): : 262 - 268
  • [6] Polarity correction factor for flattening filter free photon beams in several cylindrical ionization chambers
    Ogata T.
    Uehara K.
    Nakayama M.
    Tsudou S.
    Masutani T.
    Okayama T.
    [J]. Radiological Physics and Technology, 2016, 9 (2) : 187 - 192
  • [7] The impact of flattening filter free (FFF) photon beams to ion recombination correction factor
    Hentihu, F. K.
    Ryangga, D.
    Pawiro, S. A.
    [J]. 18TH ASIA-OCEANIA CONGRESS OF MEDICAL PHYSICS (AOCMP) & 16TH SOUTH-EAST ASIA CONGRESS OF MEDICAL PHYSICS (SEACOMP), 2019, 1248
  • [8] Ion-recombination correction factor k(sat) for spherical ion chambers irradiated by continuous photon beams
    Piermattei, A
    Azario, L
    Arcovito, G
    Toni, MP
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (06): : 1025 - 1035
  • [9] Comparison of Head Scatters in Flattening-Filter-Free and Flattened Photon Beams
    Chen, Y.
    Shih, R.
    Afghan, M.
    Gopal, A.
    Chen, H.
    [J]. MEDICAL PHYSICS, 2013, 40 (06)
  • [10] Ion recombination correction in ionization chambers exposed to high dose per pulse electron beams
    Marini, P.
    Poggiu, A.
    D'Avenia, P.
    Fattori, S.
    Rossi, G.
    Felici, G.
    Proietti, A.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S139 - S139