In vitro validation of MR measurements of arterial pulse-wave velocity in the presence of reflected waves

被引:24
|
作者
Stevanov, M [1 ]
Baruthio, J [1 ]
Gounot, D [1 ]
Grucker, D [1 ]
机构
[1] Univ Strasbourg 1, Inst Phys Biol, Fac Med, UPRES A 7004,CNRS, F-67085 Strasbourg, France
关键词
compliance; flow propagation; projective method; velocity imaging; reflections;
D O I
10.1002/jmri.1161
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A magnetic resonance imaging projective velocity encoding sequence was used to determine the pulse-wave velocity in an artery model. To this end, a well-defined flow phantom simulating flow propagation in large arteries was used. In order to validate the measurement method in the presence of large reflected waves, these were deliberately created in the phantom. The projective sequence was applied to two measurement sites and the wave velocity was determined from the spatial and temporal separations of the foot of the velocity waveform. A theoretical model describing reflection and attenuation phenomena was compared with experimental velocity waveforms. The model showed that reflections and attenuation can explain the important changes in velocity waveforms. The model also confirmed that in the presence of reflecting waves, the foot of the waveform can be used as a characteristic point for measurements through changes in the waveform. (C) 2001 Wiley-Liss, Inc.
引用
收藏
页码:120 / 127
页数:8
相关论文
共 50 条
  • [41] Arterial path length measurements required for the pulse wave velocity
    Sugawara, Jun
    Tanaka, Hirofumi
    [J]. JOURNAL OF HYPERTENSION, 2009, 27 (05) : 1102 - 1102
  • [42] In Vivo Measurement of Local Aortic Pulse-Wave Velocity in Mice With MR Microscopy at 17.6 Tesla
    Herold, Volker
    Parczyk, Marco
    Moerchel, Philipp
    Ziener, Christian H.
    Klug, Gert
    Bauer, Wolfgang R.
    Rommel, Eberhard
    Jakob, Peter M.
    [J]. MAGNETIC RESONANCE IN MEDICINE, 2009, 61 (06) : 1293 - 1299
  • [43] Inflammation does not influence arterial stiffness and pulse-wave velocity in patients with coronary artery disease
    Blann, A. D.
    Kuzniatsova, N.
    Lip, G. Y. H.
    [J]. JOURNAL OF HUMAN HYPERTENSION, 2013, 27 (10) : 629 - 634
  • [44] Comparison of techniques for measuring pulse-wave velocity in the rat
    Mitchell, GF
    Pfeffer, MA
    Finn, PV
    Pfeffer, JM
    [J]. JOURNAL OF APPLIED PHYSIOLOGY, 1997, 82 (01) : 203 - 210
  • [45] PULSE-WAVE VELOCITY AS AN INDEX OF AGING IN THE CARDIOVASCULAR SYSTEM
    PAINE, RM
    [J]. JOURNALS OF GERONTOLOGY, 1948, 3 (04): : 303 - 305
  • [46] CALCULATION OF PULSE-WAVE VELOCITY USING CROSS-CORRELATION - EFFECTS OF REFLEXES IN THE ARTERIAL TREE
    BENTHIN, M
    DAHL, P
    RUZICKA, R
    LINDSTROM, K
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 1991, 17 (05): : 461 - 469
  • [47] A retrospective method for pulse-wave velocity measurement in the mouse
    Williams, R
    Needles, A
    Cherin, E
    Foster, FS
    Zhou, YQ
    Henkelman, M
    [J]. 2005 IEEE Ultrasonics Symposium, Vols 1-4, 2005, : 385 - 388
  • [48] Reproducibility of regional pulse-wave velocity in uremic subjects
    Di Iorio, Biagio R.
    Cucciniello, Emanuele
    Alinei, Pietro
    Torraca, Serena
    [J]. HEMODIALYSIS INTERNATIONAL, 2010, 14 (04) : 441 - 446
  • [49] THE COUNTER CHRONOMETER METHOD FOR RECORDING PULSE-WAVE VELOCITY
    HALE, JF
    MCDONALD, DA
    TAYLOR, MG
    WOMERSLEY, JR
    [J]. JOURNAL OF PHYSIOLOGY-LONDON, 1955, 129 (03): : P27 - P28
  • [50] REPRODUCIBILITY AT 3 SITES OF PULSE-WAVE VELOCITY AT REST
    SICHE, JP
    DURANDET, P
    DEGAUDEMARIS, R
    MANSOUR, P
    MALLION, JM
    [J]. ARCHIVES DES MALADIES DU COEUR ET DES VAISSEAUX, 1990, 83 (08): : 1195 - 1199