Low-Dose Three-Dimensional Rotational Angiography for Evaluating Intracranial Aneurysms: Analysis of Image Quality and Radiation Dose

被引:3
|
作者
Ki, Hee Jong [1 ]
Kim, Bum-Soo [2 ]
Kim, Jun-Ki [1 ]
Choi, Jai Ho [1 ]
Shin, Yong Sam [1 ]
Choi, Yangsean [2 ]
Shin, Na-Young [2 ]
Jang, Jinhee [2 ]
Ahn, Kook-Jin [2 ]
机构
[1] Catholic Univ Korea, Coll Med, Dept Neurosurg, Seoul St Marys Hosp, Seoul, South Korea
[2] Catholic Univ Korea, Coll Med, Dept Radiol, Seoul St Marys Hosp, 222 Banpo Daero, Seoul 06591, South Korea
关键词
Low-dose 3D rotational angiography; Image quality; Radiation dose; Intracranial aneurysm; DIGITAL-SUBTRACTION-ANGIOGRAPHY; ENDOVASCULAR TREATMENT; CEREBRAL-ANGIOGRAPHY; FOLLOW-UP; REDUCTION;
D O I
10.3348/kjr.2021.0162
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: This study aimed to evaluate the image quality and dose reduction of low-dose three-dimensional (3D) rotational angiography (RA) for evaluating intracranial aneurysms. Materials and Methods: We retrospectively evaluated the clinical data and 3D RA datasets obtained from 146 prospectively registered patients (male:female, 46:100; median age, 58 years; range, 19-81 years). The subjective image quality of 79 examinations obtained from a conventional method and 67 examinations obtained from a low-dose (5-seconds and 0.10 mu Gy/frame) method was assessed by two neurointerventionists using a 3-point scale for four evaluation criteria. The total image quality score was then obtained as the average of the four scores. The image quality scores were compared between the two methods using a noninferiority statistical testing, with a margin of -0.2 (i.e., score of low-dose group - score of conventional group). For the evaluation of dose reduction, dose-area product (DAP) and air kerma (AK) were analyzed and compared between the two groups. Results: The mean total image quality score +/- standard deviation of the 3D RA was 2.97 +/- 0.17 by reader 1 and 2.95 +/- 0.20 by reader 2 for conventional group and 2.92 +/- 0.30 and 2.95 +/- 0.22, respectively, for low-dose group. The image quality of the 3D RA in the low-dose group was not inferior to that of the conventional group according to the total image quality score as well as individual scores for the four criteria in both readers. The mean DAP and AK per rotation were 5.87 Gy-cm2 and 0.56 Gy, respectively, in the conventional group, and 1.32 Gy-cm2 (p < 0.001) and 0.17 Gy (p < 0.001), respectively, in the low-dose group. Conclusion: Low-dose 3D RA was not inferior in image quality and reduced the radiation dose by 70%-77% compared to the conventional 3D RA in evaluating intracranial aneurysms.
引用
收藏
页码:256 / 263
页数:8
相关论文
共 50 条
  • [1] Three-dimensional cardiac rotational angiography: effective radiation dose and image quality implications
    Wielandts, Jean-Yves
    De Buck, Stijn
    Ector, Joris
    LaGerche, Andre
    Willems, Rik
    Bosmans, Hilde
    Heidbuchel, Hein
    [J]. EUROPACE, 2010, 12 (02): : 194 - 201
  • [2] Cardiac three-dimensional rotational angiography can be performed with low radiation dose while preserving image quality
    De Buck, Stijn
    Alzand, Becker S. N.
    Wielandts, Jean-Yves
    Garweg, Christophe
    Phlips, Thomas
    Ector, Joris
    Nuyens, Dieter
    Heidbuchel, Hein
    [J]. EUROPACE, 2013, 15 (12): : 1718 - 1724
  • [3] Three-dimensional rotational angiography of the left atrium: effects of decreasing frame-rate on radiation dose and image quality
    Alzand, B. S. N.
    De Buck, S.
    Wielandts, J. Y.
    Garweg, G.
    Ector, J.
    Phlips, T.
    Willems, R.
    Nuyens, D.
    Heidbuchel, H.
    [J]. ACTA CARDIOLOGICA, 2012, 67 (05) : 624 - 625
  • [4] Image Quality of Low-Dose Cerebral Angiography and Effectiveness of Clinical Implementation on Diagnostic and Neurointerventional Procedures for Intracranial Aneurysms
    Choi, J.
    Kim, B.
    Choi, Y.
    Shin, N. Y.
    Jang, J.
    Choi, H. S.
    Jung, S. L.
    Ahn, K. J.
    [J]. AMERICAN JOURNAL OF NEURORADIOLOGY, 2019, 40 (05) : 827 - 833
  • [5] Low-dose CT angiography of the lower extremities: a comparison study of image quality and radiation dose
    Park, S.
    Park, S. H.
    Hwang, J. H.
    Kim, J. H.
    Lee, K. H.
    Shin, J. H.
    Pak, S. Y.
    Kang, J. M.
    [J]. CLINICAL RADIOLOGY, 2021, 76 (02) : 156.e19 - 156.e26
  • [6] A novel protocol for three-dimensional rotational venography with low-dose contrast media in preoperative angiography of brain tumours
    Kashimoto, Kimiaki
    Asai, Katsunori
    Kinoshita, Manabu
    Okita, Yoshiko
    Tanabe, Shogo
    Yamane, Yasuhiko
    Kawamata, Minoru
    Yoneda, Akitoshi
    Nakanishi, Katsuyuki
    [J]. NEURORADIOLOGY JOURNAL, 2019, 32 (06): : 452 - 457
  • [7] Application of low injection rate and low contrast agent dose in three-dimensional rotational digital subtraction angiography of the intracranial aneurysm
    Gao, Zhimei
    Zeng, Yongming
    Sun, Jingkun
    Yang, Jingjie
    Zhou, Yang
    Zhou, Mi
    Song, Tingni
    [J]. INTERVENTIONAL NEURORADIOLOGY, 2016, 22 (03) : 287 - 292
  • [8] Evaluating Image Quality and Radiation Dose in Low-Dose Thoraco-Abdominal CT Angiography with a Tin Filter for Patients with Aortic Disease
    Oh, Chang Hoon
    Cho, Soo Buem
    Kwon, Hyeyoung
    [J]. JOURNAL OF CLINICAL MEDICINE, 2024, 13 (04)
  • [9] Three-dimensional rotational angiography for craniotomy planning and postintervention evaluation of intracranial aneurysms
    Pedicelli, A.
    Desiderio, F.
    Esposito, G.
    Rollo, M.
    Albanese, A.
    Verdolotti, T.
    D'Argento, F.
    Bonomo, L.
    Maira, G.
    Colosimo, C.
    [J]. RADIOLOGIA MEDICA, 2013, 118 (03): : 415 - 430
  • [10] Preoperative evaluation of intracranial aneurysms: Usefulness of three-dimensional rotational angiography in conjunction with conventional angiography
    Hirai, T
    Korogi, Y
    Ono, K
    Kawanaka, K
    Uemura, S
    Takahashi, M
    [J]. RADIOLOGY, 2001, 221 : 242 - 242