摘要
Dried blood spots (DBS) may be a promising alternative specimen type to plasma for measuring the viral load (VL) in HIV-infected individuals in resource-limited settings. However, characterization of assay performance using DBS is incomplete. In this prospective study, the VL was measured in parallel using plasma and DBS specimens collected at the same time from 157 HIV-1-infected individuals. DBS were prepared by dispensing 50 mu l of blood onto filter paper cards and were stored desiccated at -20 degrees C. Nucleic acid extraction from plasma and DBS was performed automatically using the Abbott m2000sp instrument, and the VL was measured using the RealTime HIV-1 VL assay, which has a lower limit of detection of 40 HIV RNA copies/ml. The correlation between plasma and DBS results was good (R = 0.91; P < 0.001). The mean difference in the VL (DBS minus plasma) was 0.35 log copies (standard deviation [SD], 0.47 log copies). A total of 40 (26%) paired specimens had a difference of >0.5 log copy, and in 12 (7.8%) it was >1 log copy. the VL from DBS was measurable in 95.7% of specimens with a plasma VL of >2.74 log copies (550 HIV RNA copies/ml). In summary, the VL can reliably be measured using DBS with the Abbott RealTime HIV-1 assay. The estimated lower limit of detection of this automated methodology on DBS is 550 copies/ml, a threshold that may be acceptable for periodic VL monitoring in patients on antiretroviral therapy in resource-limited settings, where early detection of virologic treatment failure is often problematic.
引用
收藏
页码:569 / 572
页数:4
相关论文
共 50 条
  • [41] Comparison of the Cepheid GeneXpert and Abbott M2000 HIV-1 real time molecular assays for monitoring HIV-1 viral load and detecting HIV-1 infection
    Ceffa, Susanna
    Luhanga, Richard
    Andreotti, Mauro
    Brambilla, Davide
    Erba, Fulvio
    Jere, Haswel
    Mancinelli, Sandro
    Giuliano, Marina
    Palombi, Leonardo
    Marazzi, Maria Cristina
    JOURNAL OF VIROLOGICAL METHODS, 2016, 229 : 35 - 39
  • [42] Separation of Plasma from Whole Blood by Use of the cobas Plasma Separation Card: a Compelling Alternative to Dried Blood Spots for Quantification of HIV-1 Viral Load
    Carmona, Sergio
    Seiverth, Britta
    Magubane, Dieketseng
    Hans, Lucia
    Hoppler, Matthias
    JOURNAL OF CLINICAL MICROBIOLOGY, 2019, 57 (04)
  • [43] Comparison of HIV-1 detection in plasma specimens and dried blood spots using the Roche COBAS Ampliscreen HIV-1 test in Kisumu, Kenya
    Okonji, Jully A.
    Basavaraju, Sridhar V.
    Mwangi, Jane
    Shiraishi, Ray W.
    Odera, Matthew
    Ouma, Kenneth
    Pitman, John P.
    Marum, Lawrence H.
    Ou, Chin-Yih
    Zeh, Clement
    JOURNAL OF VIROLOGICAL METHODS, 2012, 179 (01) : 21 - 25
  • [44] Design and Implementation of an External Quality Assessment Program for HIV Viral Load Measurements Using Dried Blood Spots
    Prach, Lisa M.
    Puren, Adrian
    Lippman, Sheri A.
    Carmona, Sergio
    Stephenson, Sophie
    Cutler, Ewalde
    Barnhart, Scott
    Liegler, Teri
    JOURNAL OF CLINICAL MICROBIOLOGY, 2015, 53 (03) : 964 - 966
  • [45] Use of Dried Plasma Spots for HIV-1 Viral Load Determination and Drug Resistance Genotyping in Mexican Patients
    Pablo Rodriguez-Auad, Juan
    Rojas-Montes, Othon
    Maldonado-Rodriguez, Angelica
    Teresa Alvarez-Munoz, Ma.
    Munoz, Onofre
    Torres-Ibarra, Rocio
    Vazquez-Rosales, Guillermo
    Lira, Rosalia
    BIOMED RESEARCH INTERNATIONAL, 2015, 2015
  • [46] HIV-1 and HCV detection in dried blood spots by SYBR Green multiplex real-time RT-PCR
    De Crignis, Elisa
    Re, Maria Carla
    Cimatti, Laura
    Zecchi, Lisa
    Gibellini, Davide
    JOURNAL OF VIROLOGICAL METHODS, 2010, 165 (01) : 51 - 56
  • [47] Evaluation of dried blood spots for HIV-1 viral load and drug resistance genotyping in patients on antiretroviral therapy
    Rottinghaus, E. K.
    Ugbenga, R.
    Diallo, K.
    Bassey, O.
    Vedapuri, S.
    Azeez, A.
    DeVos, J.
    Wurie, I.
    Aberle-Grasse, J.
    Nkengasong, J.
    Knight, N.
    Yang, C.
    ANTIVIRAL THERAPY, 2010, 15 : A148 - A148
  • [48] Field evaluation of Abbott Real Time HIV-1 Qualitative test for early infant diagnosis using dried blood spots samples in comparison to Roche COBAS Ampliprep/COBAS TaqMan HIV-1 Qual Test in Kenya
    Chang, Joy
    Omuomo, Kenneth
    Anyango, Emily
    Kingwara, Leonard
    Basiye, Frank
    Morwabe, Alex
    Shanmugam, Vedapuri
    Shon Nguyen
    Sabatier, Jennifer
    Zeh, Clement
    Ellenberger, Dennis
    JOURNAL OF VIROLOGICAL METHODS, 2014, 204 : 25 - 30
  • [49] Qualitative Real-Time PCR Assay for HIV-1 and HIV-2 RNA
    Yamazaki, Sayaka
    Kondo, Makiko
    Sudo, Koji
    Ueda, Tomoyuki
    Fujiwara, Hiroshi
    Hasegawa, Naoki
    Kato, Shingo
    JAPANESE JOURNAL OF INFECTIOUS DISEASES, 2016, 69 (05) : 367 - 372
  • [50] A field evaluation of HIV-1 RNA viral load in plasma and dried blood spots using the COBAS Amplicor Analyzer and the Nucleic Acid Extraction COBAS Ampliprep/Taqman in Kisumu, Kenya
    Ouma, K.
    Basavaraju, S.
    Okonji, J.
    Williamson, J.
    Mills, L.
    Zeh, C.
    JOURNAL OF THE INTERNATIONAL AIDS SOCIETY, 2012, 15 : 37 - 38