Comparison of non-contact infrared forehead thermometer to standard temperature measurement in neonatal intensive care unit patients

被引:8
|
作者
Can, Emrah [1 ]
Bulbul, Ali [1 ]
Uslu, Sinan [1 ]
Nuhoglu, Asiye [1 ]
机构
[1] Arastirma Hastanesi Yenidogan Klin, Istanbul, Turkey
关键词
Body temperature; digital thermometer; mercury thermometer; newborn; non-contact infrared forehead thermometer; TYMPANIC THERMOMETRY; URINARY-BLADDER; FEVER; AXILLARY; CHILDREN; MERCURY;
D O I
10.4274/tpa.45.257
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Aim: In this study, we aimed to compare non-contact infrared forehead thermometer measurement with the mercury thermometer and digital thermometer measurements in the neonatal period. Material and Method: A total of 179 newborns who were admitted to our hospital were enrolled, and simultaneous temperature measurements were performed via three devices. Mesaurements were made from forehead with non-contact infrared thermometer and from axillar region with digital thermometers and conventional thermometers. Thermoflash LX-26 (France-China) thermometer was used for non-contact infrared mesasurement, MC-203-E (Omron Health Care, Europe) thermometer for digital measurement and glass mercury thermometer for conventional axillary measurement. A total of 6273 measurements were performed, being 2091 measurement for each method. Results: The mean results of the non-contact infrared forehead thermometer, mercury thermometers and digital thermometer were found as 37.2 +/- 0.5 degrees C, 36.7 +/- 0.4 degrees C, and 36.6 +/- 0.4 degrees C respectively. The Bland-Altman plots differences suggest that 95% of the non-contact infrared forhead thermometer readings were within the limits of agreement, which is -0.37 and +1.54 degrees C range of mercury thermometer and -0.40 and +1.54 degrees C range of digital thermometer respectively. The sensitivity and specificity of the non-contact infrared thermometer were found as 81% and 50% respectively [ 95% CI (41.6-58.4)]. Conclusions: We conclude that the non-contact infrared thermometer can not be recommended for assessment of body temperature in newborns admitted to neonatal intensive care unit. (Turk Arch Ped 2010; 45: 257-63)
引用
收藏
页码:257 / 263
页数:7
相关论文
共 50 条
  • [1] Correction of Human Forehead Temperature Variations Measured by Non-Contact Infrared Thermometer
    Shajkofci, Adrian
    [J]. IEEE SENSORS JOURNAL, 2022, 22 (17) : 16750 - 16755
  • [2] Comparison of Infrared Tympanic Thermometer with Non-Contact Infrared Thermometer
    Bayhan, Cihangul
    Ozsurekci, Yasemin
    Tekcam, Necmiye
    Guloglu, Ayse
    Ehliz, Gulten
    Ceyhan, Mehmet
    Kara, Ates
    [J]. JOURNAL OF PEDIATRIC INFECTION, 2014, 8 (02): : 52 - 55
  • [3] Comparison of Infrared Tympanic Thermometer with Non-Contact Infrared Thermometer
    Ilce, Arzu
    [J]. JOURNAL OF PEDIATRIC INFECTION, 2014, 8 (04): : 200 - 200
  • [4] Clinical Accuracy of Non-Contact Forehead Infrared Thermometer Measurement in Children: An Observational Study
    Kim, Yeon-Mi
    Jang, Myung-Roul
    Moon, Ju-Ryoung
    Park, Goeun
    An, Ye-Jin
    Seo, Jeong-Meen
    [J]. CHILDREN-BASEL, 2022, 9 (09):
  • [5] Non-contact monitoring of respiration in the neonatal intensive care unit
    Jorge, Joao
    Villarroel, Mauricio
    Chaichulee, Sitthichok
    Guazzi, Alessandro
    Davis, Sara
    Green, Gabrielle
    McCormick, Kenny
    Tarassenko, Lionel
    [J]. 2017 12TH IEEE INTERNATIONAL CONFERENCE ON AUTOMATIC FACE AND GESTURE RECOGNITION (FG 2017), 2017, : 286 - 293
  • [6] Clinical accuracy of non-contact forehead infrared thermometer and infrared tympanic thermometer in postoperative adult patients: A comparative study.
    Kameda, Norihiro
    [J]. JOURNAL OF PERIOPERATIVE PRACTICE, 2022, 32 (06) : 142 - 148
  • [7] Performance characterization of non-contact infrared thermometers (NCITs) for forehead temperature measurement
    Sullivan, Stacey J. L.
    Seay, Nathanael
    Zhu, Liang
    Rinaldi, Jean E.
    Hariharan, Prasanna
    Vesnovsky, Oleg
    Topoleski, L. D. Timmie
    [J]. MEDICAL ENGINEERING & PHYSICS, 2021, 93 : 93 - 99
  • [8] Donor Temperature Screening with a Non-Contact Infrared Thermometer
    Navarro, Marie
    Molfese, Paul M.
    Rioveros, Jowin
    Villasin, Emylee
    Godoy, Jessica
    Ziman, Alyssa
    McGonigle, Andrea M.
    Mei, Zhen Wei
    Ward, Dawn C.
    [J]. TRANSFUSION, 2022, 62 : 151A - 151A
  • [9] Assessment of Non-Contact Infrared Thermometer Measurement Sites in Birds
    Anderson, Jessica
    Kaplan-Stein, Sara
    Adolph, Stephen
    Peralta, Jose M.
    [J]. JOURNAL OF APPLIED ANIMAL WELFARE SCIENCE, 2020, 23 (02) : 131 - 139
  • [10] Comparison of infrared tympanic, non-contact infrared skin, and axillary thermometer to rectal temperature measurements in a pediatric emergency observation unit
    Berksoy, Emel Atas
    Anil, Murat
    Bicilioglu, Yuksel
    Gokalp, Gamze
    Bal, Alkan
    [J]. INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2018, 11 (02): : 567 - 573