Calibration transfer for bioprocess Raman monitoring using Kennard Stone piecewise direct standardization and multivariate algorithms

被引:17
|
作者
Petillot, Laure [1 ]
Pewny, Fiona [2 ]
Wolf, Martin [2 ]
Sanchez, Celia [1 ]
Thomas, Fabrice [1 ]
Sarrazin, Johan [1 ]
Fauland, Katharina [2 ]
Katinger, Hermann [2 ]
Javalet, Charlotte [1 ]
Bonneville, Christophe [1 ]
机构
[1] RESOLUT Spectra Syst, Meylan, France
[2] Polymun Sci, Klosterneuburg, Austria
基金
欧盟地平线“2020”;
关键词
cell culture monitoring; model transferability; multivariate model maintenance; process analytical technology; Raman spectroscopy; ORTHOGONAL SIGNAL CORRECTION; LEAST-SQUARES; SPECTROSCOPY; IMPROVEMENT; PREDICTION;
D O I
10.1002/eng2.12230
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
In the biopharmaceutical industry, Raman spectroscopy is now a proven PAT tool that enables in-line simultaneous monitoring of several CPPs and CQAs in real-time. However, as Raman monitoring requires multivariate modeling, variabilities unknown by models can impact the monitoring prediction accuracy. With the widespread use of Raman PAT tools, it is necessary to fix instrumental variability impacts, encountered for instance during a device replacement. In this work, we investigated the impact of instrumental variability between probes inside a multi-channel analyzer and between two analyzers, and explored solutions to correct them on model prediction errors in cell cultures. It is shown that the Kennard Stone piecewise direct standardization (KS PDS) method enables to lower model prediction errors between probes of a multi-channel analyzer from 20% to 10% on the cell densities (TCD/VCD). Considering the integration of a new device or the replacement of a previous one, it has been determined that a first cell culture monitoring can be directly performed with the new analyzer calibrated by the KS PDS method based on the dataset from the previous analyzer, with an accuracy better than 10% on the main components of the culture like glucose, lactate, and the cell densities. Then, the new data obtained by the new analyzer can be inserted in a global calibration dataset to integrate instrumental variability in the chemometric model: it is shown that only one batch with the new device in a consistent and equilibrated calibration dataset was sufficient to correct the prediction gap induced by instrumental variability, allowing to exploit the data from previous analyzers considering optimized methods. This methodology provides good multivariate calibration model prediction errors throughout the instrumental changes which is a requirement for model maintenance.
引用
收藏
页数:19
相关论文
共 11 条
  • [1] The Application of Piecewise Direct Standardization with SNV in Calibration Transfer of Raman Spectra
    Huang Cheng-wei
    Dai Lian-kui
    Dong Xue-feng
    SPECTROSCOPY AND SPECTRAL ANALYSIS, 2011, 31 (05) : 1279 - 1282
  • [2] Improvement of the piecewise direct standardisation procedure for the transfer of NIR spectra for multivariate calibration
    Bouveresse, E
    Massart, DL
    CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1996, 32 (02) : 201 - 213
  • [3] Pharmaceutical analysis in solids using front face fluorescence spectroscopy and multivariate calibration with matrix correction by piecewise direct standardization
    Alves, Julio Cesar L.
    Poppi, Ronei J.
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2013, 103 : 311 - 318
  • [4] On-line bioprocess monitoring with a multi-wavelength fluorescence sensor using multivariate calibration
    Skibsted, E
    Lindemann, C
    Roca, C
    Olsson, L
    JOURNAL OF BIOTECHNOLOGY, 2001, 88 (01) : 47 - 57
  • [5] Calibration transfer and drift counteraction in chemical sensor arrays using Direct Standardization
    Fonollosa, J.
    Fernandez, L.
    Gutierrez-Galvez, A.
    Huerta, R.
    Marco, S.
    SENSORS AND ACTUATORS B-CHEMICAL, 2016, 236 : 1044 - 1053
  • [6] Pharmaceutical analysis in solids using front face fluorescence spectroscopy and multivariate calibration with matrix correction by piecewise direct standardization (vol 103, pg 311, 2013)
    Alves, Julio Cesar L.
    Poppi, Ronei J.
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2014, 121 : 751 - 751
  • [7] Bioprocess in-line monitoring using Raman spectroscopy and Indirect Hard Modeling (IHM): A simple calibration yields a robust model
    Mueller, David Heinrich
    Flake, Carsten
    Brands, Thorsten
    Koss, Hans-Juergen
    BIOTECHNOLOGY AND BIOENGINEERING, 2023, 120 (07) : 1857 - 1868
  • [8] A calibration transfer methodology for Standardization of Raman instruments with different spectral resolutions using Double Digital Projection Slit
    Li, Qifeng
    Sun, Xueqing
    Ma, Xiangyun
    Li, Bin
    Wang, Huijie
    Lv, Haiyue
    Wang, Qing
    Xu, Kexin
    Chen, Da
    CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2019, 191 : 143 - 147
  • [9] Advanced monitoring and control of pharmaceutical production processes with Pichia pastoris by using Raman spectroscopy and multivariate calibration methods
    Voss, Jan-Patrick
    Mittelheuser, Nina E.
    Lemke, Roman
    Luttmann, Reiner
    ENGINEERING IN LIFE SCIENCES, 2017, 17 (12): : 1281 - 1294
  • [10] A novel multivariate approach using science-based calibration for direct coating thickness determination in real-time NIR process monitoring
    Moeltgen, C. -V.
    Herdling, T.
    Reich, G.
    EUROPEAN JOURNAL OF PHARMACEUTICS AND BIOPHARMACEUTICS, 2013, 85 (03) : 1056 - 1063