Constructing a Device for Calculating the Atmospheric Correction Factor for the Calibration the Radiation-Absorbed Dose in Radiotherapy

被引:0
|
作者
Hong, Joo-Wan [1 ]
Lee, Jin-Hyeok [2 ]
Ahn, Jae-Ouk [3 ]
Yu, Dong-Su [3 ]
Heo, Yeong-Cheol [4 ]
机构
[1] Seoul Natl Univ, Dept Radiat Oncol, Bundang Hosp, Seongnam 13620, South Korea
[2] Seoul Vet Hosp, Dept Nucl Med, Seoul 05368, South Korea
[3] Soonchunhyang Univ, Dept Med IT Engn, Asan 31538, South Korea
[4] EulJi Univ, Dept Radiol Sci, Coll Hlth Sci, Seongnam 13135, South Korea
基金
新加坡国家研究基金会;
关键词
MCU; Atmospheric correction factor QA device; Atmospheric correction factor;
D O I
10.3938/jkps.72.353
中图分类号
O4 [物理学];
学科分类号
0702 ;
摘要
The purpose of this study was to measure the air temperature and the air pressure, develop an atmospheric correction factor calculation device (ACFCD), and utilize it. Because the radiation dose varies with the temperature, pressure, and humidity, is essential calibrating doses after calculating the atmospheric correction factor. In this study, we constructed an ACFCD by using a micro controller, an air thermometer, and an air pressure sensor. We measured the temperature and the air pressure in nine different spaces and confirmed the changes in the factors. We obtained the measured data as comma-separated-values files, and we created a Microsoft Excel macro for the analysis to calculate the average temperature, pressure, and atmospheric correction factor; we determined trends by using graphs. As a result of the performance evaluation, we confirmed that the space temperature and the air pressure measurement were the same (p > 0.05). In addition, we confirmed that the micro controller unit measured the heating, which we observed to be 3.9 au broken vertical bar C. The ACFCD that we developed in this study confirmed the possibility of replacing existing air thermometers and barometers and can contribute to improving work efficiency. In addition, in light of the basic concept of the ACFCD, we consider that the device will be useful for ensuring safe, accurate use of radiation once it is made more practical through additional studies.
引用
收藏
页码:353 / 358
页数:6
相关论文
共 50 条
  • [31] A model for calculating the radiation absorbed dose (RAD) to the myocardium from radioimmunotherapy (RAIT) with monoclonal antibodies (MABS)
    Juweid, M
    Mardirossian, G
    Petorius, H
    Goldenberg, DM
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE, 1999, 26 (09): : 1049 - 1049
  • [32] Calculation of the Radiation Dose Absorbed by Human Biological Tissue When Using Imaging Equipment in Radiotherapy
    Lebedenko, I. M.
    Khromov, S. S.
    Bondarenko, T. V.
    Chertenkov, E. M.
    MEASUREMENT TECHNIQUES, 2021, 64 (01) : 60 - 63
  • [33] Calculation of the Radiation Dose Absorbed by Human Biological Tissue When Using Imaging Equipment in Radiotherapy
    I. M. Lebedenko
    S. S. Khromov
    T. V. Bondarenko
    E. M. Chertenkov
    Measurement Techniques, 2021, 64 : 60 - 63
  • [34] RADIATION DOSE-RATE - FACTOR OF IMPORTANCE IN RADIOBIOLOGY AND RADIOTHERAPY
    HALL, EJ
    BRITISH JOURNAL OF RADIOLOGY, 1972, 45 (530): : 81 - +
  • [35] Design and assembly of an IoT-based device to determine the absorbed dose of gamma and UV radiation
    Baena-Navarro, R.
    Torres-Hoyos, F.
    Uc-Rios, C.
    Colmenares-Quintero, R. F.
    APPLIED RADIATION AND ISOTOPES, 2020, 166
  • [36] Evaluation of resistance of CMOS LSIC to the factor of absorbed dose under the pulsed radiation effect
    Sogoyan A.V.
    Russian Microelectronics, 2011, 40 (3) : 185 - 193
  • [37] THE ATMOSPHERIC PRESSURE CORRECTION FACTOR IN ROENTGEN-RAY DOSE MEASUREMENT
    LOWRY, WSB
    AMERICAN JOURNAL OF ROENTGENOLOGY RADIUM THERAPY AND NUCLEAR MEDICINE, 1958, 79 (01): : 149 - 150
  • [38] Performance of a device to minimise radiation dose to the hands during radioactive syringe calibration
    Lindner, O
    Busch, F
    Burchert, W
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2003, 30 (06) : 819 - 825
  • [39] Performance of a device to minimise radiation dose to the hands during radioactive syringe calibration
    Oliver Lindner
    Frank Busch
    Wolfgang Burchert
    European Journal of Nuclear Medicine and Molecular Imaging, 2003, 30 : 819 - 825
  • [40] Comparisons of air kerma and absorbed dose to water standards in Co-60 radiation beams for radiotherapy
    Kottler, Christian
    Steurer, Andreas
    Delaunay, Frank
    de Pooter, Jacco
    Mulas, Cristina Garcia
    Mihailescu, Liviu-Cristian
    METROLOGIA, 2020, 57 (1A)