An automated ionization chamber for secondary radioactivity standards

被引:15
|
作者
Fitzgerald, R. [1 ]
机构
[1] Natl Inst Stand & Technol, Phys Lab, Gaithersburg, MD 20899 USA
关键词
Ionization chamber; Radioactivity;
D O I
10.1016/j.apradiso.2009.12.008
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
I report on the operation and characterization of a new ionization chamber system. "AUTOIC", featuring a commercial digital electrometer and a commercial robotic sample changer The relative accuracy of the electrometer was improved significantly beyond the manufacturer's specifications through an in-house calibration of the various ranges, applied via software The measurement precision and repeatability of the system have been determined by measuring multiple samples of the same radionuclide over the span of two or three years The linearity of the system was examined by following the decay of Tc-99m, Mo-99 and Xe-133 sources for up to 19 half-lives and determining half-life values All of these values agree with the accepted literature values, within their combined uncertainties. Published by Elsevier Ltd
引用
收藏
页码:1507 / 1509
页数:3
相关论文