Analytic variability due to change of deficient plasma vials: application to one-stage clotting factor VIII assay

被引:5
|
作者
Nougier, Christophe [1 ]
Sobas, Frederic [1 ]
Nguyen, Thi K. [1 ]
Carage, Marie L. [1 ]
Lienhart, Anne [1 ]
Negrier, Claude [1 ]
机构
[1] Hop Edouard Herriot, Haemostasis Lab, F-69437 Lyon, France
关键词
efficiency; factor VIII deficient plasmas; one-stage clotting factor VIII assay;
D O I
10.1097/MBC.0b013e328343300e
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The statistical process control required under International Organization for Standardization 15189 as well as economic considerations necessitates having robust methods that do not need systematic recalibration for each series of analyses. Using the concrete example of one-stage clotting factor VIII assay, we assessed the analytic variability specifically linked to changing factor VIII deficient plasma vials. The study used freeze-dried (Instrumentation Laboratory, Siemens, Stago and T-Coag) and frozen (Affinity Biologicals and Precision Biologic) factor VIII deficient plasmas. On the most widely recognized acceptability criteria and methods (i.e. those of Kallner et al. and Kasper et al.), the Stago and Instrumentation Laboratory plasmas require systematic recalibration at each vial changeover. Blood Coagul Fibrinolysis 22:151-154 (C) 2011 Wolters Kluwer Health vertical bar Lippincott Williams & Wilkins.