Excitation module for orthogonal fundamental mode fluxgate sensor

被引:5
|
作者
Song, S. X. [1 ]
Deng, M. [1 ]
Chen, K. [1 ]
Yuan, Z. Z. [3 ]
Xu, Z. F. [2 ]
Zhang, Q. S. [1 ]
机构
[1] China Univ Geosci Beijing, Sch Geophys & Informat Technol, Beijing, Peoples R China
[2] Beijing Huahang Radio Measurement Inst, Beijing, Peoples R China
[3] China Key Syst & Integrated Circuit Co Ltd, Wuxi, Jiangsu, Peoples R China
来源
JOURNAL OF INSTRUMENTATION | 2020年 / 15卷 / 05期
关键词
Analogue electronic circuits; Modular electronics; Front-end electronics for detector readout; NOISE; MECHANISM; WIRE;
D O I
10.1088/1748-0221/15/05/P05007
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A fluxgate is a device that has important applications in weak magnetic field measurements. Compared with the second harmonic mode, the orthogonal fundamental mode fluxgate exhibits a better performance in terms of lower noise and wider bandwidth, making it popular among researchers in this field. An amorphous wire is a vital part of any orthogonal fundamental mode fluxgate, and its excitation parameters determine the performance of the sensor (especially the noise level). For different structures and sizes of amorphous wires as probes, the optimal excitation parameters of orthogonal fundamental mode fluxgate are different. The conventional excitation module development scheme suffers from problems such as a complicated structure and low flexibility. To overcome these problems, a simple, compact, and effective excitation module is proposed, which is composed of a waveform generator, signal conditioning circuit, and voltage-controlled current source. Owing to the high flexibility, the frequency of the direct digital frequency synthesis is fully programmable. We tested the design performance indexes of key parameters, such as excitation signal waveform, amplitude, noise level, and bandwidth. The excitation current noise level is as low as 12.42 nA/sqrt (Hz)@ 20 kHz without I-ac and 17.35 nA/sqrt (Hz)@ 20 kHz with I-ac. The effective bandwidth of the alternating current output is 10-200 kHz. Moreover, a system test platform was developed, and this module was used to analyze the variation trend of the noise level of the fluxgate sensor under different excitation parameters (such as I-ac, I-dc, and excitation current frequency). It has been proved that the excitation module helps in effectively and quickly selecting the best excitation parameters to improve the performance of orthogonal fundamental mode fluxgates and shorten their development cycle.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Research on the Orthogonal Fundamental Mode Fluxgate Sensor Circuit
    Yuan, Zhenzhong
    Zhang, Yuxin
    Wang, Dong
    Jiang, Yingdan
    Guo, Ronghai
    IEEE ACCESS, 2020, 8 : 150665 - 150671
  • [2] Modeling and Performances of the Orthogonal Fluxgate Sensor Operated in Fundamental Mode
    Jiang, Hao
    Zhang, Jilong
    Xu, Zhenfeng
    Xu, Yufeng
    Chen, Kai
    Li, Guangmin
    Zeng, Pan
    Qu, Timing
    IEEE TRANSACTIONS ON MAGNETICS, 2020, 56 (02)
  • [3] Fundamental Mode Orthogonal Fluxgate Gradiometer
    Sasada, Ichiro
    Harada, Shoumu
    IEEE TRANSACTIONS ON MAGNETICS, 2014, 50 (11)
  • [4] Noise correlation in fundamental mode orthogonal fluxgate
    Butta, M.
    Sasada, I.
    JOURNAL OF APPLIED PHYSICS, 2012, 111 (07)
  • [5] Towards digital fundamental mode orthogonal fluxgate
    Butta, Mattia
    2016 IEEE SENSORS APPLICATIONS SYMPOSIUM (SAS 2016) PROCEEDINGS, 2016, : 144 - 148
  • [6] The fundamental mode orthogonal fluxgate and its applications
    Sasada I.
    Sasada, Ichiro (sasada.ichiro.i65@m.kyushu-u.ac.jp), 1600, Institute of Electrical Engineers of Japan (137): : 448 - 453
  • [7] Fluxgate Sensor with Bifactor Excitation Mode
    Bryakin, Ivan V.
    Bochkarev, Igor V.
    Khramshin, Vadim R.
    Gasiyarov, Vadim R.
    SENSORS, 2023, 23 (04)
  • [8] Gradiometer and Magnetometer Integration Using a Pair of Fundamental Mode Orthogonal Fluxgate Sensor Heads
    Elrefai, Ahmed L.
    Sasada, Ichiro
    Harada, Shoumu
    IEEE TRANSACTIONS ON MAGNETICS, 2015, 51 (11)
  • [9] Gradiometer and Magnetometer Integration by Using a Pair of Fundamental Mode Orthogonal Fluxgate Sensor Heads
    Sasada, I.
    Harada, S.
    Elrefai, A. L.
    2015 IEEE MAGNETICS CONFERENCE (INTERMAG), 2015,
  • [10] Magnetocardiogram measured by fundamental mode orthogonal fluxgate array
    Karo, Hikaru
    Sasada, Ichiro
    JOURNAL OF APPLIED PHYSICS, 2015, 117 (17)