Electrochemiluminescent biosensors array for the concomitant detection of choline, glucose, glutamate, lactate, lysine and urate

被引:102
|
作者
Marquette, CA [1 ]
Degiuli, A [1 ]
Blum, LJ [1 ]
机构
[1] Univ Lyon 1, CNRS, UMR 5013, Lab Genie Enzymat & Biomol, F-69622 Villeurbanne, France
来源
BIOSENSORS & BIOELECTRONICS | 2003年 / 19卷 / 05期
关键词
choline; electrochimiluminescence; glucose; glutamate; lactate; lysine; multifunctional bio-sensing chip; uric acid;
D O I
10.1016/S0956-5663(03)00225-2
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A multifunctional bio-sensing chip was designed based on the electrochemiluminescent (ECL) detection of enzymatically produced hydrogen peroxide. Six different oxidases specific for choline, glucose, glutamate, lactate, lysine and urate were non-covalently immobilised on imidodiacetic acid chelating beads (glucose oxidase only) or on diethylaminoethyl (DEAE) anion exchanger beads, and spotted on the surface of a glassy carbon foil (25 mm(2) square), entrapped in PVA-SbQ photopolymer. The chip measurement was achieved by applying during 3 min a +850 mV potential between the glassy carbon electrode and a platinum pseudo-reference, while capturing a numeric image of the multifunctional bio-sensing chip with a CCD camera. The use of luminol supporting beads (DEAE-Sepharose) included in the sensing layer was shown to enable the achievement of spatially well defined signals, and to solve the hydrogen peroxide parasite signal which appeared between contiguous spots using luminol free in solution. The detection limits of the different biosensor were found to be 1 muM for glutamate, lysine and uric acid, 20 muM for glucose and 2 muM for choline and lactate. The detection ranges were 1-25 muM (uric acid), 1 muM-0.5 mM (glutamate and lysine), 20 muM-2 mM (glucose) and 2 muM-0.2 mM (choline and lactate). The ECL chip was used for the detection of glucose, lactate and uric acid in human serum matrix. Good correlations between measured and expected values were found without the need of internal calibration of the sample, demonstrating the potentiality of the ECL multifunctional bio-sensing chip. (C) 2003 Elsevier B.V. All rights reserved.
引用
收藏
页码:433 / 439
页数:7
相关论文
共 21 条
  • [1] Amperometric biosensors for detection of choline and lactate by Choline Oxidase and Lactate Oxidase immobilisation in microporous membranes
    Makela, Timo
    Albers, Martin
    VTT Tiedotteita - Valtion Teknillinen Tutkimuskeskus, (1775):
  • [2] Self-containing reactant biochips for the electrochemiluminescent determination of glucose, lactate and choline
    Marquette, CA
    Blum, LJ
    SENSORS AND ACTUATORS B-CHEMICAL, 2003, 90 (1-3) : 112 - 117
  • [3] Visualization analysis of glucose, lactate and cholesterol based on an electrochemiluminescent biosensor array
    Jiajia Chen
    Fengting Dai
    Haotian Liu
    Zhengguang He
    Hui Gao
    Gen Liu
    Journal of Materials Science: Materials in Electronics, 2023, 34
  • [4] Visualization analysis of glucose, lactate and cholesterol based on an electrochemiluminescent biosensor array
    Chen, Jiajia
    Dai, Fengting
    Liu, Haotian
    He, Zhengguang
    Gao, Hui
    Liu, Gen
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN ELECTRONICS, 2023, 34 (05)
  • [5] A highly selective amperometric biosensor array for the simultaneous determination of glutamate, glucose, choline, acetylcholine, lactate and pyruvate
    Kucherenko, D. Yu.
    Kucherenko, I. S.
    Soldatkin, O. O.
    Topolnikova, Ya. V.
    Dzyadevych, S. V.
    Soldatkin, A. P.
    BIOELECTROCHEMISTRY, 2019, 128 : 100 - 108
  • [6] MICROFABRICATED GLUCOSE, LACTATE, GLUTAMATE AND GLUTAMINE THIN-FILM BIOSENSORS
    SCHALKHAMMER, T
    LOBMAIER, C
    ECKER, B
    WAKOLBINGER, W
    KYNCLOVA, E
    HAWA, G
    PITTNER, F
    SENSORS AND ACTUATORS B-CHEMICAL, 1994, 19 (1-3) : 587 - 5591
  • [7] Microfabricated glucose, lactate, glutamate and glutamine thin-film biosensors
    Schalkhammer, Th.
    Lobamaier, Ch.
    Ecker, B.
    Wakolbinger, W.
    Kynclova, E.
    Hawa, G.
    Pittner, F.
    Sensors and Actuators, B: Chemical, 1994, B19 (1 -3 pt 2) : 587 - 591
  • [8] Electrochemiluminescence bipolar electrode array for the multiplexed detection of glucose, lactate and choline based on a versatile enzymatic approach
    Xiao, Yi
    Xu, Linru
    Qi, Lian-Wen
    TALANTA, 2017, 165 : 577 - 583
  • [9] On-line biosensors for simultaneous determination of glucose, choline, and glutamate integrated with a microseparation system
    Shi, GY
    Yamamoto, K
    Zhou, TS
    Xu, F
    Kato, T
    Jin, JY
    Jin, LT
    ELECTROPHORESIS, 2003, 24 (18) : 3266 - 3272
  • [10] Oxygen insensitive biosensors for real time detection of glutamate and glucose
    Ispas, Cristina R.
    Andreescu, Silvana
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2010, 240