共 22 条
A Modified Transvenous Single Mechanical Dilatation Technique to Remove a Chronically Implanted Active-Fixation Coronary Sinus Pacing Lead
被引:9
|作者:
Bongiorni, Maria Grazia
[1
]
Di Cori, Andrea
[1
]
Zucchelli, Giulio
[1
]
Segreti, Luca
[1
]
De Lucia, Raffaele
[1
]
Paperini, Luca
[1
]
Soldati, Ezio
[1
]
机构:
[1] Hosp Univ Pisa, Dept Cardiovasc Dis 2, Pisa, Italy
来源:
关键词:
lead extraction;
coronary sinus;
active-fixation lead;
StarFix lead;
cardiac resynchronization therapy;
CARDIAC RESYNCHRONIZATION THERAPY;
SUCCESS RATE;
SAFETY;
D O I:
10.1111/j.1540-8159.2010.02784.x
中图分类号:
R5 [内科学];
学科分类号:
1002 ;
100201 ;
摘要:
We described a 77-year-old patient, previously implanted with a dual-chamber pacemaker later upgraded to a cardiac resynchronization therapy-defibrillator (CRT-D) device with an active-fixation coronary sinus pacing lead, who underwent a transvenous mechanical extraction procedure for a device-related systemic infection. All leads were removed successfully with a transvenous approach. With regard to the coronary sinus (CS) lead (Attain 4195 StarFix, Medtronic Inc., Minneapolis, MN, USA), manual traction was ineffective and extraction required long and challenging mechanical dilatation up to distal CS using either conventional sheaths or modified CS lead delivery. (PACE 2011; 34: e66-e69)