Quantitative Analysis of Airway Tree in Low-dose Chest CT with a New Model-based Iterative Reconstruction Algorithm: Comparison to Adaptive Statistical Iterative Reconstruction in Routine-dose CT
被引:3
|作者:
Jia, Yongjun
[1
,2
]
Ji, Xing
[3
]
He, Taiping
[2
]
Yu, Yong
[2
]
Yu, Nan
[2
]
Duan, Haifeng
[2
]
Guo, Youmin
[1
]
机构:
[1] Xi An Jiao Tong Univ, Affiliated Hosp 1, Dept Radiol, Yanta Western Rd, Xian 710061, Shaanxi, Peoples R China
[2] Shaanxi Chinese Med Univ, Affiliated Hosp, Dept Radiol, Xianyang 712000, Peoples R China
[3] Yanan Univ, Affiliated Hosp, Dept Radiol, Yanan 716000, Peoples R China
关键词:
Model-based iterative reconstruction;
computer-aided detection;
tomography;
X-ray computed;
radiation dosage;
COMPUTER-AIDED DETECTION;
ABDOMINAL CT;
IDENTIFICATION;
REDUCTION;
WALL;
D O I:
10.1016/j.acra.2018.03.021
中图分类号:
R8 [特种医学];
R445 [影像诊断学];
学科分类号:
1002 ;
100207 ;
1009 ;
摘要:
Objective: We aimed to evaluate a new model-based iterative reconstruction (MBIRn) algorithm either with spatial resolution and noise reduction balance (MBIRSTND) or spatial resolution preference (MBIRRP20) for quantitative analysis of airway in low-dose chest computed tomography (CT) with a computer-aided detection (CAD) software, in comparison to adaptive statistical iterative reconstruction (ASIR) in routine-dose CT. Methods: Thirty patients who underwent both the routine-dose (noise index [NI] = 14 HU) and low-dose (at 30% level with NI = 28 HU) CT examination for pulmonary disease were included. Image acquisition was performed with 120 kVp tube voltage and automatic tube current modulation. Routine-dose scans were reconstructed with ASIR, whereas low-dose scans were reconstructed with ASIR, MBIRSTND, and MBIRRP20. Airway dimensions of the right middle lobe bronchus from the four reconstructions were measured using CAD software. Two radiologists used a semiquantitative 5 scoring criteria (-2, inferior to; +2, superior to; -1 slightly inferior to; +1, slightly superior to; and 0, equal to ASIR in routine-dose CT) to rate the subjective image quality of MBIRSTND\ and MBIRRP20 of airway trees. The paired t test and Wilcoxon signed-rank test were used for statistical comparison. Results: The low-dose CT provided 70.76% dose reduction compared to the routine-dose CT (0.88 +/- 0.83 mSv vs 3.01 +/- 1.89 mSv). MBIRSTND and MBIRRP20 with low-dose CT provided longer bronchial length measurements and were better in measurement variability and continuity and completeness of bronchial walls than ASIR in routine-dose CT (P < .05). MBIRSTND was better for subjective noise and MBIRRP20 for showing distal branches. Conclusions: MBIRSTND and MBIRRP20 algorithms provide better airway quantification at 30% of the radiation dose, compared to ASIR at routine-dose CT.