4D Shearforce-Based Constant-Distance Mode Scanning Electrochemical Microscopy

被引:46
|
作者
Nebel, Michaela [1 ]
Eckhard, Kathrin [1 ]
Erichsen, Thomas [2 ]
Schulte, Albert [3 ]
Schuhmann, Wolfgang [1 ]
机构
[1] Ruhr Univ Bochum, D-44780 Bochum, Germany
[2] Sensolytics GmbH, D-44799 Bochum, Germany
[3] Suranaree Univ Technol, Inst Sci, Sch Chem & Biochem, Biochem Electrochem Res Unit, Nakhon Ratchasima 30000, Thailand
关键词
SECM; CELLS; TIP; TOPOGRAPHY;
D O I
10.1021/ac1008805
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
4D shearforce-based constant-distance mode scanning electrochemical microscopy (4D SF/CD-SECM) is designed to assess SECM tip currents at several but constant distances to the sample topography at each point of the x,y-scanning grid. The distance dependent signal is achieved by a shearforce interaction between the in. resonance vibrating SECM tip and the sample surface. A 4.D SF/CD-SECM measuring cycle at each grid point involves a shearforce controlled SECM tip z-approach to a point of closest distance and subsequent stepwise tip retractions. At the point of closest approach and during the retraction steps, pairs of tip current (1) and position are acquired for various distances above the sample surface. Such a sequence provides x,y,I maps, that can be compiled and displayed for each selected data acquisition distance. Thus, multiple SECM images are obtained at known and constant distances above the sample topography. 4D SF/CD-SECM supports distance-controlled tip operation while continuous scanning of the SECM tip in the shear-force distance is avoided. In this way, constant-distance mode SECM imaging can be performed at user-defined, large tip-to-sample distances. The feasibility and the potential of the proposed 4D SF/CD-SECM imaging is demonstrated using on the one hand amperometric feedback mode imaging of a Pt band electrode array and on the other hand the visualization of the diffusion zone of a redox active species above a microelectrode in a generator/collector arrangement.
引用
收藏
页码:7842 / 7848
页数:7
相关论文
共 50 条
  • [21] Detection of nitric oxide release from single cells by using constant-distance-mode scanning electrochemical microscopy
    Isik, Sonnur
    Schuhmann, Wolfgang
    ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 2006, 45 (44) : 7451 - 7454
  • [22] SCANNING ELECTROCHEMICAL MICROSCOPY - A NEW SCANNING-MODE BASED ON CONVECTIVE EFFECTS
    BORGWARTH, K
    EBLING, DG
    HEINZE, J
    BERICHTE DER BUNSEN-GESELLSCHAFT-PHYSICAL CHEMISTRY CHEMICAL PHYSICS, 1994, 98 (10): : 1317 - 1321
  • [23] A Leveling Method Based on Current Feedback Mode of Scanning Electrochemical Microscopy
    Han, Lianhuan
    Yuan, Ye
    Zhang, Jie
    Zhao, Xuesen
    Cao, Yongzhi
    Hu, Zhenjiang
    Yan, Yongda
    Dong, Shen
    Tian, Zhong-Qun
    Tian, Zhao-Wu
    Zhan, Dongping
    ANALYTICAL CHEMISTRY, 2013, 85 (03) : 1322 - 1326
  • [24] 4D Scanning Ultrafast Electron Microscopy: Visualization of Materials Surface Dynamics
    Mohammed, Omar F.
    Yang, Ding-Shyue
    Pal, Samir Kumar
    Zewail, Ahmed H.
    JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2011, 133 (20) : 7708 - 7711
  • [25] The 4D Camera: An 87 kHz Direct Electron Detector for Scanning/Transmission Electron Microscopy
    Ercius, Peter
    Johnson, Ian J.
    Pelz, Philipp
    Savitzky, Benjamin H.
    Hughes, Lauren
    Brown, Hamish G.
    Zeltmann, Steven E.
    Hsu, Shang-Lin
    Pedroso, Cassio C. S.
    Cohen, Bruce E.
    Ramesh, Ramamoorthy
    Paul, David
    Joseph, John M.
    Stezelberger, Thorsten
    Czarnik, Cory
    Lent, Matthew
    Fong, Erin
    Ciston, Jim
    Scott, Mary C.
    Ophus, Colin
    Minor, Andrew M.
    Denes, Peter
    MICROSCOPY AND MICROANALYSIS, 2024, 30 (05) : 903 - 912
  • [26] 4D Scanning Transmission Ultrafast Electron Microscopy: Single-Particle Imaging and Spectroscopy
    Ortalan, Volkan
    Zewail, Ahmed H.
    JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2011, 133 (28) : 10732 - 10735
  • [27] 4D scanning transmission electron microscopy (4D-STEM) reveals crystallization mechanisms of organic semiconductors on graphene
    Zixuan Guo
    Colin Ophus
    Karen C. Bustillo
    Ryan Fair
    Stefan C. B. Mannsfeld
    Alejandro L. Briseno
    Enrique D. Gomez
    MRS Communications, 2023, 13 : 47 - 54
  • [28] Frequency-dependent alternating-current scanning electrochemical microscopy (4D AC-SECM) for local visualisation of corrosion sites
    Eckhard, Kathrin
    Erichsen, Thomas
    Stratmann, Martin
    Schuhmann, Wolfgang
    CHEMISTRY-A EUROPEAN JOURNAL, 2008, 14 (13) : 3968 - 3976
  • [29] 4D scanning transmission electron microscopy (4D-STEM) reveals crystallization mechanisms of organic semiconductors on graphene
    Guo, Zixuan
    Ophus, Colin
    Bustillo, Karen C.
    Fair, Ryan
    Mannsfeld, Stefan C. B.
    Briseno, Alejandro L.
    Gomez, Enrique D.
    MRS COMMUNICATIONS, 2023, 13 (01) : 47 - 54
  • [30] Moire Superlattice Structure of Pleated Trilayer Graphene Imaged by 4D Scanning Transmission Electron Microscopy
    Wen, Yi
    Coupin, Matthew J.
    Hou, Linlin
    Warner, Jamie H.
    ACS NANO, 2023, 17 (20) : 19600 - 19612