The Influence of Acute Pulmonary Hypertension on Cardiac Output Measurements: Calibrated Pulse Contour Analysis, Transpulmonary and Pulmonary Artery Thermodilution Against a Modified Fick Method in an Animal Model

被引:7
|
作者
Kutter, Annette P. N. [1 ]
Mosing, Martina [1 ]
Hartnack, Sonja [2 ]
Raszplewicz, Joanna [1 ]
Renggli, Martina [1 ]
Mauch, Jacqueline Y. [3 ]
Hofer, Christoph K. [4 ]
机构
[1] Univ Zurich, Vetsuisse Fac, Equine Dept, Sect Anesthesiol, CH-8057 Zurich, Switzerland
[2] Univ Zurich, Vetsuisse Fac, Epidemiol Sect, CH-8057 Zurich, Switzerland
[3] Kantonsspital Luzern, Dept Anesthesiol, Luzern, Switzerland
[4] Triemli City Hosp, Inst Anesthesiol & Intens Care Med, Zurich, Switzerland
来源
ANESTHESIA AND ANALGESIA | 2015年 / 121卷 / 01期
关键词
RELIABILITY; PRECISION; ACCURACY; MONITORS; ABILITY;
D O I
10.1213/ANE.0000000000000655
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
BACKGROUND: In critically ill patients with significant pulmonary hypertension (PH), close perioperative cardiovascular monitoring is mandatory, considering the increased morbidity and mortality in this patient group. Although the pulmonary artery catheter is still the standard for the diagnosis of PH, its use to monitor cardiac output (CO) in patients with PH is decreasing as a result of increased morbidity and possible influence of tricuspid regurgitation on the measurements. However, continuous CO measurement methods have never been evaluated under PH regarding their agreement and trending ability. In this study, we evaluated the influence of acute PH and different CO states on transpulmonary thermodilution (TPTD) and calibrated pulse contour analysis (PiCCO; both assessed with PiCCO plus (TM)), intermittent pulmonary artery thermodilution (PATD), and continuous thermodilution (CCO) compared with a modified Fick method (FICK) in an animal model. METHODS: Nine healthy pigs were studied under anesthesia. PH of 25 and 40 mm Hg (by administration of the thromboxane analog U46619), CO decreases, and CO increases were induced to test the different CO measurement techniques over a broad range of hemodynamic situations. Before each step, a new baseline data set was collected. CO values were compared using Bland-Altman analysis; trending abilities were assessed via concordance and polar plot analysis. The influence of pulmonary pressure on CO measurements was analyzed using linear mixed models. RESULTS: A mean bias of -0.26 L/min with prediction intervals of -0.88 to 1.4 L/min was measured between TPTD and FICK. Their concordance rate was 100% (94%-100% confidence interval), and the mean polar angle -3 degrees with radial limits of agreement of +/-28 degrees indicated good trending abilities. PATD compared with FICK also showed good trending ability. Comparisons of PiCCO and CCO versus FICK revealed low agreement and poor trending results with concordance rates of 84% (71%-93%) and 88% (74%-95%), mean polar angles from -17 degrees and -19 degrees, and radial limits of agreement of +/-45 degrees and 40 degrees. Pulmonary pressures influenced only the difference between FICK and PiCCO, as assessed by linear mixed models. CONCLUSIONS: TPTD compared with FICK was able to track all changes induced during the study period, including those by PH. It yielded better agreement than PATD both compared with FICK. PiCCO and CCO were not mapping all changes correctly, and when used clinically in unstable patients, regular controls with intermittent techniques are required. Acute pharmacologically induced PH did influence the difference between FICK and PiCCO.
引用
收藏
页码:99 / 107
页数:9
相关论文
共 40 条
  • [1] Comparison of Transpulmonary Thermodilution and Calibrated Pulse Contour Analysis with Pulmonary Artery Thermodilution Cardiac Output Measurements in Anesthetized Dogs
    Garofalo, N. A.
    Teixeira-Neto, F. J.
    Rodrigues, J. C.
    Cerejo, S. A.
    Aguiar, A. J. A.
    Becerra-Velasquez, D. R.
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 2016, 30 (04) : 941 - 950
  • [2] Evaluation of agreement and trending ability between transpulmonary thermodilution and calibrated pulse contour and pulse power cardiac output monitoring methods against pulmonary artery thermodilution in anesthetized dogs
    Kutter, Annette P. N.
    Bettschart-Wolfensberger, Regula
    Romagnoli, Noemi
    Bektas, Rima N.
    JOURNAL OF VETERINARY EMERGENCY AND CRITICAL CARE, 2016, 26 (04) : 531 - 540
  • [3] Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: Comparison with pulmonary arterial thermodilution
    Goedje, O
    Hoeke, K
    Lichtwarck-Aschoff, M
    Faltchauser, A
    Lamm, P
    Reichart, B
    CRITICAL CARE MEDICINE, 1999, 27 (11) : 2407 - 2412
  • [4] Cardiac output by femoral arterial thermodilution-calibrated pulse contour analysis during liver transplantation:: Comparison with pulmonary artery thermodilution
    Tzenkov, IG
    Velasco, DA
    Peña, JMP
    Arnal, LO
    Martínez, IG
    Fernández, JS
    TRANSPLANTATION PROCEEDINGS, 2003, 35 (05) : 1920 - 1922
  • [5] The level of cardiac output affects the relationship and agreement between pulmonary artery and transpulmonary aortic thermodilution measurements in an animal model
    Hueter, Lars
    Schwarzkopf, Konrad R.
    Preussler, Niels P.
    Schubert, Harald
    Schreiber, Torsten
    JOURNAL OF CARDIOTHORACIC AND VASCULAR ANESTHESIA, 2007, 21 (05) : 659 - 663
  • [6] Cardiac output-pulse contour analysis vs. pulmonary artery thermodilution
    Ostergaard, M.
    Nielsen, J.
    Rasmussen, J. P.
    Berthelsen, P. G.
    ACTA ANAESTHESIOLOGICA SCANDINAVICA, 2006, 50 (09) : 1044 - 1049
  • [7] Determination of cardiac output by the Fick method, thermodilution, and acetylene rebreathing in pulmonary hypertension
    Hoeper, MM
    Maier, R
    Tongers, J
    Niedermeyer, J
    Hohlfeld, JM
    Hamm, M
    Fabel, H
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 1999, 160 (02) : 535 - 541
  • [8] Cardiac output measurement using a modified carbon dioxide Fick method: comparison analysis with pulmonary artery catheter method and pulse induced contour cardiac output method
    Li, Tong
    Cai, Hongliu
    Pan, Hui
    Pu, Qibin
    INTERNATIONAL JOURNAL OF CLINICAL AND EXPERIMENTAL MEDICINE, 2015, 8 (03): : 3530 - 3537
  • [9] Differences in Direct Fick and Thermodilution Measurements of Cardiac Output: Impact on Pulmonary Hypertension Classification
    Hamilton, Garry W.
    Fletcher, Luke R.
    Harley, William
    Azzopardi, Robert
    Chan, R. Kimberly
    Fulcher, Jordan
    Miles, Lachlan F.
    Farouque, Omar
    Horrigan, Mark C. G.
    PULMONARY CIRCULATION, 2025, 15 (01)
  • [10] Comparison of pulmonary artery and transpulmonary thermodilution cardiac output measurements in unsedated newborn calves
    Kutter, Annette P. N.
    Schefer, Rahel S. Jud
    Bircher, Barbara
    Bleul, Uli
    Bettschart-Wolfensberger, Regula
    VETERINARY ANAESTHESIA AND ANALGESIA, 2015, 42 (06) : 614 - 622