Effectiveness of four electronic apex locators to determine distance from the apical foramen

被引:42
|
作者
Stoll, R. [1 ]
Urban-Klein, B. [1 ]
Roggendorf, M. J. [1 ]
Jablonski-Momeni, A. [2 ]
Strauch, K. [3 ]
Frankenberger, R. [1 ]
机构
[1] Univ Marburg, Dept Operat Dent & Endodont, Sch Dent, D-35033 Marburg, Germany
[2] Univ Marburg, Dept Pediat Dent, Sch Dent, D-35033 Marburg, Germany
[3] Univ Marburg, Inst Med Biometry & Epidemiol, D-35033 Marburg, Germany
关键词
apical region; electronic apex locator; meter reading display; root canal length determination; IN-VITRO EVALUATION; WORKING LENGTH DETERMINATION; ENLARGED ROOT CANALS; VIVO ACCURACY; ZX; CONSTRICTION; SIZE;
D O I
10.1111/j.1365-2591.2010.01765.x
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
P>Aim To evaluate the accuracy of four electronic apex locators (EAL) in the apical region (0-3 mm short of the foramen) and to compare the precision of the readings on the display with the real position of the file in the root canal. Methodology Twenty single-rooted extracted teeth with round root canals were used. The canal orifices were preflared, and the length to the major foramen was determined visually under a microscope. Canals were enlarged, so that a size 15 file fitted well inside the canal. Teeth were mounted in acrylic test tubes filled with physiologic saline solution. Electronic length was determined in the region between the major foramen and 3 mm short of it in 0.5 mm steps with the Dentaport ZX, Root ZX mini, Elements Diagnostic Unit and Apex Locator and Raypex 5 using files of size 10 and size 15. The data were analysed using linear regression between true length and EAL reading, Bland-Altman plots and nonparametric tests at a significance level of alpha = 0.05. Results The major foramen was detected by all EALs. With a measurement file positioned directly at the major foramen, meter readings were equivalent to a position 0.01-0.38 mm away. For the Dentaport ZX, a better accuracy using the size 15 file for the area 0-1.5 mm short of the apex was found. The differences in measurements between the two files were smaller for the other EALs. In linear regression, a good linearity for Dentaport ZX and Root ZX mini and moderate linearity for Elements Diagnostic Unit and Apex Locator and Raypex 5 were found. The slope of the measurement curve was too low (0.37-0.57) for the Raypex 5 and almost optimal for the Dentaport ZX (1.01-1.05). The Root ZX mini and the Elements Obturation Unit produced lower slope values and especially the Elements Obturation Unit revealed much higher SDs at the different measurement levels. Conclusion Amongst the four EALs, the Dentaport ZX and Root ZX mini had the best agreement between true lengths and meter readings. For the Raypex 5, an interpretation of the colour-coded zones as distance to the foramen cannot be recommended and might lead to erroneous interpretations.
引用
收藏
页码:808 / 817
页数:10
相关论文
共 36 条
  • [1] Influence of Different Apical Foramen Morphologies on the Accuracy of Four Electronic Foramen Locators
    Furlan, Renan D.
    Alcalde, Murilo P.
    Vivan, Rodrigo R.
    Klymus, Michel E.
    Limoeiro, Ana G. S.
    Duarte, Marco A. H.
    de Vasconcelos, Bruno C.
    EUROPEAN JOURNAL OF DENTISTRY, 2024,
  • [2] In vitro comparison of four different electronic apex locators to determine the major foramen using the clearing technique
    Altunbas, D.
    Kustarci, A.
    Arslan, D.
    Er, K.
    NIGERIAN JOURNAL OF CLINICAL PRACTICE, 2014, 17 (06) : 706 - 710
  • [3] Distance from file tip to the major apical foramen in relation to the numeric meter reading on the display of three different electronic apex locators
    Higa, R. A.
    Adorno, C. G.
    Ebrahim, A. K.
    Suda, H.
    INTERNATIONAL ENDODONTIC JOURNAL, 2009, 42 (12) : 1065 - 1070
  • [4] The ability of two apex locators to locate the apical foramen: An in vitro study
    Cunha D'Assuncao, Fabio Luiz
    de Albuquerque, Diana Santana
    de Queiroz Ferreira, Linalda Correia
    JOURNAL OF ENDODONTICS, 2006, 32 (06) : 560 - 562
  • [5] Ex vivo evaluation of the ability of four different electronic apex locators to determine the working length in teeth with various foramen diameters
    Ebrahim, A. K.
    Wadachi, R.
    Suda, H.
    AUSTRALIAN DENTAL JOURNAL, 2006, 51 (03) : 258 - 262
  • [6] Accuracy of four electronic apex locators:: an in vitro evaluation
    De Moor, RJG
    Hommez, GMG
    Martens, LC
    De Boever, JG
    ENDODONTICS & DENTAL TRAUMATOLOGY, 1999, 15 (02): : 77 - 82
  • [7] An ex vivo comparison of root canal length determination by three electronic apex locators at positions short of the apical foramen
    de Vasconcelos, Bruno Carvalho
    do Vale, Taina Macedo
    Teixeira de Menezes, Antonio Sergio
    Pinheiro-Junior, Elilton Cavalcante
    Vivacqua-Gomes, Nilton
    Bernardes, Ricardo Affonso
    Hungaro Duarte, Marco Antonio
    ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 2010, 110 (02): : E57 - E61
  • [8] Impact of Apical Patency on Accuracy of Electronic Apex Locators: In Vitro Study
    Abdelsalam, Nelly
    Hashem, Nasr
    JOURNAL OF ENDODONTICS, 2020, 46 (04) : 509 - 514
  • [9] Accuracy of three electronic apex locators in determining the apical foramen in multi-rooted teeth: Randomised clinical and laboratory study
    Milanovic, Ivana
    Ivanovic, Vladimir
    Vujaskovic, Mirjana
    Ignjatovic, Snezana
    Miletic, Vesna
    AUSTRALIAN ENDODONTIC JOURNAL, 2015, 41 (01) : 35 - 43
  • [10] Ex Vivo Evaluation of the Accuracy of Electronic Foramen Locators in Root Canals with an Obstructed Apical Foramen
    de Vasconcelos, Bruno Carvalho
    Verissimo Chaves, Rebeca Dibe
    Vivacqua-Gomes, Nilton
    de Miranda Candeiro, George Taccio
    Bernardes, Ricardo Affonso
    Vivan, Rodrigi Ricci
    Hungaro Duarte, Marco Antonio
    JOURNAL OF ENDODONTICS, 2015, 41 (09) : 1551 - 1554