Assessment of a Novel 3T MRI Compatible Cochlear Implant Magnet: Torque, Forces, Demagnetization, and Imaging

被引:16
|
作者
Tysome, James Russell [1 ]
Tam, Yu Chuen [2 ]
Patterson, Ilse [3 ]
Graves, Martin John [1 ]
Gazibegovic, Dzemal [4 ]
机构
[1] Cambridge Univ Hosp, Cambridge Biomed Campus,Hills Rd, Cambridge, England
[2] Emmeline Ctr, Hills Rd, Cambridge, England
[3] Radiol Dept, Hills Rd, Cambridge, England
[4] AB GmbH, ERC, Hannover, Germany
关键词
Cochlear implant; Imaging; Magnet; Magnetic resonance imaging; Torque; Ultra; 3D; AUDITORY BRAIN-STEM; SAFETY;
D O I
10.1097/MAO.0000000000002407
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
Hypothesis: A novel cochlea implant (CI) device magnet providing alignment to the static field of a magnetic resonance imaging (MRI) will lead to reduced torque, longitudinal forces, and demagnetization effects. The image void and distortion will be comparable to those obtained with standard CI magnets. Background: MRI investigations of CI users pose several challenges such as magnet dislocation, demagnetization, and may cause pain. The presence of a CI magnet within MRI field causes image distortions and may diminish the diagnostic value of an MRI procedure. Objective of this work is to evaluate magnetic forces and imaging properties of the novel CI magnet within 1.5 and 3T MRI. Methods: Forces and torque of the novel CI magnet were measured in both 1.5 and 3T MRI and compared with the standard magnet in 1.5T. One cadaver head was implanted with the CI devices containing the novel and standard magnets in different configurations reflecting clinical scenarios and imaging properties were assessed and compared. Results: In particular the torque has been reduced with the novel CI magnet in comparison to the standard one. Both CI magnets have not shown any signs of demagnetization. The image void and distortion was comparable between the two magnets for the main MRI clinical scanning protocols in 1.5T MRI. Conclusions: The novel CI magnet is safe to use for MRI investigations of CI users in 3T MRI without a need for bandaging and has acceptable level of image artefacts.
引用
收藏
页码:E966 / E974
页数:9
相关论文
共 50 条
  • [1] 3T MR imaging of the cochlear nerve and labyrinth in cochlear implant candidates
    Lane, JI
    Kocharian, A
    Witte, RJ
    Bernstein, MA
    Felmlee, JP
    Lin, C
    RADIOLOGY, 2001, 221 : 571 - 571
  • [2] When an MRI compatible cochlear implant magnet can hide important findings
    Eltawil, Yasmin
    Aaron, Ksenia A.
    Maria, Peter L. Santa
    OTOLARYNGOLOGY CASE REPORTS, 2022, 23
  • [3] When an MRI compatible cochlear implant magnet can hide important findings
    Eltawil, Yasmin
    Aaron, Ksenia A.
    Maria, Peter L. Santa
    OTOLARYNGOLOGY CASE REPORTS, 2022, 23
  • [4] 3T MRI-based estimation of scalar cochlear implant electrode position
    Tek, F.
    Mueller, S.
    Boga, E.
    Gehl, H. B.
    Seitz, D.
    Scholtz, L. U.
    Sudhoff, H.
    Todt, I.
    ACTA OTORHINOLARYNGOLOGICA ITALICA, 2019, 39 (04) : 269 - 273
  • [5] Significant Artifact Reduction at 1.5T and 3T MRI by the Use of a Cochlear Implant with Removable Magnet: An Experimental Human Cadaver Study
    Wagner, Franca
    Wimmer, Wilhelm
    Leidolt, Lars
    Vischer, Mattheus
    Weder, Stefan
    Wiest, Roland
    Mantokoudis, Georgios
    Caversaccio, Marco D.
    PLOS ONE, 2015, 10 (07):
  • [6] Cochlear Implant Magnet Dislocation: Simulations and Measurements of Force and Torque at 1.5T Magnetic Resonance Imaging
    Eerkens, Hedwig J.
    Smits, Cas
    Hofman, Mark B. M.
    EAR AND HEARING, 2021, 42 (05): : 1276 - 1283
  • [7] MRI-Compatible Microcirculation System Using Ultrasonic Pumps for Microvascular Imaging on 3T MRI
    Jung, Ju-Yeon
    Seo, Dong-Kyu
    Lee, Yeong-Bae
    Kang, Chang-Ki
    SENSORS, 2022, 22 (16)
  • [8] Cochlear implant with a non-removable magnet: preliminary research at 3-T MRI
    F. Dubrulle
    A. Sufana Iancu
    C. Vincent
    G. Tourrel
    O. Ernst
    European Radiology, 2013, 23 : 1510 - 1518
  • [9] Cochlear implant with a non-removable magnet: preliminary research at 3-T MRI
    Dubrulle, F.
    Iancu, A. Sufana
    Vincent, C.
    Tourrel, G.
    Ernst, O.
    EUROPEAN RADIOLOGY, 2013, 23 (06) : 1510 - 1518
  • [10] Magnet dislocation during 3 T magnetic resonance imaging in a pediatric case with cochlear implant
    Ozgur, Abdulkadir
    Dursun, Engin
    Celiker, Fatma Beyazal
    Terzi, Suat
    BRAZILIAN JOURNAL OF OTORHINOLARYNGOLOGY, 2019, 85 (06) : 799 - 802