Comparison of two- and three-dimensional ultrasonography in lung volume measurement of normal fetuses

被引:15
|
作者
Araujo, Edward, Jr. [1 ]
Machado Nardozza, Luciano Marcondes [1 ]
Pires, Claudio Rodrigues [1 ]
Guimaraes Filho, Helio Antonio [1 ]
Moron, Antonio Fernandes [1 ]
机构
[1] Univ Fed Sao Paulo, Dept Obstet, BR-04023062 Sao Paulo, Brazil
关键词
fetus; lung volume measurements; three-dimensional image; ultrasound;
D O I
10.1515/JPM.2007.073
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Aim: The purpose of this study was to compare the two-and three-dimensional methods for measuring fetal lung volume of normal fetuses. Methods: A cross-sectional study was performed with 51 normal pregnant women between 20 and 35 weeks. The ellipsoid formula (X*Y*Z*0.52) was used for volume calculation with the two-dimensional (2D) method. With the VOCAL (TM) (Virtual Organ Computer-aided Analysis) method, a rotation angle of 30 degrees was used. Intraclass correlation coefficient (ICC), paired Student's t-test. and Bland-Altman plots were used to compare the techniques. To calculate the intraobserver variability we used the ICC and compared the means between the two measures using the paired Student's p-test. Results: VOCAL (TM) and 2D methods were highly correlated (ICC = 0.919, and 0.873 for the right and left lungs, respectively), however, there was a disagreement. The fetal lung volume means calculated by the 2D method were always overestimated in relation to the means obtained by the VOCAL (TM), for the right lung (24.02 mL x 19.15 mL; P < 0.001), as well as for the left (16.03 mL x 13.77 mL; P = 0.002). As for the intraobserver variability, a good reproducibility was observed for the volume measurement of the left lung by the 2D technique (mean=0.40 mL.; P=0.57) and by the VOCAL (TM) (mean= -0.22 mL; P=0.63). The 2D method, however, presented low reproducibility for the right lung (mean = 1.73 mL; P = 0.31). Conclusion: The two-dimensional method had low agreement and low reproducibility in relation to the threedimensional method for measurement of fetal lung volume in normal fetuses.
引用
收藏
页码:415 / 421
页数:7
相关论文
共 50 条
  • [31] Comparison of nuchal translucency measurements obtained using Volume NTTM and two- and three-dimensional ultrasound
    Cho, H. Y.
    Kwon, J. -Y.
    Kim, Y. H.
    Lee, K. H.
    Kim, J.
    Kim, S. Y.
    Park, Y. W.
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2012, 39 (02) : 175 - 180
  • [33] Fetal liver volume measurement by three-dimensional ultrasonography: a preliminary study
    Laudy, JAM
    Janssen, MMM
    Struyk, PC
    Stijnen, T
    Wallenburg, HCS
    Wladimiroff, JW
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 1998, 12 (02) : 93 - 96
  • [34] Fetal weight prediction by thigh volume measurement with three-dimensional ultrasonography
    Song, TB
    Moore, TR
    Lee, JY
    Kim, YH
    Kim, EK
    [J]. OBSTETRICS AND GYNECOLOGY, 2000, 96 (02): : 157 - 161
  • [35] Two-dimensional measurement misidentifies alignment outliers in total knee arthroplasty: a comparison of two- and three-dimensional measurements
    Hideki Ueyama
    Yukihide Minoda
    Ryo Sugama
    Yoichi Ohta
    Kazumasa Yamamura
    Suguru Nakamura
    Susumu Takemura
    Hiroaki Nakamura
    [J]. Knee Surgery, Sports Traumatology, Arthroscopy, 2019, 27 : 1497 - 1503
  • [36] Two-dimensional measurement misidentifies alignment outliers in total knee arthroplasty: a comparison of two- and three-dimensional measurements
    Ueyama, Hideki
    Minoda, Yukihide
    Sugama, Ryo
    Ohta, Yoichi
    Yamamura, Kazumasa
    Nakamura, Suguru
    Takemura, Susumu
    Nakamura, Hiroaki
    [J]. KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2019, 27 (05) : 1497 - 1503
  • [37] Comparison of viscoelastic flows in two- and three-dimensional serpentine channels
    Garg, Himani
    Fureby, Christer
    [J]. PHYSICAL REVIEW E, 2024, 109 (05)
  • [38] Comparison between a two- and a three-dimensional arc plasma configuration
    Freton, P
    Gonzalez, JJ
    Gleizes, A
    [J]. JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2000, 33 (19) : 2442 - 2452
  • [39] Comparison of two- and three-dimensional reconstruction methods in optical tomography
    Schweiger, M
    Arridge, SR
    [J]. APPLIED OPTICS, 1998, 37 (31): : 7419 - 7428
  • [40] Measurement of fetal urine production by three-dimensional ultrasonography in normal pregnancy
    Lee, S. M.
    Park, S. K.
    Shim, S. S.
    Jun, J. K.
    Park, J. S.
    Syn, H. C.
    [J]. ULTRASOUND IN OBSTETRICS & GYNECOLOGY, 2007, 30 (03) : 281 - 286