A quantitative metrology for performance characterization of breast tomosynthesis systems based on an anthropomorphic phantom

被引:0
|
作者
Ikejimba, Lynda [1 ]
Chen, Yicheng [2 ]
Oberhofer, Nadia [3 ]
Kiarashi, Nooshin [1 ]
Lo, Joseph Y. [1 ]
Samei, Ehsan [1 ]
机构
[1] Duke Univ, Carl E Ravin Adv Imaging Labs, Med Phys Grad Program, Dept Radiol, Durham, NC 27706 USA
[2] Tsinghua Univ, Dept Engn Phys, Beijing, Peoples R China
[3] Hlth Serv South, Tyrol, Italy
关键词
tomosynthesis; breast phantom; image quality; non-uniformity; IMAGING-SYSTEMS; MAMMOGRAPHY; IMAGES;
D O I
10.1117/12.2082594
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Common methods for assessing image quality of digital breast tomosynthesis (DBT) devices utilize uniform, simplified or otherwise unrealistic phantoms. This study proposes a new imaging performance paradigm using a three-dimensional, anthropomorphic breast phantom and novel quantitative metrology. Images of the phantom were acquired and computationally characterizing in terms of a contrast index, a contrast variability index, a contrast to noise ratio, and a background non-uniformity index. Indices were measured at ROI sizes of 10 mm and 37 mm. The method was applied to two commercial systems, referred to as system X and system Y. All images were acquired at fixed dose of 1.5 mGy AGD. Results showed differences between systems. At all window sizes, system X had a greater contrast index than Y, where median values were 8% and 5%, respectively. Contrast variability tracked with contrast but change slightly with window size. At 10 mm, systems X and Y showed CVIs of 6% and 4%, respectively, which at 37 mm became 4% and 3%. Like the contrast index, CNR was found to be independent of window size. CNR was 0.8 for system X and 0.5 for system Y. Lastly, NI at 10 mm was 7% for system X and 6% for Y. At 37mm these reduced to 3% and 2%, respectively. The metrics showed system-dependencies correlated with tissue contrast and visual appearance of clutter from out-of-plane artifacts. The findings illustrate that the anthropomorphic phantom can be used as an image quality tool with results that are quantitative and are targeted to be reflective of clinical performance of breast tomosynthesis systems.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] A phantom study comparing technical image quality of five breast tomosynthesis systems
    Sundell, Veli-Matti
    Jousi, Mikko
    Hukkinen, Katja
    Blanco, Roberto
    Makela, Teemu
    Kaasalainen, Touko
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2019, 63 : 122 - 130
  • [22] Anthropomorphic Physical Breast Phantom Based on Patient Breast CT Data: Preliminary Results
    Daskalov, Sivo
    Okkalidis, Nikiforos
    Boone, John M.
    Marinov, Stoyko
    Bliznakov, Zhivko
    Mettivier, Giovanni
    Bosmans, Hilde
    Russo, Paolo
    Bliznakova, Kristina
    XV MEDITERRANEAN CONFERENCE ON MEDICAL AND BIOLOGICAL ENGINEERING AND COMPUTING - MEDICON 2019, 2020, 76 : 367 - 374
  • [23] A semi-anthropomorphic breast phantom with tunable blood oxygenation levels for use in quantitative photoacoustics
    Dantuma, Maura
    Julia, Javier Ortega
    Manohar, Srirang
    PROCEEDINGS OF THE 2019 IEEE REGION 10 CONFERENCE (TENCON 2019): TECHNOLOGY, KNOWLEDGE, AND SOCIETY, 2019, : 114 - 117
  • [24] Channelized Hotelling observer for multi-vendor breast tomosynthesis image quality estimation: detection of calcification clusters in an anthropomorphic phantom
    Petrov, Dimitar
    Marshall, Nicholas
    Bosmans, Hilde
    15TH INTERNATIONAL WORKSHOP ON BREAST IMAGING (IWBI2020), 2020, 11513
  • [25] Image simulation and realism evaluation for mammography and tomosynthesis based on the detailed breast phantom
    Wang, Jiahao
    Liu, Yeqi
    Hu, Ankang
    Wei, Shuoyang
    Wu, Tao
    Li, Junli
    Qiu, Rui
    RADIATION PHYSICS AND CHEMISTRY, 2023, 213
  • [26] Quantitative Breast Tomosynthesis: Development of An Estimation Performance Metric and Optimization Framework
    Richard, S.
    Samei, E.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [27] Development of a Dynamic 4D Anthropomorphic Breast Phantom for Contrast-based Breast Imaging
    Kiarashi, Nooshin
    Lin, Yuan
    Segars, Willliam P.
    Ghate, Sujata V.
    Ikejimba, Lynda
    Chen, Baiyu
    Lo, Joseph Y.
    Dobbins, James T., III
    Nolte, Loren W.
    Samei, Ehsan
    MEDICAL IMAGING 2012: PHYSICS OF MEDICAL IMAGING, 2012, 8313
  • [28] Objective assessment of task performance: a comparison of two FFDM detectors using an anthropomorphic breast phantom
    Makeev, Andrey
    Ikejimba, Lynda C.
    Salad, Jesse
    Glick, Stephen J.
    JOURNAL OF MEDICAL IMAGING, 2019, 6 (04)
  • [29] Design and application of a structured phantom for detection performance comparison between breast tomosynthesis and digital mammography
    Cockmartin, L.
    Marshall, N. W.
    Zhang, G.
    Lemmens, K.
    Shaheen, E.
    Van Ongeval, C.
    Fredenberg, E.
    Dance, D. R.
    Salvagnini, E.
    Michielsen, K.
    Bosmans, H.
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (03): : 758 - 780
  • [30] A Comparison of Methods to Evaluate Gray Scale Response of Tomosynthesis Systems using a Software Breast Phantom
    Sousa, Maria A. Z.
    Bakic, Predrag R.
    Schiabel, Homero
    Maidment, Andrew D. A.
    MEDICAL IMAGING 2017: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2017, 10136