Development of a dedicated phantom for multi-target single-isocentre stereotactic radiosurgery end to end testing

被引:22
|
作者
Poder, Joel [1 ]
Brown, Ryan [1 ]
Porter, Harry [1 ]
Gupta, Rashmi [1 ]
Ralston, Anna [1 ]
机构
[1] St George Hosp, Canc Care Ctr, Kogarah, NSW, Australia
来源
关键词
dosimetric validation; end-to-end testing; MRI deformation; multi-target; single isocentre; stereotactic radiosurgery; RADIOCHROMIC FILM DOSIMETRY; BRAIN RADIATION-THERAPY; MODULATED ARC THERAPY; IMAGE REGISTRATION; LINEAR-ACCELERATOR; TARGET DELINEATION; CT SCANS; SRS QA; RADIOTHERAPY; METASTASES;
D O I
10.1002/acm2.12452
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose The aim of this project was to design and manufacture a cost-effective end-to-end (E2E) phantom for quantifying the geometric and dosimetric accuracy of a linear accelerator based, multi-target single-isocenter (MTSI) frameless stereotactic radiosurgery (SRS) technique. Method A perspex Multi-Plug device from a Sun Nuclear ArcCheck phantom (Sun Nuclear, Melbourne, FL) was enhanced to make it more applicable for MTSI SRS E2E testing. The following steps in the SRS chain were then analysed using the phantom: magnetic resonance imaging (MRI) distortion, planning computed tomography (CT) scan and MRI image registration accuracy, phantom setup accuracy using CBCT, dosimetric accuracy using ion chamber, planar film dose measurements and coincidence of linear accelerator mega-voltage (MV), and kilo-voltage (kV) isocenters using Winston-Lutz testing (WLT). Results The dedicated E2E phantom was able to successfully quantify the geometric and dosimetric accuracy of the MTSI SRS technique. MRI distortions were less than 0.5 mm, or half a voxel size. The average MRI-CT registration accuracy was 0.15 mm (+/- 0.31 mm), 0.20 mm (+/- 0.16 mm), and 0.39 mm (+/- 0.11 mm) in the superior/inferior, left/right and, anterior/posterior directions, respectively. The phantom setup accuracy using CBCT was better than 0.2 mm and 0.1 degrees. Point dose measurements were within 5% of the treatment planning system predicted dose. The comparison of planar film doses to the planning system dose distributions, performed using gamma analysis, resulted in pass rates greater than 97% for 3%/1 mm gamma criteria. Finally, off-axis WLT showed MV/kV coincidence to be within 1 mm for off-axis distances up to 60 mm. Conclusion A novel, versatile and cost-effective phantom for comprehensive E2E testing of MTSI SRS treatments was developed, incorporating multiple detector types and fiducial markers. The phantom is capable of quantifying the accuracy of each step in the MTSI SRS planning and treatment process.
引用
收藏
页码:99 / 108
页数:10
相关论文
共 50 条
  • [1] End-to-end testing for stereotactic radiotherapy including the development of a MULTI-MODALITY phantom
    Shariff, Maya
    Grigo, Johanna
    Masitho, Siti
    Brandt, Tobias
    Weiss, Alexander
    Lambrecht, Ulrike
    Stillkrieg, Willi
    Lotter, Michael
    Putz, Florian
    Fietkau, Rainer
    Bert, Christoph
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2024, 34 (03): : 477 - 484
  • [2] An open-source, customizable Winston-Lutz-style quality assurance system for multi-target, single isocentre, stereotactic radiosurgery
    Wood, Tania
    Oliver, Pak
    Baldwin, Ln
    [J]. MEDICAL PHYSICS, 2022, 49 (08) : 5693 - 5693
  • [3] Development and Validation of End-to-End Test for Stereotactic Radiosurgery
    Jadhav, A. R., Sr.
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2019, 105 (01): : E692 - E692
  • [4] A Multi-Clinic Validation of An End-To-End Procedure for MLC-Based Stereotactic Radiosurgery with a Novel Phantom
    Brown, T.
    Fagerstrom, J.
    Beck, C.
    Holloway, C.
    Kerns, J.
    Kaurin, D.
    Kielar, K.
    [J]. MEDICAL PHYSICS, 2021, 48 (06)
  • [5] Evaluation of PTV margins and plan robustness for single isocentre multiple target stereotactic radiosurgery
    Lam, Cheuk Him Michael
    Bernstein, David
    Wells, Emma
    [J]. PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2023, 114
  • [6] End-to-End tests in linac based Stereotactic Radiosurgery and Radiotherapy using a new phantom
    Scheib, S.
    Schmelzer, P.
    Krayenbuehl, J.
    Lang, S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S112 - S113
  • [7] Gamma Knife Perfexion End-To-End Testing with a Commercial Stereotactic Phantom
    Yount, C.
    Lyons, K.
    Rahnema, S.
    McDonald, D.
    Peng, J.
    Vanek, K.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3697 - 3697
  • [8] Robustness of single-isocenter multiple-metastasis stereotactic radiosurgery end-to-end testing across institutions
    Saenz, Daniel
    Papanikolaou, Niko
    Zoros, Emmanouil
    Pappas, Evangelos
    Reiner, Michael
    Chew, Lip Teck
    Lim, Hooi Yin
    Hancock, Sam
    Nebelsky, Alex
    Njeh, Christopher
    Anagnostopoulos, Georgios
    [J]. JOURNAL OF RADIOSURGERY AND SBRT, 2020, 7 (03): : 223 - 232
  • [9] A QA device to perform end-to-end spatial accuracy tests of target irradiation in stereotactic radiosurgery
    Kanellitsas, C.
    Gupta, N.
    [J]. MEDICAL PHYSICS, 2006, 33 (06) : 2060 - 2060
  • [10] Proton Radiosurgery End-To-End Testing Using Lucy 3D QA Phantom
    Choi, D.
    Gordon, I.
    Ghebremedhin, A.
    Wroe, A.
    Schulte, R.
    Bush, D.
    Slater, J.
    Patyal, B.
    [J]. MEDICAL PHYSICS, 2014, 41 (06) : 285 - 285