Innovative processes for electropolishing of medical devices made of stainless steels

被引:32
|
作者
Eliaz, Noam [1 ]
Nissan, Oded [1 ]
机构
[1] Tel Aviv Univ, Sch Mech Engn, Biomat & Corros Lab, IL-69978 Tel Aviv, Israel
关键词
electropolishing; stainless steel; atomic force microscope; surface roughness; cell adhesion;
D O I
10.1002/jbm.a.31429
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Currently, many medical devices are made of implantable metals such as 316LVM stainless steel. Electropolishing is a common process to obtain a smooth surface, free of contaminants and more passive, which allows for minimizing the foreign body response and cell adhesion. However, polishing of small implants with a highly complicated geometry and nonuniform metallurgy might result in unsatisfactory results. The objective of this work was to develop an electropolishing process effective for complex metallic implants such as artificial heart valve frames and miniature glaucoma implants. Polishing in an ultrasonic bath and pulsed voltage polishing processes were studied and compared to the standard ASTM process. Currentvoltage curves were constructed for different solutions and bath temperatures. The polished parts were evaluated by stereomicroscopy, optical microscopy, atomic force microscopy, noncontact surface profilometry, and X-ray diffraction. Pulse polishing was found useful in eliminating the erosion effects of gas bubbles in solution. Electropolishing in an ultrasonic bath was found useful when a rough, patterned surface is needed, e.g. for osseointegration purposes. Preliminary animal studies followed by histopathology indicated that the polished surfaces stimulated only a moderate body reaction, as desired in such applications. The pronounced dependence of the measured roughness values on both the measurement technique and scanned area should inspire the preparation of a new test-method standard. (c) 2007 Wiley Periodicals, Inc.
引用
收藏
页码:546 / 557
页数:12
相关论文
共 50 条
  • [31] DISCUSSION OF PROCESSES FOR THE PRODUCTION OF STAINLESS AND ACID-RESISTING STEELS
    BURGHARDT, H
    PIETSCHMANN, G
    KOCH, F
    NEUE HUTTE, 1979, 24 (06): : 214 - 216
  • [32] Micromagnetic investigation of the fatigue processes of superduplex and austenitic stainless steels
    Mészáros, I
    Dobránszky, J
    Nagy, G
    ELECTROMAGNETIC NONDESTRUCTIVE EVALUATION (V), 2001, 21 : 267 - 274
  • [33] Study of restoration processes and texture evolution in ferritic stainless steels
    Rege, JS
    Garcia, CI
    DeArdo, AJ
    37TH MECHANICAL WORKING AND STEEL PROCESSING CONFERENCE PROCEEDINGS, 1996, 33 : 871 - 881
  • [34] Processes of Physical Treatment of Stainless Steels Obtained by Additive Manufacturing
    Babaev, Artem
    Promakhov, Vladimir
    Schulz, Nikita
    Semenov, Artem
    Bakhmat, Vladislav
    Vorozhtsov, Alexander
    METALS, 2022, 12 (09)
  • [35] Materials and Processes for Medical Devices - Abstracts
    不详
    ADVANCED MATERIALS & PROCESSES, 2005, 163 (07): : 68 - +
  • [36] Nickel-free austenitic stainless steels for medical applications
    Yang, Ke
    Ren, Yibin
    SCIENCE AND TECHNOLOGY OF ADVANCED MATERIALS, 2010, 11 (01)
  • [37] Plasma immersion ion implantation: Innovative technology for surface treatment of stainless steels
    Blawert, C
    SURFACE ENGINEERING, 1999, 15 (02) : 154 - 158
  • [38] Electrochemical studies of welded joints corrosion resistance made from stainless steels
    Kovalev, Mark
    Shakhmatov, Aleksey
    Alhimenko, Aleksey
    MATERIALS TODAY-PROCEEDINGS, 2020, 30 : 501 - 505
  • [40] Stainless steel tubular parts designed for medical devices
    不详
    ADVANCED MATERIALS & PROCESSES, 2006, 164 (01): : 81 - 81