Electron dose distributions caused by the contact-type metallic eye shield: Studies using Monte Carlo and pencil beam algorithms

被引:2
|
作者
Kang, Sei-Kwon [1 ]
Yoon, Jai-Woong [1 ]
Hwang, Taejin [1 ]
Park, Soah [1 ]
Cheong, Kwang-Ho [1 ]
Han, Tae Jin [1 ]
Kim, Haeyoung [1 ]
Lee, Me-Yeon [1 ]
Kim, Kyoung Ju [1 ]
Bae, Hoonsik [1 ]
机构
[1] Hallym Univ, Coll Med, Dept Radiat Oncol, Seoul, South Korea
基金
新加坡国家研究基金会;
关键词
Eye shield; Electron treatment; Monte Carlo; Pinnacle; RADIOTHERAPY; SIMULATION; LYMPHOMA;
D O I
10.1016/j.meddos.2015.01.006
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
A metallic contact eye shield has sometimes been used for eyelid treatment, but dose distribution has never been reported for a patient case. This study aimed to show the shield-incorporated CT-based dose distribution using the Pinnacle system and Monte Carlo (MC) calculation for 3 patient cases. For the artifact-free CT scan, an acrylic shield machined as the same size as that of the tungsten shield was used. For the MC calculation, BEAMnrc and DOSXYZnrc were used for the 6-MeV electron beam of the Varian 21 EX, in which information for the tungsten, stainless steel, and aluminum material for the eye shield was used. The same plan was generated on the Pinnacle system and both were compared. The use of the acrylic shield produced clear CT images, enabling delineation of the regions of interest, and yielded CT-based dose calculation for the metallic shield. Both the MC and the Pinnacle systems showed a similar dose distribution downstream of the eye shield, reflecting the blocking effect of the metallic eye shield. The major difference between the MC and the Pinnacle results was the target eyelid dose upstream of the shield such that the Pinnacle system underestimated the dose by 19 to 28% and 11 to 18% for the maximum and the mean doses, respectively. The pattern of dose difference between the MC and the Pinnacle systems was similar to that in the previous phantom study. In conclusion, the metallic eye shield was successfully incorporated into the CT-based planning, and the accurate dose calculation requires MC simulation. (C) 2015 American Association of Medical Dosimetrists.
引用
收藏
页码:240 / 243
页数:4
相关论文
共 50 条
  • [21] 3D-Monte Carlo simulations of clinical electron beam dose distributions using CT-numbers
    Schneider, W
    Bortfeld, T
    Schlegel, W
    [J]. PROCEEDINGS OF THE XIITH INTERNATIONAL CONFERENCE ON THE USE OF COMPUTERS IN RADIATION THERAPY, 1997, : 163 - 165
  • [22] Comparison of dose calculations between pencil-beam and Monte Carlo algorithms in arc therapy with 3DVH software
    Son, S. H.
    Song, J. H.
    Shin, H. J.
    Kay, C. S.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S178 - S178
  • [23] Influence of initial electron beam characteristics on Monte Carlo calculated absorbed dose distributions for linear accelerator electron beams
    Björk, P
    Knöös, T
    Nilsson, P
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (22): : 4019 - 4041
  • [24] Denoising of electron beam Monte Carlo dose distributions using digital filtering techniques (vol 45, pg 1765, 2000)
    Deasy, JO
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2000, 45 (08): : 2441 - 2441
  • [25] THE APPLICATION OF CORRELATED SAMPLING TO THE COMPUTATION OF ELECTRON-BEAM DOSE DISTRIBUTIONS IN HETEROGENEOUS PHANTOMS USING THE MONTE-CARLO METHOD
    HOLMES, MA
    MACKIE, TR
    SOHN, W
    RECKWERDT, PJ
    KINSELLA, TJ
    BIELAJEW, AF
    ROGERS, DWO
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 1993, 38 (06): : 675 - 688
  • [26] A Comparative Dosimetric Study of Spine SABR Using Two MLC Delivery Systems with Pencil Beam and Monte Carlo Algorithms
    Rahimian, J.
    Zhang, J.
    Vinci, J.
    [J]. MEDICAL PHYSICS, 2022, 49 (06) : E965 - E965
  • [27] A Golden Beam Data Commissioning Framework of Monte Carlo Dose Calculation Algorithms of Two Pencil Beam Scanning Proton Therapy Treatment Planning Systems
    Chang, C.
    Harms, J.
    Zhang, R.
    Zhou, J.
    Lin, Y.
    Slopsema, R.
    Dhabaan, A.
    Liu, T.
    McDonald, M.
    Langen, K.
    Lin, L.
    [J]. MEDICAL PHYSICS, 2019, 46 (06) : E167 - E167
  • [28] Influence of initial electron beam parameters on Monte Carlo calculated absorbed dose distributions for radiotherapy photon beams
    Tzedakis, A
    Damilakis, JE
    Mazonakis, M
    Stratakis, J
    Varveris, H
    Gourtsoyiannis, N
    [J]. MEDICAL PHYSICS, 2004, 31 (04) : 907 - 913
  • [29] The Effect of Pencil Beam, Collapsed Cone, and Monte Carlo Algorithms on Dose-Volume Parameters in Esophagus Cancer: The Digital Phantom Study
    Bukulmez, Taylan
    Ozdemir, Beyza Sirin
    [J]. TURK ONKOLOJI DERGISI-TURKISH JOURNAL OF ONCOLOGY, 2021, 36 (02): : 191 - 199
  • [30] Monte Carlo Versus Pencil Beam Dose Calculation for Spine SBRT Treatments Using HybridARC and Sliding Windows IMRT
    Venencia, C.
    Pino, M.
    Caussa, L.
    Molineu, A.
    Garrigo, E.
    [J]. MEDICAL PHYSICS, 2016, 43 (06) : 3549 - +