A primary field compensation scheme for planar array magnetic induction tomography

被引:45
|
作者
Watson, S [1 ]
Morris, A
Williams, RJ
Griffiths, H
Gough, W
机构
[1] Univ Glamorgan, Sch Elect, Pontypridd CF37 1DL, M Glam, Wales
[2] Singleton Hosp, Dept Med Phys & Clin Engn, Swansea SA2 8QA, W Glam, Wales
[3] Cardiff Univ, Dept Phys & Astron, Cardiff CF2 3YB, S Glam, Wales
关键词
magnetic induction tomography; electrical impedance tomography;
D O I
10.1088/0967-3334/25/1/031
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
In biomedical magnetic induction tomography (MIT), measurement precision may be improved by incorporating some form of primary field compensation/cancellation scheme. Schemes which have been described previously include gradiometric approaches and the use of 'back-off' coils. In each of these methods, however, the primary field cancellation was achieved only for a single transmitter/receiver combination. For the purpose of imaging, it would be desirable for a fully electronically scanned MIT system to provide a complete set of measurements, all with the primary field cancelled. A single channel suitable for incorporation into an MIT system with planar-array geometry is described. The transmitter is a 6-turn coil of wire 5 cm in diameter. The receiver is a surface mount inductor, of inductance 10 muH, mounted such that, in principle, no net primary field flux threads it. The results of measurements carried out with the single channel system suggest that the signal due to the primary excitation field can be reduced on average by a factor of 298 by the sensor geometry over the operating frequency range 1-10 MHz. The standard deviation and drift of the signal with the system adjusted for maximum primary field cancellation, expressed as a percentage of the signal when the receiver coil was rotated until its axis of sensitivity lay along the primary field, were 0.0009% and 0.009%, respectively. The filter time constant used was 30 ms.
引用
收藏
页码:271 / 279
页数:9
相关论文
共 50 条
  • [21] Magnetic induction tomography
    Korzhenevskii, AV
    Cherepenin, VA
    RADIOTEKHNIKA I ELEKTRONIKA, 1997, 42 (04): : 506 - 512
  • [22] A multifrequency magnetic induction tomography system using planar gradiometers: data collection and calibration
    Rosell-Ferrer, J.
    Merwa, R.
    Brunner, P.
    Scharfetter, H.
    PHYSIOLOGICAL MEASUREMENT, 2006, 27 (05) : S271 - S280
  • [23] Planar Array of Electrical Capacitance Tomography With Rotation
    Zhang, Maomao
    Liu, Yijun
    Soleimani, Manuchehr
    IEEE SENSORS JOURNAL, 2023, 23 (12) : 13548 - 13554
  • [24] Mutual coupling compensation in small planar array antennas
    Darwood, P
    Fletcher, PN
    Hilton, GS
    IEE PROCEEDINGS-MICROWAVES ANTENNAS AND PROPAGATION, 1998, 145 (01) : 1 - 6
  • [25] COMPENSATION OF SPHERICAL REFLECTOR ABERRATIONS BY PLANAR ARRAY FEEDS
    AMITAY, N
    ZUCKER, H
    IEEE TRANSACTIONS ON ANTENNAS AND PROPAGATION, 1972, AP20 (01) : 49 - &
  • [26] Mutual coupling compensation in small planar array antennas
    Univ of Bristol, Bristol, United Kingdom
    IEE Proc Microwaves Antennas Propag, 1 (1-6):
  • [27] The Phase Compensation Scheme for Multiple Planar Antenna on the SOTM
    Wang, Qiong
    Yao, Minli
    Lin, Zhiqiang
    Xu, Huachun
    ICIEA: 2009 4TH IEEE CONFERENCE ON INDUSTRIAL ELECTRONICS AND APPLICATIONS, VOLS 1-6, 2009, : 1806 - 1809
  • [28] Sector sensor array technique for high conductivity materials imaging in magnetic induction tomography
    Chen, Jia
    Ke, Li
    Du, Qiang
    Zu, Wanni
    Ding, Xiaodi
    BIOMEDICAL ENGINEERING ONLINE, 2019, 18 (01)
  • [29] Sector sensor array technique for high conductivity materials imaging in magnetic induction tomography
    Jia Chen
    Li Ke
    Qiang Du
    Wanni Zu
    Xiaodi Ding
    BioMedical Engineering OnLine, 18
  • [30] Magnetic Field Analysis of Permanent Magnet Array for Planar Motor Based on Equivalent Magnetic Charge Method
    Rui, Huang
    Jian, Feng
    PROCEEDINGS OF THE 10TH WORLD CONGRESS ON INTELLIGENT CONTROL AND AUTOMATION (WCICA 2012), 2012, : 3966 - 3970