A fast nonlinear method for parametric imaging of myocardial perfusion by dynamic 13N-ammonia PET

被引:0
|
作者
Golish, SR
Hove, JD
Schelbert, HR
Gambhir, SS
机构
[1] Univ Calif Los Angeles, Crump Inst Mol Imaging, Sch Med, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Dept Comp Sci, Los Angeles, CA 90095 USA
[3] Univ Calif Los Angeles, Dept Mol & Med Pharmacol, Los Angeles, CA 90095 USA
[4] Univ Calif Los Angeles, Dept Biomath, Los Angeles, CA 90095 USA
[5] Rigshosp, DK-2100 Copenhagen, Denmark
关键词
parametric imaging; N-13-ammonia PET; sigmoidal networks; artificial neural networks; function estimation; nonlinear regression;
D O I
暂无
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A parametric image of myocardial perfusion (mL/min/g) is a quantitative image generated by fitting a tracer kinetic model to dynamic(13)N-ammmonia PET data on a pixel-by-pixel basis. There are several methods for such parameter estimation problems, including weighted nonlinear regression (WNLR) and a fast linearizing method known as Patlak analysis. Previous work showed that sigmoidal networks can be used for parameter estimation of mono- and biexponential models. The method used in this study is a hybrid of WNLR and sigmoidal networks called nonlinear regression estimation (NRE). The purpose of the study is to compare NRE with WNLR and Patlak analysis for parametric imaging of perfusion in the canine heart by N-13-ammonia PET. Methods: A simulation study measured the statistical performance of NRE, WNLR, and Patlak analysis for a probabilistic model of time-activity curves. Four canine subjects were injected with 740 MBq N-13-ammonia and scanned dynamically. Images were reconstructed with filtered backprojection and resliced into short-axis cuts. Parametric images of a single midventricular plane per subject were generated by NRE, WNLR, and Patlak analysis. Small regions of interest (ROIs) were drawn on each parametric image (8 ROls per subject for a total of 32). Results: For the simulation study, the median absolute value of the relative error for a perfusion value of 1.0 mL/min/g was 16.6% for NRE, 17.9% for WNLR, 19.5% for Patlak analysis, and 14.5% for an optimal WNLR method (computable by simulation only). All methods are unbiased conditioned on a wide range of perfusion values. For the canine studies, the least squares line fits comparing NRE (y) and Patlak analysis (z) with WNLR (x) for all 32 ROls were y = 1.02x - 0.028 and z = 0.90x + 0.019, respectively. Both NRE and Patlak analysis generate 128 x 128 parametric images in seconds. Conclusion: The statistical performance of NRE is competitive with WNLR and superior to Patlak analysis for parametric imaging of myocardial perfusion. NRE is a fast nonlinear alternative to Patlak analysis and other fast linearizing methods for parametric imaging. NRE should be applicable to many other tracers and tracer kinetic models.
引用
收藏
页码:924 / 931
页数:8
相关论文
共 50 条
  • [21] Evaluation of perfusion and image quality changes by reconstruction methods in 13N-ammonia myocardial perfusion PET/CT
    Do, Yong Ho
    Lee, Hong Jae
    Kim, Jin Eui
    Im, Hyung Jun
    Lee, Jae Sung
    Lee, Dong Soo
    [J]. JOURNAL OF NUCLEAR MEDICINE, 2014, 55
  • [22] Nonlinear function estimators allow real time parametric imaging of canine myocardial perfusion from dynamic N-13 ammonia PET data.
    Golish, SR
    Hove, J
    Schelbert, HR
    Phelps, ME
    Gambhir, SS
    [J]. JOURNAL OF NUCLEAR MEDICINE, 1999, 40 (05) : 73P - 73P
  • [23] Utility of Dynamic Perfusion PET Using 13N-Ammonia in Diagnosis of Asymptomatic Recurrence of Fibrosarcoma
    Harisankar, Chidambaram Natrajan Balasubramanian
    Mittal, Bhagwant Rai
    Watts, Ankit
    Bhattacharya, Anish
    Sen, Ramesh
    [J]. CLINICAL NUCLEAR MEDICINE, 2011, 36 (02) : 150 - 151
  • [24] Residual Activity Correction in Quantitative Myocardial Perfusion 13N-Ammonia PET Imaging: A Study in Post-MI Patients
    Markousis-Mavrogenis, George
    Juarez-Orozco, Luis Eduardo
    Alexanderson, Erick
    [J]. HELLENIC JOURNAL OF CARDIOLOGY, 2017, 58 (03) : 245 - 249
  • [25] 13N-ammonia myocardial perfusion imaging with a PET/CT scanner: impact on clinical decision making and cost-effectiveness
    Patrick T. Siegrist
    Lars Husmann
    Martina Knabenhans
    Oliver Gaemperli
    Ines Valenta
    Tobias Hoefflinghaus
    Hans Scheffel
    Paul Stolzmann
    Hatem Alkadhi
    Philipp A. Kaufmann
    [J]. European Journal of Nuclear Medicine and Molecular Imaging, 2008, 35 : 889 - 895
  • [26] Splenic switch off as a novel marker for adenosine response in 13N-ammonia PET myocardial perfusion imaging - a pilot study
    Bakula, A.
    Patriki, D.
    Von Felten, E.
    Benetos, G.
    Sustar, A.
    Benz, D. C.
    Wiedemann-Buser, M.
    Treyer, V
    Pazhenkottil, A. P.
    Graeni, C.
    Gebhard, C.
    Kaufmann, P. A.
    Buechel, R. R.
    Fuchs, T. A.
    [J]. EUROPEAN HEART JOURNAL, 2020, 41 : 289 - 289
  • [27] 13N-ammonia myocardial perfusion imaging with a PET/CT scanner:: impact on clinical decision making and cost-effectiveness
    Siegrist, Patrick T.
    Husmann, Lars
    Knabenhans, Martina
    Gaemperli, Oliver
    Valenta, Ines
    Hoefflinghaus, Tobias
    Scheffel, Hans
    Stolzmann, Paul
    Alkadhi, Hatem
    Kaufmann, Philipp A.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2008, 35 (05) : 889 - 895
  • [28] Nonlinear function estimation and other numerical methods for imaging myocardial perfusion by N-13 ammonia PET
    Golish, SR
    Jove, JD
    Schelbert, HR
    Gambhir, SS
    [J]. COMPUTERS IN CARDIOLOGY 1999, VOL 26, 1999, 26 : 651 - 654
  • [29] Interobserver reproducibility of myocardial perfusion quantification with 13N-ammonia and 82Rb PET studied with Carimas
    Nesterov, S. V.
    Deshayes, E.
    Settimo, L.
    Han, C.
    Ryzhkova, D. V.
    Kostina, I. S.
    Maki, M.
    Prior, J. O.
    Sciagra, R.
    Knuuti, J.
    [J]. EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2013, 40 : S155 - S155
  • [30] Human 13N-ammonia PET studies: the importance of measuring 13N-ammonia metabolites in blood
    Keiding, Susanne
    Sorensen, Michael
    Munk, Ole Lajord
    Bender, Dirk
    [J]. METABOLIC BRAIN DISEASE, 2010, 25 (01) : 49 - 56