A method for the evaluation of the change in volume of retrieved acetabular cups

被引:11
|
作者
Raimondi, MT [1 ]
Sassi, R [1 ]
Pietrabissa, R [1 ]
机构
[1] Politecn Milan, Lab Meccan Strutture Biol, Dipartimento Bioingn, I-20133 Milan, Italy
关键词
hip prosthesis; retrieved acetabular cup; estimation of wear; measurement; accuracy;
D O I
10.1243/0954411001535606
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Studies on retrieved hip prostheses are currently performed in order to assess the wear mechanisms and the overall wear rate of such artificial joints. Several reported studies on the survival of artificial hips have been based on the measurement of the amount of worn material directly on retrieved acetabular cups. The estimation of the change in volume, V, of the cup cavity is particularly difficult in the case of slight wear due to several factors of which the most critical is the lack of information on the unworn geometry of the cup. This paper presents a new measuring technique, which is described in detail and has been applied to estimate the wear of 65 acetabular cups harvested from revised hip arthroplasty. The coordinate data of several points on the articular surface are sampled using a coordinate measuring machine (CMM). The value of V is calculated mathematically from the measurements by using the hypothesis that the actual shape of both slightly worn and highly worn surfaces has a small departure from a truly spherical shape. The wear volume is estimated with reasonable accuracy mainly on those cups showing penetration depths greater than 0.2 mm, corresponding to an amount of wear greater than 100 mm(3) for a 32 mm cup. The uncertainty in the results is estimated for each cup. The repeatability of the technique is studied for a case showing very slight wear. The advantages, disadvantages and limitations of the method are presented and discussed.
引用
收藏
页码:577 / 587
页数:11
相关论文
共 50 条
  • [1] Evaluation of oxidation and fatigue damage of retrieved crossfire polyethylene acetabular cups
    Currier, Barbara H.
    Currier, John H.
    Mayor, Michael B.
    Lyford, Kimberly A.
    Collier, John P.
    Van Citters, Douglas W.
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2007, 89A (09): : 2023 - 2029
  • [2] POLYETHYLENE WEAR FROM RETRIEVED ACETABULAR CUPS
    WEIGHTMAN, B
    SWANSON, SAV
    ISAAC, GH
    WROBLEWSKI, BM
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1991, 73 (05): : 806 - 810
  • [3] Wear mode and wear mechanism of retrieved acetabular cups
    Yamamoto, Kengo
    Imakiire, Atsuhiro
    Masaoka, Toshinori
    Shishido, Takaaki
    Mizoue, Tatsuro
    Clarke, Ian C.
    Shoji, Hiromu
    Kawanabe, Keiichi
    Tamura, Jiro
    INTERNATIONAL ORTHOPAEDICS, 2003, 27 (05) : 286 - 290
  • [4] Wear mode and wear mechanism of retrieved acetabular cups
    Kengo Yamamoto
    Atsuhiro Imakiire
    Toshinori Masaoka
    Takaaki Shishido
    Tatsuro Mizoue
    Ian C. Clarke
    Hiromu Shoji
    Keiichi Kawanabe
    Jiro Tamura
    International Orthopaedics, 2003, 27 : 286 - 290
  • [5] A new optical method for determining deformation and volumetric wear in retrieved polyethylene acetabular cups
    Thomsen, M
    Burandt, C
    Görtz, A
    Dathe, H
    Meesenburg, DK
    Spiering, S
    Görtz, W
    Näger, H
    BIOMEDIZINISCHE TECHNIK, 1999, 44 (09): : 247 - 254
  • [6] Wear effects in retrieved acetabular UHMW-PE cups
    Ambrosio, L
    Carotenuto, G
    Marletta, G
    Nicolais, L
    Scandurra, A
    JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1996, 7 (12) : 723 - 729
  • [7] Wear effects in retrieved acetabular UHMW-PE cups
    Dept. of Mat. and Prod. Engineering, University of Naples Federico II, Inst. of Compos. Mat. Technology CNR, Piazzale Tecchio, 80125 Napoli, Italy
    不详
    不详
    J. MATER. SCI. MATER. MED., 12 (723-729):
  • [8] CMM-based method for assessing the volume change of retrieved polyethylene cups in MoP total hip replacements
    Hua X.
    Li J.
    Hua, Xijin (xijinhua@outlook.com), 1600, Institution of Engineering and Technology, United States (06): : 37 - 42
  • [9] Measurement of extent of bone ongrowth and hydroxyapatite absorption in retrieved acetabular cups
    Chung, Young-Yool
    Kim, Jin
    Lim, Chae Hyun
    Kim, Ki-Soo
    Lee, Yeon Soo
    JOURNAL OF ORTHOPAEDIC SCIENCE, 2008, 13 (03) : 198 - 201
  • [10] Association between dislocation, impingement, and articular geometry in retrieved acetabular polyethylene cups
    Tanino, Hiromasa
    Harman, Melinda K.
    Banks, Scott A.
    Hodge, W. Andrew
    JOURNAL OF ORTHOPAEDIC RESEARCH, 2007, 25 (11) : 1401 - 1407