Anatomical Measurement and Finite Element Study on Screw Channel Parameter in Percutaneous Fixation of Canulated Screw for Symphyseolysis

被引:3
|
作者
Yu, Kehe [1 ,2 ]
Hong, Jianjun [2 ]
Sun, Yuefei [3 ]
Shi, Chengdi [2 ]
Guo, Xiaoshan [2 ]
Zhou, Dongsheng [1 ]
机构
[1] Shandong Univ, Shandong Prov Hosp, Dept Traumat Orthoped, Jinan 250021, Shandong, Peoples R China
[2] Med Univ, Affiliated Hosp Wenzhou 2, Dept Orthoped Surg, Wenzhou 325000, Zhejiang, Peoples R China
[3] Chashan Univ Town, Wenzhou Med Univ, Wenzhou 325000, Zhejiang, Peoples R China
关键词
Symphyseolysis; Percutaneous; Canulated screw; Anatomy; Finite element; PELVIC RING; INTERNAL-FIXATION; FRACTURES; EXPERIENCE; REDUCTION;
D O I
10.1007/s12013-014-0335-4
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
To provide anatomical basement for symphyseolysis treatment with percutaneous fixation of canulated screw, through anatomical measurement on pubic symphysis and the surrounding tissues, and conduct the finite element studies on screw channel parameters. 20 cases of normal pelvic specimens from embalmed adult cadavers were taken to measure the anatomical parameter of bony remark of pubic symphysis and the space between spermatic cord (round ligament of the uterus) and pubic tubercle. Anatomical measurement results showed that the narrowest diameter of the superior ramus of pubis was 9.127 +/- A 1.189 mm, distance between two pubic tubercles was 55.656 +/- A 3.780 mm, thickness of the upper pubic symphysis was 10.510 +/- A 0.814 mm, and distance between upper and lower pubic symphysis was 40.872 +/- A 1.211 mm; the distance between round ligament of the uterus and pubic tubercle was 4.408 +/- A 0.304 mm, and the distance between spermatic cord and pubic tubercle was 5.196 +/- A 0.251 mm. The angle between canulated screw guide pin and horizontal plane was 8.342 +/- A 2.152A degrees, the one between guide pin and coronal plane was 5.236 +/- A 1.612A degrees, and the distance from entry point to the outer edge of pubic tubercle was 10.023 +/- A 1.245 mm, which was measured by Mimics software. Percutaneous surgery at horizontal position was simulated on cadaver. And the screw was correctly placed in postoperative imaging examination. According to the anatomical data and finite element studies of screw channel parameter in percutaneous fixation of canulated screw for symphyseolysis, the method can improve the accuracy of screw placement and reduce complications.
引用
收藏
页码:1243 / 1248
页数:6
相关论文
共 50 条
  • [31] Patient-Specific Finite Element Models of Posterior Pedicle Screw Fixation: Effect of Screw's Size and Geometry
    Sensale, Marco
    Vendeuvre, Tanguy
    Schilling, Christoph
    Grupp, Thomas
    Rochette, Michel
    Dall'Ara, Enrico
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2021, 9
  • [32] Screw Stress Distribution in a Clavicle Fracture with Plate Fixation: A Finite Element Analysis
    Alito, Angelo
    Fenga, Domenico
    Tropeano, Giada
    Milardi, Demetrio
    Leonetti, Danilo
    Migliorato, Alba
    Tisano, Adriana
    D'Andrea, Danilo
    Filardi, Vincenzo
    BIOENGINEERING-BASEL, 2023, 10 (12):
  • [33] Biomechanical Evaluation of Clival Screw Fixation for Occipitocervical Instablity: A Finite Element Analysis
    Lin, Weipeng
    Zheng, Jianying
    Zhang, Meichao
    Xu, Panjie
    Xiao, Hang
    Ji, Wei
    ORTHOPAEDIC SURGERY, 2025, 17 (02) : 583 - 592
  • [34] Finite Element Analysis of Horizontal Screw- Screw Crosslink Used in C1-C2 Pedicle Screw- Rod Fixation
    Ouyang, Beiping
    Zou, Xiaobao
    Luo, Chunshan
    Lu, Tingsheng
    Xia, Hong
    Ma, Xiangyang
    MEDICAL SCIENCE MONITOR, 2021, 27
  • [35] Biomechanical Analysis of Scoliosis Adjusted by Screw Fixation System with Finite Element Analysis
    Somtua, Chompunut
    Aroonjarattham, Panya
    Saengpetch, Nadhaporn
    Rattanapan, Nattawat
    Aroonjarattham, Kitti
    ENGINEERING JOURNAL-THAILAND, 2024, 28 (03): : 25 - 34
  • [36] Role of the compression screw in the dynamic hip-screw system: A finite-element study
    Chang, Chih-Wei
    Chen, Yen-Nien
    Li, Chun-Ting
    Peng, Yao-Te
    Chang, Chih-Han
    MEDICAL ENGINEERING & PHYSICS, 2015, 37 (12) : 1174 - 1179
  • [37] A comparative Study of Novel Extramedullary Fixation and Dynamic Hip Screw in the Fixation of Intertrochanteric Fracture: A Finite-Element Analysis
    Ding, Kai
    Zhu, Yanbin
    Wang, Haicheng
    Li, Yonglong
    Yang, Weijie
    Cheng, Xiaodong
    Zhang, Yingze
    Chen, Wei
    Zhang, Qi
    FRONTIERS IN SURGERY, 2022, 9
  • [38] Experimental measurement and finite element analysis of screw spike fatigue loads
    Dick, Matthew G.
    McConnell, David S.
    Iwand, Hans C.
    PROCEEDINGS OF THE ASME/IEEE JOINT RAIL CONFERENCE AND THE ASME INTERNAL COMBUSTION ENGINE DIVISION SPRING TECHNICAL CONFERENCE - 2007, 2007, : 161 - 166
  • [39] A novel calcium phosphate cement used for enhanced pedicle screw fixation: A biomechanical study with finite element
    Chao-Dong Deng
    Zhi-Qiang Gong
    Shao-Li Wang
    Wei Song
    Yi-Lin Zhang
    Ting-Ting Yan
    Journal of Materials Research, 2022, 37 : 3535 - 3546
  • [40] Biomechanical analysis of a customized lumbar interspinous spacer based on transfacetopedicular screw fixation: A finite element study
    Zhao, Gaiping
    Wang, Xiaoyu
    Liu, Tielong
    Dong, Lianfeng
    MEDICAL ENGINEERING & PHYSICS, 2022, 107