Reproducibility of macular thickness and volume using zeiss optical coherence tomography in patients with diabetic macular edema

被引:111
|
作者
Krzystolik, Magdalena G. [1 ]
Strauber, Samara F. [1 ]
Ai-ello, Lloyd Paul [1 ]
Beek, Roy W. [1 ]
Berger, Brian B. [1 ]
Bressler, Neil M. [1 ]
Browning, David J. [1 ]
Chambers, Robert B. [1 ]
Danis, Ronald P. [1 ]
Davis, Matthew D. [1 ]
Glassman, Adam R. [1 ]
Gonzalez, Victor H. [1 ]
Greenberg, Paul B. [1 ]
Gross, Jeffrey G. [1 ]
Kim, Judy E. [1 ]
Kollman, Craig [1 ]
机构
[1] Jaeb Ctr Hlth Res, Tampa, FL 33647 USA
关键词
D O I
10.1016/j.ophtha.2006.10.055
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Purpose: To evaluate optical coherence tomography (OCT) reproducibility in patients with diabetic macular edema (DME). Design: Prospective 1-day observational study. Participants: Two hundred twelve eyes of 107 patients with DME involving the macular center by clinical examination and OCT central subfield thickness of >= 225 mu m. Methods: Retinal thickness was measured with the OCT3 system, and scans were evaluated by a reading center. Reproducibility of retinal thickness measurements was assessed, and 95% confidence intervals (CIs) for a change in thickness were estimated. Main Outcome Measures: Reproducibility of OCT-measured central subfield thickness. Results: Reproducibility was better for central subfield thickness than for center point thickness (half-width of the 95% CI for absolute change, 38 mu m vs. 50 mu m, and for relative change, 11 % vs. 17%, respectively; P<0.001). The median absolute difference between replicate measurements of the central subfield was 7 mu m (2%). Half-widths of the 95% CI for a change in central subfield thickness were 22, 23, 33, and 56 mu m for scans with central subfield thicknesses of <200, 200 to <250, 250 to <400, and >= 400 mu m, respectively. When expressed as percentage differences between 2 measurements, half-widths of the 95% CI for a change in central subfield thickness were 10%, 10%, 10%, and 13% for scans with central subfield thicknesses of <200, 200 to <250, 250 to <400, and >= 400 mu m, respectively. We were unable to identify an effect on reproducibility of central subfield measurements with respect to the presence of cystoid abnormalities, subretinal fluid, vitreomacular traction, or reduced visual acuity. Reproducibility was better when both scans had a standard deviation (SD) of the center point of <10.0% (half-width of the 95% CI for change, 33 mu m vs. 56 mu m; P<0.001). Conclusions: Reproducibility is better for central subfield thickness measurements than for center point measurements, and variability is less with retinal thickness when expressed as a percent change than when expressed as an absolute change. A change in central subfield thickness exceeding 11% is likely to be real. Scans with an SD of the center point of >= 10.0% are less reproducible and should be viewed with caution when assessing the validity of an observed change in retinal thickness inpatients with DME.
引用
收藏
页码:1520 / 1525
页数:6
相关论文
共 50 条
  • [1] Comparison of Central Macular Fluid Volume With Central Subfield Thickness in Patients With Diabetic Macular Edema Using Optical Coherence Tomography Angiography
    You, Qi Sheng
    Tsuboi, Kotaro
    Guo, Yukun
    Wang, Jie
    Flaxel, Christina J.
    Bailey, Steven T.
    Huang, David
    Jia, Yali
    Hwang, Thomas S.
    [J]. JAMA OPHTHALMOLOGY, 2021, 139 (07) : 734 - 741
  • [2] Reproducibility of Optovue RTVue Optical Coherence Tomography Retinal Thickness Measurements and Conversion to Equivalent Zeiss Stratus Metrics in Diabetic Macular Edema
    Bressler, Susan B.
    Edwards, Allison R.
    Andreoli, Christopher M.
    Edwards, Paul Andrew
    Glassman, Adam R.
    Jaffe, Glenn J.
    Melia, Michele
    Sun, Jennifer K.
    [J]. TRANSLATIONAL VISION SCIENCE & TECHNOLOGY, 2015, 4 (01):
  • [3] Optical coherence tomography in diabetic macular edema
    Baskin, Darrell E.
    [J]. CURRENT OPINION IN OPHTHALMOLOGY, 2010, 21 (03) : 172 - 177
  • [4] Comparison of Macular Thickness in Diabetic Macular Edema Using Spectral-Domain Optical Coherence Tomography and Time-Domain Optical Coherence Tomography
    Kakinoki, Masashi
    Miyake, Taichirou
    Sawada, Osamu
    Sawada, Tomoko
    Kawamura, Hajime
    Ohji, Masahito
    [J]. JOURNAL OF OPHTHALMOLOGY, 2012, 2012
  • [5] Reproducibility of topographic macular thickness mapping using optical coherence tomography
    Ghanta, RK
    Schuman, JS
    Guedes, VRF
    Drexler, W
    Hertzmark, E
    Miller, VM
    MacNutt, J
    Puliafito, CA
    Fujimoto, JG
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 1999, 40 (04) : S570 - S570
  • [6] RETINAL THICKNESS AND VOLUME MEASUREMENTS IN DIABETIC MACULAR EDEMA A Comparison of Four Optical Coherence Tomography Systems
    Lammer, Jan
    Scholda, Christoph
    Pruente, Christian
    Benesch, Thomas
    Schmidt-Erfurth, Ursula
    Bolz, Matthias
    [J]. RETINA-THE JOURNAL OF RETINAL AND VITREOUS DISEASES, 2011, 31 (01): : 48 - 55
  • [7] Imaging of diabetic macular edema with optical coherence tomography
    Yang, DCS
    Cheng, CY
    Lee, FL
    Hsu, WM
    Liu, JH
    [J]. INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2000, 41 (04) : S654 - S654
  • [8] Topography of diabetic macular edema with optical coherence tomography
    Hee, MR
    Puliafito, CA
    Duker, JS
    Reichel, E
    Coker, JG
    Wilkins, JR
    Schuman, JS
    Swanson, EA
    Fujimoto, JG
    [J]. OPHTHALMOLOGY, 1998, 105 (02) : 360 - 370
  • [9] Optical coherence tomography for assessment of diabetic macular edema
    Zhang, Wei
    Yamamoto, Kaori
    Hori, Sadao
    [J]. INTERNATIONAL JOURNAL OF OPHTHALMOLOGY, 2008, 1 (04) : 370 - 373
  • [10] Patterns of diabetic macular edema with optical coherence tomography
    Otani, T
    Kishi, S
    Maruyama, Y
    [J]. AMERICAN JOURNAL OF OPHTHALMOLOGY, 1999, 127 (06) : 688 - 693