The clinical significance of vacuum mixing bone cement

被引:0
|
作者
Geiger, MH [1 ]
Keating, EM [1 ]
Ritter, MA [1 ]
Ginther, JA [1 ]
Faris, PM [1 ]
Meding, JB [1 ]
机构
[1] St Francis Hosp, Ctr Hip & Knee Surg, Mooresville, IN USA
关键词
D O I
暂无
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
This controlled study compared the strength and porosity of 48 polymethylmethacrylate cement-implant constructs prepared with open bowl versus vacuum mix technique. Forty-eight blast finished stainless steel rods of 13 mm diameter were implanted with centralizers into 17-mm inner diameter tubes that had been retrograde filled with polymethylmethacrylate cement. The eight cement preparations used were open bowl and vacuum mixed Simplex, Osteobond, Zimmer Dough Type, or Palaces R. Six replications of each condition were performed. The tubes were maintained at 37 degrees C. Each tube was cut transversely into five segments. The center three segments were used for data analysis: pushout strength, cycles to failure, and interface porosity analysis. Rod pushout data showed there was no significant difference between open bowl and vacuum mixed samples when all cement brands were combined. Mean sheer force for Palaces R vacuum mixed samples was greater than open bowl (634 +/- 47 versus 423 +/- 171), whereas the force for the Zimmer Dough Type cement open howl was greater than that of the vacuum mixed samples (901 +/- 71 versus 705 +/- 82). Cycles to failure data did not show significant differences when open bowl and vacuum mixed samples were compared when cements were analyzed individually or combined. Image analysis of cement-implant interfaces showed that vacuum mixing reduced void area significantly compared with open bowl mixing in the Palaces R and Osteobond preparations. Vacuum mixing does not appear to reduce cement prosthesis interface porosity or improve its mechanical properties in all cements.
引用
收藏
页码:258 / 266
页数:9
相关论文
共 50 条
  • [21] MECHANICAL VERSUS MANUAL MIXING OF BONE-CEMENT
    LINDEN, U
    ACTA ORTHOPAEDICA SCANDINAVICA, 1988, 59 (04): : 400 - 402
  • [22] The influence of vacuum mixing on methylmethacrylate liberation from acrylic cement powder
    Bettencourt, A
    Calado, A
    Amaral, J
    Vale, FM
    Rico, JMT
    Monteiro, J
    Castro, M
    INTERNATIONAL JOURNAL OF PHARMACEUTICS, 2001, 219 (1-2) : 89 - 93
  • [23] INFLUENCE OF THE MIXING TECHNIQUE ON THE PROPERTIES OF PMMA BONE-CEMENT .1. CLINICAL-PRACTICE
    EYERER, P
    JIN, R
    BIOMEDIZINISCHE TECHNIK, 1985, 30 (12): : 316 - 325
  • [24] INFLUENCE OF THE MIXING TECHNIQUE ON THE PROPERTIES OF PMMA BONE-CEMENT .3. IMPROVED MIXING PROCEDURE - EFFECTS AND CLINICAL-TESTING
    EYERER, P
    JIN, R
    BIOMEDIZINISCHE TECHNIK, 1986, 31 (03): : 34 - 41
  • [25] The performance of three vacuum-mixing cement guns - a comparison of the fatigue properties of Simplex P cement
    Yau, WP
    Ng, TP
    Chiu, KY
    Poon, KC
    Ho, WY
    Luk, DK
    INTERNATIONAL ORTHOPAEDICS, 2001, 25 (05) : 290 - 293
  • [26] The performance of three vacuum-mixing cement guns – a comparison of the fatigue properties of Simplex P cement
    W. Yau
    T. Ng
    K. Chiu
    K. Poon
    W. Ho
    D. Luk
    International Orthopaedics, 2001, 25 : 290 - 293
  • [27] Mixing of acrylic bone cement: effect of oxygen on setting properties
    He, SL
    Scott, C
    Higham, P
    BIOMATERIALS, 2003, 24 (27) : 5045 - 5048
  • [28] Influence of mixing techniques on the physical properties of acrylic bone cement
    Dunne, NJ
    Orr, JF
    BIOMATERIALS, 2001, 22 (13) : 1819 - 1826
  • [29] AIR INCLUSION IN BONE-CEMENT - IMPORTANCE OF THE MIXING TECHNIQUE
    LINDEN, U
    GILLQUIST, J
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1989, (247) : 148 - 151
  • [30] FATIGUE PROPERTIES OF BONE-CEMENT - COMPARISON OF MIXING TECHNIQUES
    LINDEN, U