Techniques to measure rigidity of ankle-foot orthosis: A review

被引:33
|
作者
Kobayashi, Toshiki [1 ]
Leung, Aaron K. L. [2 ]
Hutchins, Stephen W. [3 ]
机构
[1] Orthocare Innovat, Mountlake Terrace, WA 98043 USA
[2] Hong Kong Polytech Univ, Dept Hlth Technol & Informat, Hong Kong, Hong Kong, Peoples R China
[3] Univ Salford, Ctr Hlth Sport & Rehabil Sci Res, Salford M5 4WT, Lancs, England
来源
关键词
AFO; ankle-foot orthosis; cerebral palsy; flexibility; hemiplegia; mechanical test; rigidity; spinal cord injury; stiffness; stroke; testing and measurement; DYNAMIC STRESS ANALYSIS; BENDING STIFFNESS; BRACES; POLYPROPYLENE; GAIT; KINEMATICS; APPARATUS; TRIMLINE; DESIGN; MOMENT;
D O I
10.1682/JRRD.2010.10.0193
中图分类号
R49 [康复医学];
学科分类号
100215 ;
摘要
We performed this review to provide a clearer understanding of how to effectively measure ankle-foot orthosis (AFO) rigidity. This information is important to ensure appropriate orthotic intervention in the treatment of patients with pathological gait. The two main approaches to the investigation of AFO rigidity are (1) bench-testing analyses, in which an AFO is fixed or attached to a measurement device, and (2) functional analyses, in which measurements are taken while a subject is walking with an AFO in situ. This review summarizes and classifies the current state of knowledge of AFO rigidity testing methods. We analyzed the strengths and weaknesses of the methods in order to recommend the most reliable techniques to measure AFO rigidity. The information obtained from this review article would, therefore, benefit both clinicians and engineers involved in the application and design of AFOs.
引用
收藏
页码:565 / 576
页数:12
相关论文
共 50 条
  • [21] A Dynamically Consistent Model of a Motorized Ankle-Foot Orthosis
    Vivian, Michele
    Reggiani, Monica
    Moreno, Juan C.
    Pons, Jose L.
    Farina, Dario
    Sartori, Massimo
    2013 6TH INTERNATIONAL IEEE/EMBS CONFERENCE ON NEURAL ENGINEERING (NER), 2013, : 1558 - 1561
  • [22] INFLUENCE OF HEEL DESIGN ON A RIGID ANKLE-FOOT ORTHOSIS
    WIEST, DR
    WATERS, RL
    BONTRAGER, EL
    QUIGLEY, MJ
    ORTHOTICS AND PROSTHETICS, 1979, 33 (04): : 3 - 10
  • [23] Commonly Used Types and Recent Development of Ankle-Foot Orthosis: A Narrative Review
    Choo, Yoo Jin
    Chang, Min Cheol
    HEALTHCARE, 2021, 9 (08)
  • [24] Ankle-foot orthosis improves the gait: case report
    Paniccia, C.
    Albertini, G.
    JOURNAL OF APPLIED RESEARCH IN INTELLECTUAL DISABILITIES, 2010, 23 (05) : 473 - 473
  • [25] A hybrid ankle-foot orthosis with soft pneumatic actuation
    Marconi, Grace P.
    Gopalai, Alpha A.
    Chauhan, Sunita
    MECHATRONICS, 2024, 99
  • [26] Use of an ankle-foot orthosis in the management of pretibial lacerations
    Kumar, P
    De Souza, B
    Carver, N
    PLASTIC AND RECONSTRUCTIVE SURGERY, 2003, 112 (05) : 1497 - 1498
  • [27] Effect of ankle-foot orthosis on postural control after stroke: a systematic review
    Guerra Padilla, M.
    Molina Rueda, F.
    Alguacil Diego, I. M.
    NEUROLOGIA, 2014, 29 (07): : 423 - 432
  • [28] Fuel Efficiency of a Portable Powered Ankle-Foot Orthosis
    Boes, Morgan K.
    Islam, Mazharul
    Li, Yifan David
    Hsiao-Wecksler, Elizabeth T.
    2013 IEEE 13TH INTERNATIONAL CONFERENCE ON REHABILITATION ROBOTICS (ICORR), 2013,
  • [29] Experimental and computational analysis of composite ankle-foot orthosis
    Zou, Dequan
    He, Tao
    Dailey, Michael
    Smith, Kirk E.
    Silva, Matthew J.
    Sinacore, David R.
    Mueller, Michael J.
    Hastings, Mary K.
    JOURNAL OF REHABILITATION RESEARCH AND DEVELOPMENT, 2014, 51 (10): : 1525 - 1535
  • [30] Multiscale modeling of variable stiffness ankle-foot orthosis
    Redekop, S
    Meguid, SA
    Sagals, G
    Stranart, JC
    Parker, K
    COMPUTATIONAL FLUID AND SOLID MECHANICS 2003, VOLS 1 AND 2, PROCEEDINGS, 2003, : 1794 - 1796