Performance of computer-aided detection applied to full-field digital mammography in detection of breast cancers

被引:33
|
作者
Sadaf, Arifa [1 ]
Crystal, Pavel [1 ]
Scaranelo, Anabel [2 ]
Helbich, Thomas [3 ]
机构
[1] Mt Sinai Hosp, Dept Med Imaging, Toronto, ON M5G 1X5, Canada
[2] Princess Margaret Hosp, Dept Med Imaging, Toronto, ON M5G 2M9, Canada
[3] Med Univ Vienna, Gen Hosp Vienna, A-1090 Vienna, Austria
关键词
Mammography; Computer-assisted detection; Breast cancer; SCREENING MAMMOGRAPHY; DETECTION SYSTEM; FILM MAMMOGRAPHY; DETECTION CAD; SENSITIVITY; APPEARANCE; PROGRAM; IMPACT;
D O I
10.1016/j.ejrad.2009.08.024
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The aim of this retrospective study was to evaluate performance of computer-aided detection (CAD) with full-field digital mammography (FFDM) in detection of breast cancers. Materials and Methods: CAD was retrospectively applied to standard mammographic views of 127 cases with biopsy proven breast cancers detected with FFDM (Senographe 2000, GE Medical Systems). CAD sensitivity was assessed in total group of 127 cases and for subgroups based on breast density, mammographic lesion type, mammographic lesion size, histopathology and mode of presentation. Results: Overall CAD sensitivity was 91% (115 of 127 cases). There were no statistical differences (p > 0.1) in CAD detection of cancers in dense breasts 90% (53/59) versus non-dense breasts 91% (62/68). There was statistical difference (p < 0.05) in CAD detection of cancers that appeared mammographically as micro-calcifications only versus other mammographic manifestations. CAD detected 100% (44/44) of cancers manifesting as microcalcifications, 89% (47/53) as no-calcified masses or asymmetries, 88% (14/16) as masses with associated calcifications, and 71% (10/14) as architectural distortions. CAD sensitivity for cancers 1-10 mm was 84% (38/45); 11-20 mm 93% (55/59); and > 20 mm 97% (22/23). Conclusion: CAD applied to FFDM showed 100% sensitivity in identifying cancers manifesting as microcalcifications only and high sensitivity 86% (71/83) for other mammographic appearances of cancer. Sensitivity is influenced by lesion size. CAD in FFDM is an adjunct helping radiologist in early detection of breast cancers. (C) 2009 Elsevier Ireland Ltd. All rights reserved.
引用
收藏
页码:457 / 461
页数:5
相关论文
共 50 条
  • [21] Application of computer-aided diagnosis to full-field digital mammography
    Yarusso, LM
    Nishikawa, RM
    Papaioannou, J
    Nagel, R
    Jokich, P
    Venta, LA
    [J]. IWDM 2000: 5TH INTERNATIONAL WORKSHOP ON DIGITAL MAMMOGRAPHY, 2001, : 421 - 426
  • [22] Computer-aided diagnosis of masses with full-field digital mammography
    Li, LH
    Clark, RA
    Thomas, JA
    [J]. ACADEMIC RADIOLOGY, 2002, 9 (01) : 4 - 12
  • [23] A comparison between digital breast tomosynthesis and full-field digital mammography for the detection of breast cancers
    Choi, Woo Jung
    Kim, Hak Hee
    Lee, Sun Young
    Chae, Eun Young
    Shin, Hee Jung
    Cha, Joo Hee
    Son, Byung Ho
    Ahn, Sei Hyun
    Choi, Young-Wook
    [J]. BREAST CANCER, 2016, 23 (06) : 886 - 892
  • [24] A comparison between digital breast tomosynthesis and full-field digital mammography for the detection of breast cancers
    Woo Jung Choi
    Hak Hee Kim
    Sun Young Lee
    Eun Young Chae
    Hee Jung Shin
    Joo Hee Cha
    Byung Ho Son
    Sei Hyun Ahn
    Young-Wook Choi
    [J]. Breast Cancer, 2016, 23 : 886 - 892
  • [25] Breast Cancer Detection in a Screening Population: Comparison of Digital Mammography, Computer-Aided Detection Applied to Digital Mammography and Breast Ultrasound
    Cho, Kyu Ran
    Seo, Bo Kyoung
    Woo, Ok Hee
    Song, Sung Eun
    Choi, Jungsoon
    Whang, Shin Young
    Park, Eun Kyung
    Park, Ah Young
    Shin, Hyeseon
    Chung, Hwan Hoon
    [J]. JOURNAL OF BREAST CANCER, 2016, 19 (03) : 316 - 323
  • [26] Sensitivity of integral computer-aided detection with full-field digital mammography for detection of breast cancer according to different histopathological tumor types and appearances
    K Pinker
    N Perry
    S Milner
    K Mokbel
    S Duffy
    [J]. Breast Cancer Research, 12
  • [27] Sensitivity of integral computer-aided detection with full-field digital mammography for detection of breast cancer according to different histopathological tumor types and appearances
    Pinker, K.
    Perry, N.
    Milner, S.
    Mokbel, K.
    Duffy, S.
    [J]. BREAST CANCER RESEARCH, 2010, 12
  • [28] Sensitivity of a Direct Computer-aided Detection System in Full-field Digital Mammography for Detection of Microcalcifications Not Associated with Mass or Architectural Distortion
    Scaranelo, Anabel M.
    Crystal, Pavel
    Bukhanov, Karina
    Helbich, Thomas H.
    [J]. CANADIAN ASSOCIATION OF RADIOLOGISTS JOURNAL-JOURNAL DE L ASSOCIATION CANADIENNE DES RADIOLOGISTES, 2010, 61 (03): : 162 - 169
  • [29] Retrospective analysis of a computer-aided detection (CAD) system in full-field digital mammography in correlation to tumor histology
    Obenauer, S
    Sohns, C
    Werner, C
    Grabbe, E
    [J]. ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2005, 177 (08): : 1103 - 1109
  • [30] Who could benefit the most from using a computer-aided detection system in full-field digital mammography?
    Jung, Na Young
    Kang, Bong Joo
    Kim, Hyeon Sook
    Cha, Eun Suk
    Lee, Jae Hee
    Park, Chang Suk
    Whang, In Young
    Kim, Sung Hun
    An, Yeong Yi
    Choi, Jae Jeong
    [J]. WORLD JOURNAL OF SURGICAL ONCOLOGY, 2014, 12