Minimizing RF heating of conducting wires in MRI

被引:51
|
作者
Yeung, Christopher J.
Karmarkar, Parag
McVeigh, Elliot R.
机构
[1] NHLBI, Cardiac Energet Lab, Natl Inst Hlth, Div Intramural Res, Bethesda, MD 20892 USA
[2] Johns Hopkins Univ, Sch Med, Dept Radiol, Baltimore, MD 21205 USA
关键词
MRI; safety; RF heating; interventional MRI; guidewires; catheters; MRI-guided surgery; moment method; conductive wires;
D O I
10.1002/mrm.21410
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Performing interventions using long conducting wires in MRI introduces the risk of focal RF heating at the wire tip. Comprehensive EM simulations are combined with carefully measured experimental data to show that method-of-moments EM field modeling coupled with heat transfer modeling can adequately predict RF heating with wires partially inserted into the patient-mimicking phantom. The effects of total wire length, inserted length, wire position in the phantom, phantom position in the scanner, and phantom size are examined. Increasing phantom size can shift a wire's length of maximum tip heating from about a half wave toward a quarter wave. In any event, with wires parallel to the scanner bore, wire tip heating is minimized by keeping the patient and wires as close as possible to the central axis of the scanner bore. At 1.5T, heating is minimized if bare wires are shorter than 0.6 m or between approximate to 2.4 m and approximate to 3.0 m. Heating is further minimized if wire insertion into phantoms equivalent to most aqueous soft tissues is less than 13 cm or greater than 40 cm (longer for fatty tissues, bone, and lung). The methods demonstrated can be used to estimate the absolute amount of heating in order to set RF power safety thresholds.
引用
收藏
页码:1028 / 1034
页数:7
相关论文
共 50 条
  • [41] Evaluation of local B1 field as dosimeter of RF heating for implant in MRI
    Wan, Sen
    Jiang, Changqing
    Ding, Jianqi
    Li, Luming
    ELECTRONICS LETTERS, 2019, 55 (06) : 302 - 303
  • [42] Evaluation of MRI RF electromagnetic field induced heating near leads of cochlear implants
    Zeng, Qi
    Wang, Qinyan
    Zheng, Jianfeng
    Kainz, Wolfgang
    Chen, Ji
    PHYSICS IN MEDICINE AND BIOLOGY, 2018, 63 (13):
  • [43] A numerical investigation of RF heating effects on implants during MRI compared to experimental measurements
    Stenschke, Jens
    Li, Dan
    Thomann, Maike
    Schaefers, Gregor
    Zylka, Waldemar
    ADVANCES IN MEDICAL ENGINEERING, 2007, 114 : 53 - +
  • [44] Spatial distribution of RF-induced E-fields and implant heating in MRI
    Nordbeck, Peter
    Fidler, Florian
    Weiss, Ingo
    Warmuth, Marcus
    Friedrich, Michael T.
    Ehses, Philipp
    Geistert, Wolfgang
    Ritter, Oliver
    Jakob, Peter M.
    Ladd, Mark E.
    Quick, Harald H.
    Bauer, Wolfgang R.
    MAGNETIC RESONANCE IN MEDICINE, 2008, 60 (02) : 312 - 319
  • [45] CONDUCTING POLYMER WIRES IN MESOPORE HOSTS
    WU, CG
    BEIN, T
    ZEOLITES AND RELATED MICROPOROUS MATERIALS: STATE OF THE ART 1994, 1994, 84 : 2269 - 2276
  • [46] Improved Electrical Conducting Wires for SOFCs
    Masomtob, M.
    Wongtida, K.
    Charoensuk, J.
    Charojrochkul, S.
    SMART MATERIALS, 2008, 55-57 : 797 - +
  • [47] RF heating in stellarators
    Natl Inst for Fusion Science, Toki-shi, Japan
    Plasma Physics and Controlled Fusion, 1998, 40 (8 A)
  • [48] RF heating in stellarators
    Watari, T
    PLASMA PHYSICS AND CONTROLLED FUSION, 1998, 40 : A13 - A34
  • [49] Impact of RF Shimming on RF-Induced Heating Near Implantable Medical Electrodes in a 3T MRI Coil
    Zeng, Qi
    Wang, Qingyan
    Kainz, Wolfgang
    Chen, Ji
    IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 2020, 62 (01) : 52 - 64
  • [50] Electrode Sensitivity for MRI-RF Induced Heating Evaluation of Active Implantable Medical Device
    Long, Tiangang
    Jiang, Changqing
    Li, Luming
    2023 IEEE MTT-S INTERNATIONAL MICROWAVE BIOMEDICAL CONFERENCE, IMBIOC, 2023, : 67 - 69