Transient evoked otoacoustic emissions in patients with normal hearing and in patients with hearing loss

被引:62
|
作者
Hussain, DM
Gorga, MP
Neely, ST
Keefe, DH
Peters, J
机构
[1] Indiana Univ, Sch Med, Dept Otolaryngol, Indianapolis, IN 46202 USA
[2] Boys Town Natl Res Hosp, Omaha, NE 68131 USA
来源
EAR AND HEARING | 1998年 / 19卷 / 06期
关键词
D O I
10.1097/00003446-199812000-00005
中图分类号
R36 [病理学]; R76 [耳鼻咽喉科学];
学科分类号
100104 ; 100213 ;
摘要
Objectives: 1) To evaluate transient evoked otoacoustic emission (TEOAE) test performance when measurements are made under routine clinical conditions. 2) To evaluate TEOAE: test performance as a function of frequency and as a function of the magnitude of hearing loss. 3) To compare test performance using univariate and multivariate approaches to data analyses. 4) To provide a means of interpreting clinical TEOAE measurements. Design: TEOAEs were measured in 452 ears of 246 patients. All measurements were made after acoustic immittance assessments, which were used to demonstrate that middle-ear function was normal at the time of the TEOAE: test. TEOAE amplitudes and signal to noise ratios (SNRs), analyzed into octave bands centered at 1, 2, and 4 kHz, were compared with the pure-tone threshold at the same frequencies. Data were analyzed with clinical decision theory, cumulative distributions, discriminant analyses, and logistic regressions. Results: Using univariate analysis techniques, TEOAEs accurately identified auditory status at 2 and 4 kHz but were less accurate at 1 kHz. Test performance was best when audiometric thresholds between 20 and 30 dB HL were used as the criteria for normal hearing. TEOAE SNR resulted in better test performance than did TEOAE amplitude alone; this effect decreased as frequency increased. Multivariate analysis methods resulted in better separation between normal and impaired ears than did univariate approaches, which relied on only TEOAE: amplitude or SNR when test frequency band and audiometric frequency were the same. This improvement in test performance was greatest at 1 kHz, decreased as frequency increased, and was negligible at 4 kHz. Conclusions: TEOAEs can be used to identify hearing loss in children under routine clinical conditions. Univariate tests accurately identified auditory status at mid and high frequencies but performed more poorly at lower frequencies. The decrease in performance as frequency decreases may be a result of increased noise at lower frequencies but also may be due to properties of the measurement paradigm ("QuickScreen," high-pass filter at 0.8 kHz), which would not be ideal for recording energy around I kHz. The improvement in test performance when SNR was used and the interaction of this effect with frequency, however, would be consistent with the view that test performance in lower frequencies is at least partially influenced by the level of background noise. Multivariate analysis techniques improved test performance compared with the more traditional univariate approaches to data analysis. An approach is provided that allows one to assign measured TEOAE amplitudes, SNRs, or outputs from multivariate analyses to one of three categories: response properties consistent with normal hearing; results consistent with hearing loss; hearing status undetermined.
引用
收藏
页码:434 / 449
页数:16
相关论文
共 50 条
  • [11] Identification of hearing loss in newborns by transient otoacoustic emissions
    Jakubíková, J
    Kabátová, Z
    Závodná, M
    [J]. INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 2003, 67 (01) : 15 - 18
  • [12] USING TRANSIENT EVOKED OTOACOUSTIC EMISSIONS FOR NEONATAL HEARING SCREENING
    MAXON, AB
    WHITE, KR
    VOHR, BR
    BEHRENS, TR
    [J]. BRITISH JOURNAL OF AUDIOLOGY, 1993, 27 (02): : 149 - 153
  • [13] SCREENING ALL NEWBORNS FOR HEARING-LOSS USING TRANSIENT EVOKED OTOACOUSTIC EMISSIONS
    WHITE, KR
    VOHR, BR
    MAXON, AB
    BEHRENS, TR
    MCPHERSON, MG
    MAUK, GW
    [J]. INTERNATIONAL JOURNAL OF PEDIATRIC OTORHINOLARYNGOLOGY, 1994, 29 (03) : 203 - 217
  • [14] Wavelet analysis of transient evoked otoacoustic emissions for differential diagnosis of cochlear hearing loss
    Weiss, S
    Hoppe, U
    Schabert, M
    Eysholdt, U
    [J]. CONFERENCE RECORD OF THE THIRTY-FIFTH ASILOMAR CONFERENCE ON SIGNALS, SYSTEMS AND COMPUTERS, VOLS 1 AND 2, 2001, : 824 - 828
  • [15] Evaluation of hearing loss in at risk neonates by evoked otoacoustic emissions
    Ashraf Husain
    [J]. Indian Journal of Otolaryngology and Head and Neck Surgery, 1997, 49 (Suppl 1): : 32 - 35
  • [16] Normal evoked otoacoustic emissions with a profound hearing loss due to a juvenile pilocytic astrocytoma
    Monroe, JAB
    Krauth, L
    Arenberg, IK
    Prenger, E
    Philpott, P
    [J]. AMERICAN JOURNAL OF OTOLOGY, 1996, 17 (04): : 639 - 642
  • [17] EVOKED OTOACOUSTIC EMISSIONS IN DIAGNOSIS OF SENSORY HEARING-LOSS
    SIMONCELLI, C
    RICCI, G
    MOLINI, E
    VONGARREL, C
    [J]. LARYNGO-RHINO-OTOLOGIE, 1992, 71 (06) : 319 - 322
  • [18] Multi variant analysis of otoacoustic emissions and estimation of hearing thresholds: Transient evoked otoacoustic emissions
    Vinck, BM
    Van Cauwenberge, PB
    Corthals, P
    De Vel, E
    [J]. AUDIOLOGY, 1998, 37 (06): : 315 - 334
  • [19] Otoacoustic emissions (TEOAE) in diabetic patients (IDDM) with normal hearing
    Hosch, H
    Ottaviani, F
    [J]. SCHWEIZERISCHE MEDIZINISCHE WOCHENSCHRIFT, 2000, 130 (49) : 83S - 85S
  • [20] CLICK-EVOKED OTOACOUSTIC EMISSIONS AND HEARING THRESHOLD IN SENSORINEURAL HEARING-LOSS
    COLLET, L
    LEVY, V
    VEUILLET, E
    TRUY, E
    MORGON, A
    [J]. EAR AND HEARING, 1993, 14 (02): : 141 - 143