Jet ventilation in upper airway obstruction: Description and model lung testing of a new jetting device

被引:7
|
作者
Garry, B
Woo, P
Perrault, DF
Shapshay, SM
Wurm, WH
机构
[1] New England Med Ctr, Dept Anesthesia, Boston, MA 02111 USA
[2] New England Med Ctr, Dept Otolaryngol, Boston, MA 02111 USA
[3] Tufts Univ, Sch Med, Boston, MA 02111 USA
来源
ANESTHESIA AND ANALGESIA | 1998年 / 87卷 / 04期
关键词
D O I
10.1097/00000539-199810000-00032
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Patients with critical upper airway stenosis require a tracheotomy for corrective surgery. We describe a new transtracheal device that permits safe ventilation of these patients without tracheotomy. It is based on a coaxial bicannular design that allows "push-pull" ventilation by jetting gas through the inner cannula and applying suction through the outer cannula. It further allows monitoring of airway pressure, tidal volume, and end-tidal CO2. The device was placed in the "trachea" of an artificial lung, and the preparation was made airtight by sealing the proximal end of the trachea. Tidal volumes and their associated pressures were measured simultaneously at different parts of the airway at several lung compliances and airway resistance settings while varying the jet and suction pressures. A large range of tidal volumes was achieved at safe airway pressures using clinically relevant airway resistance and lung compliance settings. Airway pressures measured through the device correlated well with pressures measured directly in the airways at the same time. Tidal volumes, measured through a Wright respirometer in the suction line, exceeded actual values at high suction settings and decreased below actual values at low suction settings. This new form of jet ventilation allowed efficient ventilation of the artificial lung with a totally occluded upper airway. Implications: Tracheotomy is required for surgery to relieve strider because gas forced into the trachea at high pressures through a percutaneously placed needle (jetting) cannot be exhaled quickly enough for respiration. We describe a device that allows jetting in the stridorous patient by actively assisting expiration, thereby eliminating the tracheotomy requirement.
引用
收藏
页码:915 / 920
页数:6
相关论文
共 50 条
  • [31] INTRANASAL AIRWAY PACK - DESCRIPTION OF A NEW DEVICE
    DOYLE, DE
    STOLLER, KP
    LARYNGOSCOPE, 1983, 93 (06): : 808 - 809
  • [32] CERVICAL LUNG HERNIATION ASSOCIATED WITH UPPER AIRWAY-OBSTRUCTION
    DELREY, JG
    CUNHA, C
    ANNALS OF EMERGENCY MEDICINE, 1990, 19 (08) : 935 - 937
  • [33] One lung ventilation in a patient with an upper and lower airway abnormality
    Kaza, Sangita R.
    Maddali, Madan M.
    Albahrani, Maher J.
    Vaghari, Adel A.
    INDIAN JOURNAL OF ANAESTHESIA, 2012, 56 (06) : 567 - 569
  • [34] Airway obstruction during one lung ventilation: A shocking twist in the tube
    Ahluwalia, C. S.
    Kiran, S.
    Chopra, V.
    Kar, Soumita
    INDIAN JOURNAL OF ANAESTHESIA, 2014, 58 (04) : 497 - 499
  • [35] Detection and relief of upper airway obstruction in a dog model
    Durand, DM
    Sahin, M
    Haxhiu, M
    Yoo, PB
    SECOND JOINT EMBS-BMES CONFERENCE 2002, VOLS 1-3, CONFERENCE PROCEEDINGS: BIOENGINEERING - INTEGRATIVE METHODOLOGIES, NEW TECHNOLOGIES, 2002, : 1548 - 1549
  • [36] Upper Airway Obstruction during Noninvasive Ventilation Induced by the Use of an Oronasal Mask
    Vrijsen, Bart
    Buyse, Bertien
    Belge, Catharina
    Testelmans, Dries
    JOURNAL OF CLINICAL SLEEP MEDICINE, 2014, 10 (09): : 1033 - 1035
  • [37] FUNCTIONAL UPPER AIRWAY-OBSTRUCTION - A NEW SYNDROME
    APPELBLATT, NH
    BAKER, SR
    ARCHIVES OF OTOLARYNGOLOGY-HEAD & NECK SURGERY, 1981, 107 (05) : 305 - 306
  • [38] THE ESOPHAGEAL OBTURATOR AIRWAY, A NEW DEVICE FOR EMERGENCY VENTILATION
    HOFLEHNER, G
    GRUNDNER, H
    ROOB, P
    WIENER MEDIZINISCHE WOCHENSCHRIFT, 1984, 134 (12) : 305 - 308
  • [39] TRANSLARYNGEAL JET VENTILATION AND END-TIDAL PCO2 MONITORING DURING VARYING DEGREES OF UPPER AIRWAY-OBSTRUCTION
    WARD, KR
    MENEGAZZI, JJ
    YEALY, DM
    KLAIN, MM
    MOLNER, RL
    GOODE, JS
    ANNALS OF EMERGENCY MEDICINE, 1991, 20 (11) : 1193 - 1197
  • [40] A bench study of ventilation via two self-assembled jet devices and the Oxygen Flow Modulator in simulated upper airway obstruction
    Hamaekers, A. E. W.
    Borg, P. A. J.
    Enk, D.
    ANAESTHESIA, 2009, 64 (12) : 1353 - 1358