Primary stability of self-tapping dual etched implants

被引:0
|
作者
Kassim, Zethy H. [1 ]
Nor Hisham, Nor D. [2 ]
Dardiri, Nur A. [2 ]
Goot Heah, Khor [3 ]
Hazwani Baharuddin, Izyan [3 ]
De Angelis, Nicola [1 ,4 ]
机构
[1] MARA Technol Univ, Fac Dent, Ctr Studies Restorat Dent, Jalan Hosp, Sungai Buloh, Malaysia
[2] Minist Hlth Malaysia, Putrajaya, Malaysia
[3] MARA Technol Univ, Fac Dent, Ctr Studies Preclin Sci, Jalan Hosp, Sungai Buloh, Malaysia
[4] Univ Genoa, Dept Surg Sci & Integrated Diagnost DISC, Largo R Benzi 10, I-16132 Genoa, Italy
来源
MINERVA STOMATOLOGICA | 2019年 / 68卷 / 06期
关键词
Dental implants; Osseointegration; Osteotomy; Validation studies; TRABECULAR BONE; OSSEOINTEGRATION; ANCHORAGE; DIAMETER; DENSITY; LENGTH;
D O I
10.23736/S0026-4970.19.04242-0
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
BACKGROUND: The aims of this study were to enumerate the primary implant stability quotient (ISQ) value of self-tapping dual etched implants and to explore the influence of parameters such as implant length, implant diameter, age, gender, implant location and osteotomy preparation on the ISQ value. METHODS: Retrospective data from clinical worksheets given to participants during two implant courses held between the periods of 2013 to 2014 were evaluated. A total of 61 implants were considered based on the inclusion criteria. The effects of parameters such as implant diameter, implant length, age, gender, implant location and osteotomy protocol on ISQ values were analyzed. RESULTS: Mean ISQ value for all implants was 67.21 +/- 9.13. Age of patients (P=0.016) and location of implants (P=0.041) had a significant linear relationship with the ISQ values. Within the age limit of the patients in this study, it was found that an increase in one year of patient's age results in 0.20 decrease in ISQ value (95% CI: -0.36, -0.04). However, placing an implant in the posterior maxilla may negatively affect the ISQ with a likely decrease in primary stability by 6.76 ISQ value (95% CI: -13.22, -0.30). CONCLUSIONS: The results suggest that the mean ISQ achieved by the participants were comparable with the range reported for this particular type of implants. The patient's age and location of implants were elucidated as the determinant factors of primary implant stability.
引用
收藏
页码:291 / 296
页数:6
相关论文
共 50 条
  • [1] Influence of Residual Bone Thickness on Primary Stability of Hybrid Self-Tapping and Cylindric Non-Self-Tapping Implants in Vitro
    Divac, Marija
    Stawarczyk, Bogna
    Sahrmann, Philipp
    Attin, Thomas
    Schmidlin, Patrick R.
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2013, 28 (01) : 84 - 88
  • [2] Evaluation of Primary Stability of Self-Tapping and Non-Self-Tapping Dental Implants. A 12-Week Clinical Study
    Markovic, Aleksa
    Luis Calvo-Guirado, Jose
    Lazic, Zoran
    Gomez-Moreno, Gerardo
    Calasan, Dejan
    Guardia, Javier
    Colic, Snjezana
    Aguilar-Salvatierra, Antonio
    Gacic, Bojan
    Delgado-Ruiz, Rafael
    Janjic, Bojan
    Misic, Tijana
    CLINICAL IMPLANT DENTISTRY AND RELATED RESEARCH, 2013, 15 (03) : 341 - 349
  • [3] Comparison of primary stability of used and unused self-tapping and self-drilling orthodontic mini-implants
    Baser, Baris
    Ozel, Mehmet Birol
    ADVANCES IN CLINICAL AND EXPERIMENTAL MEDICINE, 2024, 33 (05): : 483 - 489
  • [4] Does Length, Diameter, or Bone Quality Affect Primary and Secondary Stability in Self-Tapping Dental Implants?
    Gomez-Polo, Miguel
    Ortega, Rocio
    Gomez-Polo, Cristina
    Martin, Cristina
    Celemin, Alicia
    del Rio, Jaime
    JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2016, 74 (07) : 1344 - 1353
  • [5] Primary stability and self-tapping blades: biomechanical assessment of dental implants in medium-density bone
    Kim, Yung-Soo
    Lim, Young-Jun
    CLINICAL ORAL IMPLANTS RESEARCH, 2011, 22 (10) : 1179 - 1184
  • [6] Effect of implant design on the initial biomechanical stability of two self-tapping dental implants
    Hsieh, Min-Chieh
    Huang, Chang-Hung
    Lin, Chun-Li
    Hsu, Ming-Lun
    CLINICAL BIOMECHANICS, 2020, 74 : 124 - 130
  • [7] Comparative study of the primary stability of self-drilling and self-tapping orthodontic miniscrews
    Son, Seil
    Motoyoshi, Mitsuru
    Uchida, Yasuki
    Shimizu, Noriyoshi
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2014, 145 (04) : 480 - 485
  • [8] On the stability efficiency of anchorage self-tapping screws: Ex vivo experiments on miniscrew implants used in orthodontics
    Migliorati, Marco
    Drago, Sara
    Dalessandri, Domenico
    Lagazzo, Alberto
    Gallo, Fabio
    Capurro, Marco
    Silvestrini-Biavati, Armando
    JOURNAL OF THE MECHANICAL BEHAVIOR OF BIOMEDICAL MATERIALS, 2018, 81 : 46 - 51
  • [9] Self-tapping bolts
    Ashley, S
    MECHANICAL ENGINEERING, 1996, 118 (05) : 26 - 26
  • [10] Self-tapping screws as reinforcement
    Trautz, Martin
    Koj, Christoph
    BAUTECHNIK, 2008, 85 (03) : 190 - 196