Pacemaker Lead Tip Heating in Abandoned and Pacemaker-Attached Leads at 1.5 Tesla MRI

被引:113
|
作者
Langman, Deborah A. [1 ,2 ]
Goldberg, Ira B. [3 ]
Finn, J. Paul [1 ,2 ]
Ennis, Daniel B. [1 ,2 ]
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, Dept Radiol Sci, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Biomed Phys Interdept Program, Los Angeles, CA USA
[3] Univ Calif Los Angeles, Dept Elect Engn, Los Angeles, CA 90024 USA
关键词
magnetic resonance imaging; medical device safety; pacemaker; RF-induced heating; abandoned leads; IMPLANTABLE-CARDIOVERTER-DEFIBRILLATORS; MAGNETIC-RESONANCE; CARDIAC-PACEMAKERS; MANAGE COMPLICATIONS; IN-VIVO; SAFETY; EXTRACTION; ISSUES; RECOGNIZE; 1.5-TESLA;
D O I
10.1002/jmri.22463
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the risk of RF-induced heating in pacemaker-attached and abandoned leads using in vitro temperature measurements at 1.5 Tesla as a function of lead length. Materials and Methods: Five custom lead lengths, 20-60 cm, were exposed to a uniform magnitude and phase radio-frequency electric field to examine the effect of lead length on pacemaker lead tip heating for pacemaker-attached and abandoned pacemaker leads. Results: Abandoned and pacemaker-attached leads show resonant heating behavior and maximum heating occurs at different lead lengths due to the differences in termination conditions. For clinical lead lengths (40-60 cm) abandoned leads exhibited greater lead tip heating compared with pacemaker-attached leads. Conclusion: Current recommendations for MRI pacemaker safety should highlight the possible increased risk for patients with abandoned leads as compared to pacemaker-attached leads.
引用
收藏
页码:426 / 431
页数:6
相关论文
共 50 条
  • [31] In vivo heating of pacemaker leads during magnetic resonance imaging
    Luechinger, R
    Zeijlemaker, VA
    Pedersen, EM
    Mortensen, P
    Falk, E
    Duru, F
    Candinas, R
    Boesiger, P
    EUROPEAN HEART JOURNAL, 2005, 26 (04) : 376 - 383
  • [32] Percutaneous pacemaker lead extraction and stent implantation for superior vena cava occlusion due to pacemaker leads
    Rosenthal, E
    Qureshi, SA
    Tynan, M
    Bucknall, CA
    AMERICAN JOURNAL OF CARDIOLOGY, 1996, 77 (08): : 670 - &
  • [33] Old is gold: Tip electrograms to diagnose pacemaker lead perforation
    Chandan, K
    Ponde, C
    Lokhandwala, Y
    Nandkumar, K
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2002, 13 (10) : 1063 - 1063
  • [34] Late pacemaker lead tip perforation documented by chest CT
    Jain, Neeraj
    Ravipati, Vijay
    Kerut, Edmund K.
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2016, 33 (09): : 1419 - 1421
  • [35] Use of an endoscopic bioptome for extraction of a retained pacemaker lead tip
    Ji, Sang Yong
    Catanzaro, John
    Crosson, Jane E.
    Brinker, Jeffrey
    Cheng, Alan
    ARCHIVES OF MEDICAL SCIENCE, 2014, 10 (04) : 853 - 854
  • [36] A Rare Case of Rectus Muscle Twitching Due to Abandoned Uncapped Pacemaker Leads
    Al-Khdour, Issa
    Alaqra, Amro
    Nairat, Moath
    Marai, Ibrahim
    Yaghi, Nadine
    Awwad, Fateh
    CUREUS JOURNAL OF MEDICAL SCIENCE, 2023, 15 (11)
  • [37] Safety and Outcomes of Magnetic Resonance Imaging in Patients with Abandoned Pacemaker and Defibrillator Leads
    Higgins, John V.
    Gard, Joseph J.
    Sheldon, Seth H.
    Espinosa, Raul E.
    Wood, Christopher P.
    Felmlee, Joel P.
    Cha, Yong-Mei
    Asirvatham, Samuel J.
    Dalzell, Connie
    Acker, Nancy
    Watson, Robert E., Jr.
    Friedman, Paul A.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2014, 37 (10): : 1284 - 1290
  • [38] Monocenter feasibility study of the MRI compatibility of the Evia pacemaker in combination with Safio S pacemaker lead
    Wollmann, Christian G.
    Steiner, Erich
    Vock, Paul
    Ndikung, Bonaventure
    Mayr, Harald
    JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2012, 14
  • [39] Monocenter feasibility study of the MRI compatibility of the Evia pacemaker in combination with Safio S pacemaker lead
    Christian G Wollmann
    Erich Steiner
    Paul Vock
    Bonaventure Ndikung
    Harald Mayr
    Journal of Cardiovascular Magnetic Resonance, 14
  • [40] Safety and Outcomes of Magnetic Resonance Imaging in Patients With Abandoned Pacemaker and Defibrillator Leads
    Higgins, John V.
    Gard, Joseph J.
    Espinosa, Raul E.
    Sheldon, Seth H.
    Asirvatham, Samuel I.
    Acker, Nancy
    Dalzell, Connie
    Cha, Yong-Mei
    Felmlee, Joel P.
    Wood, Christopher P.
    Watson, Robert E.
    Friedman, Paul A.
    CIRCULATION, 2013, 128 (22)