Pacemaker Lead Tip Heating in Abandoned and Pacemaker-Attached Leads at 1.5 Tesla MRI

被引:113
|
作者
Langman, Deborah A. [1 ,2 ]
Goldberg, Ira B. [3 ]
Finn, J. Paul [1 ,2 ]
Ennis, Daniel B. [1 ,2 ]
机构
[1] Univ Calif Los Angeles, David Geffen Sch Med, Dept Radiol Sci, Los Angeles, CA 90095 USA
[2] Univ Calif Los Angeles, Biomed Phys Interdept Program, Los Angeles, CA USA
[3] Univ Calif Los Angeles, Dept Elect Engn, Los Angeles, CA 90024 USA
关键词
magnetic resonance imaging; medical device safety; pacemaker; RF-induced heating; abandoned leads; IMPLANTABLE-CARDIOVERTER-DEFIBRILLATORS; MAGNETIC-RESONANCE; CARDIAC-PACEMAKERS; MANAGE COMPLICATIONS; IN-VIVO; SAFETY; EXTRACTION; ISSUES; RECOGNIZE; 1.5-TESLA;
D O I
10.1002/jmri.22463
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the risk of RF-induced heating in pacemaker-attached and abandoned leads using in vitro temperature measurements at 1.5 Tesla as a function of lead length. Materials and Methods: Five custom lead lengths, 20-60 cm, were exposed to a uniform magnitude and phase radio-frequency electric field to examine the effect of lead length on pacemaker lead tip heating for pacemaker-attached and abandoned pacemaker leads. Results: Abandoned and pacemaker-attached leads show resonant heating behavior and maximum heating occurs at different lead lengths due to the differences in termination conditions. For clinical lead lengths (40-60 cm) abandoned leads exhibited greater lead tip heating compared with pacemaker-attached leads. Conclusion: Current recommendations for MRI pacemaker safety should highlight the possible increased risk for patients with abandoned leads as compared to pacemaker-attached leads.
引用
收藏
页码:426 / 431
页数:6
相关论文
共 50 条
  • [1] MRI-related heating of pacemaker at 1.5-and 3-Tesla: Evaluation with and without pulse generator attached to leads
    Shellock, FG
    Valencerina, S
    Fischer, L
    CIRCULATION, 2005, 112 (17) : U623 - U623
  • [2] MRI induced heating of pacemaker leads: effect of temperature probe positioning and pacemaker placement on lead tip heating and local SAR
    Mattei, E.
    Calcagnini, G.
    Triventi, M.
    Censi, F.
    Bartolini, P.
    Kainz, W.
    Bassen, H.
    2006 28TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-15, 2006, : 4398 - +
  • [3] Abandoned Pacemaker Leads Are a Potential Risk for Patients Undergoing MRI
    Langman, Deborah A.
    Finn, J. Paul
    Ennis, Daniel B.
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2011, 34 (09): : 1051 - 1053
  • [4] Randomized trial of pacemaker and lead system for safe scanning at 1.5 Tesla
    Gimbel, J. Rod
    Bello, David
    Schmitt, Matthias
    Merkely, Bela
    Schwitter, Juerg
    Hayes, David L.
    Sommer, Torsten
    Schloss, Edward J.
    Chang, Yanping
    Willey, Sarah
    Kanal, Emanuel
    HEART RHYTHM, 2013, 10 (05) : 685 - 691
  • [5] MRI-related Heating at Cardiac Pacemaker Leads in vivo
    Nordbeck, Peter
    Bauer, Wolfgang R.
    Warmuth, Marcus
    Hiller, Karl-Heinz
    Jakob, Peter M.
    Ritter, Oliver
    CIRCULATION, 2009, 120 (18) : S371 - S371
  • [6] The dependence of radiofrequency induced pacemaker lead tip heating on the electrical conductivity of the medium at the lead tip
    Langman, Deborah A.
    Goldberg, Ira B.
    Judy, Jack
    Finn, J. Paul
    Ennis, Daniel B.
    MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (02) : 606 - 613
  • [7] Dislocation of an Infected and Abandoned Pacemaker Lead
    Kostopoulou, Anna G.
    Kyrzopoulos, Stamatios D.
    Livanis, Efthimios G.
    JOURNAL OF INVASIVE CARDIOLOGY, 2012, 24 (11): : E297 - E299
  • [8] MRI-induced lead heating of an MRI conditional pacemaker system
    Mouchawar, G.
    Sison, S.
    Chen, S.
    Min, X.
    Chen, J.
    Nyenhuis, J.
    Williamson, R.
    EUROPEAN HEART JOURNAL, 2015, 36 : 350 - 350
  • [9] Asystole caused by vegetation and abscess of right ventricle attached to a tip of pacemaker lead
    Yamaoka, Masaya
    Noda, Takashi
    Nagasawa, Hikaru
    Naganuma, Masaki
    Otsubo, Ryoichi
    Okamura, Hideo
    Ishibashi-Ueda, Hatsue
    Kamakura, Shiro
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2008, 31 (08): : 1083 - 1084
  • [10] Complications due to abandoned noninfected pacemaker leads
    Böhm, A
    Pintér, A
    Duray, G
    Lehoczky, D
    Dudás, G
    Tomcsányi, I
    Préda, I
    PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2001, 24 (12): : 1721 - 1724