Investigation of optimal scanning protocol for X-ray computed tomography polymer gel dosimetry

被引:8
|
作者
Sellakumar, P. [1 ]
Samuel, E. James Jebaseelan [2 ]
Supe, Sanjay S. [3 ]
机构
[1] Bangalore Inst Oncol, Bangalore 560027, Karnataka, India
[2] VIT Univ, Sch Sci & Humanities, Vellore 632014, Tamil Nadu, India
[3] Kidwai Mem Inst Oncol, Dept Radiat Phys, Bangalore 560027, Karnataka, India
关键词
gel dosimetry; X-ray CT; polymer gel dosimetry; image noise; artifacts;
D O I
10.1016/j.nimb.2007.09.022
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
X-ray computed tomography is one of the potential tool used to evaluate the polymer gel dosimeters in three dimensions. The purpose of this study is to investigate the factors which affect the image noise for X-ray CT polymer gel dosimetry. A cylindrical water filled phantom was imaged with single slice Siemens Somatom Emotion CT scanner. The imaging parameters like tube voltage, tube current, slice scan time, slice thickness and reconstruction algorithm were varied independently to study the dependence of noise on each other. Reductions of noise with number of images to be averaged and spatial uniformity of the image were also investigated. Normoxic polymer gel PAGAT was manufactured and irradiated using Siemens Primus linear accelerator. The radiation induced change in CT number was evaluated using X-ray CT scanner. From this study it is clear that image noise is reduced with increase in tube voltage, tube current, slice scan time, slice thickness and also reduced with increasing the number of images averaged. However to reduce the tube load and total scan time, it was concluded that tube voltage of 130 kV, tube current of 200 mA, scan time of 1.5 s, slice thickness of 3 mm for high dose gradient and 5 mm for low dose gradient were optimal scanning protocols for this scanner. Optimum number of images to be averaged was concluded to be 25 for X-ray CT polymer gel dosimetry. Choice of reconstruction algorithm was also critical. From the study it is also clear that CT number increase with imaging tube voltage and shows the energy dependency of polymer gel dosimeter. Hence for evaluation of polymer gel dosimeters with X-ray CT scanner needs the optimization of scanning protocols to reduce the image noise. (C) 2007 Elsevier B.V. All rights reserved.
引用
收藏
页码:395 / 399
页数:5
相关论文
共 50 条
  • [31] X-RAY COMPUTED TOMOGRAPHY
    HORN, E
    ELECTRONICS AND POWER, 1978, 24 (03): : 181 - 181
  • [32] Investigation of X-ray CT dose in normoxic polyacrylamide gel dosimetry
    Jirasek, A.
    Baxter, P.
    Hilts, M.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 1873 - +
  • [33] Experimental evaluation of x-ray acoustic computed tomography for radiotherapy dosimetry applications
    Hickling, Susannah
    Lei, Hao
    Hobson, Maritza
    Leger, Pierre
    Wang, Xueding
    El Naqa, Issam
    MEDICAL PHYSICS, 2017, 44 (02) : 608 - 617
  • [34] Towards quantitative in vivo dosimetry using x-ray acoustic computed tomography
    Sun, Leshan
    Gonzalez, Gilberto
    Pandey, Prabodh Kumar
    Wang, Siqi
    Kim, Kaitlyn
    Limoli, Charles
    Chen, Yong
    Xiang, Liangzhong
    MEDICAL PHYSICS, 2023, 50 (11) : 6894 - 6907
  • [35] Metrological investigation of radiation filtration in x-ray computed tomography
    Kourra, N.
    Warnett, J. A.
    Attridge, A.
    Williams, M. A.
    2017 IEEE INTERNATIONAL INSTRUMENTATION AND MEASUREMENT TECHNOLOGY CONFERENCE (I2MTC), 2017, : 915 - 920
  • [36] Contrast agents for soil investigation with X-ray computed tomography
    Van Loo, Denis
    Bouckaert, Liesbeth
    Leroux, Olivier
    Pauwels, Elin
    Dierick, Manuel
    Van Hoorebeke, Luc
    Cnudde, Veerle
    De Neve, Stefaan
    Sleutel, Steven
    GEODERMA, 2014, 213 : 485 - 491
  • [37] Investigation of the Shape and Detectability of Pores with X-ray Computed Tomography
    Baumgaertner, Benjamin
    Hussein, Juan
    Hausotte, Tino
    JOURNAL OF MANUFACTURING AND MATERIALS PROCESSING, 2023, 7 (03):
  • [38] Normoxic polymer gel dosimetry using less toxic monomer of N-isopropyl acrylamide and X-ray computed tomography for radiation therapy applications
    Ghavami, Seyed-Mostafa
    Mesbahi, Asghar
    Pesianian, Ismaeel
    Shafaee, Abbas
    Aliparasti, Mohammad-Reza
    REPORTS OF PRACTICAL ONCOLOGY AND RADIOTHERAPY, 2010, 15 (06) : 172 - 175
  • [39] Commissioning a Multislice CT Scanner for X-Ray CT Polymer Gel Dosimetry
    Johnston, H.
    Hilts, M.
    Jirasek, A.
    MEDICAL PHYSICS, 2014, 41 (06) : 238 - 238
  • [40] High spatial resolution x-ray luminescence computed tomography and x-ray fluorescence computed tomography
    Dai, Xianjin
    Sivasubramanian, Kathyayini
    Xing, Lei
    MOLECULAR-GUIDED SURGERY: MOLECULES, DEVICES, AND APPLICATIONS V, 2019, 10862