Accuracy of Oxygen Flow Delivered by Compressed-Gas Cylinders in Hospital and Prehospital Emergency Care

被引:4
|
作者
Duprez, Frederic [1 ,2 ,3 ]
Michotte, Jean Bernard [4 ]
Cuvelier, Gregory [3 ]
Legrand, Alexandre [5 ]
Mashayekhi, Sharam [1 ]
Reychler, Gregory [2 ]
机构
[1] Hop Epicura, Serv Soins Intensifs, Site Hornu, Boussu, Belgium
[2] Catholic Univ Louvain, Serv Pneumol, IREC, Pole Pneumol ORL & Dermatol, Brussels, Belgium
[3] Condorcet, Lab Effort & Mouvement, Tournai, Belgium
[4] Western Switzerland Univ Appl Sci, Haute Ecole Sante Vaud HESAV, Filiere Physiotherapie, Lausanne, Switzerland
[5] Univ Mons, Serv Physiol Physiopathol & Readaptat Resp, Mons, Belgium
关键词
oxygen gas cylinders; oxygen therapy; accuracy; hyperoxia; hypoxia; THERAPY; AUDIT;
D O I
10.4187/respcare.05657
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
BACKGROUND: Oxygen cylinders are widely used both in hospital and prehospital care. Excessive or inappropriate F-IO2 may be critical for patients with hypercapnia or hypoxia. Moreover, over-oxygenation could be deleterious in ischemic disorders. Supplemental oxygen from oxygen cylinder should therefore be delivered accurately. The aim of this study was to assess the accuracy of oxygen flows for oxygen cylinder in hospital and prehospital care. METHODS: A prospective trial was conducted to evaluate accuracy of delivered oxygen flows (2, 4, 6, 9 and 12 L/min) for different oxygen cylinder ready for use in different hospital departments. Delivered flows were analyzed randomly using a calibrated thermal mass flow meter. Two types of oxygen cylinder were evaluated: 78 oxygen cylinder with a single-stage regulator and 70 oxygen cylinder with a dual-stage regulator. Delivered flows were compared to the required oxygen flow. The residual pressure value for each oxygen cylinder was considered. A coefficient of variation was calculated to compare the variability of the delivered flow between the two types of oxygen cylinder. RESULTS: The median values of delivered flows were all >= 100% of the required flow for single stage (range 100-109%) and <100% of required flow for dual stage (range 95-97%). The median values of the delivered flow differed between single and dual stage. It was found that single stage is significantly higher than dual stage (P = .01). At low flow, the dispersion of the measures for single stage was higher than with a high oxygen flow. Delivered flow differences were also found between low and high residual pressures, but only with single stage (P = .02). The residual pressure for both oxygen cylinders (no. = 148) ranged from 73 to 2,900 pounds per square inch, and no significant difference was observed between the 2 types (P = .86). The calculated coefficient of variation ranged from 7% (+/- 1%) for dual stage to 8% (+/- 2%) for single stage. CONCLUSIONS: This study shows good accuracy of oxygen flow delivered via oxygen cylinders. This accuracy was higher with dual stage. Single stage was also accurate, however, at low flow this accuracy is slightly less. Moreover, with single stage, when residual pressure decreases, the median value of delivered flow decreased.
引用
收藏
页码:332 / 338
页数:7
相关论文
共 50 条
  • [1] Determination of strength parameters of fibre reinforced compressed-gas cylinders.
    Bonten, C
    Uebbing, M
    [J]. MATERIALPRUFUNG, 1999, 41 (04): : 124 - 127
  • [2] Accuracy between prehospital and hospital diagnosis in helicopter emergency medical services and its consequences for trauma care
    Mueller, Martin
    Hautz, Wolf
    Louma, Yves
    Knapp, Juergen
    Schnueriger, Beat
    Simmen, Hans-Peter
    Pietsch, Urs
    Jakob, Dominik A.
    [J]. EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2024,
  • [3] The prehospital and hospital costs of emergency care for frequent ED patients
    Solberg, Robert G.
    Edwards, Brandy L.
    Chidester, Jeffrey P.
    Perina, Debra G.
    Brady, William J.
    Williams, Michael D.
    [J]. AMERICAN JOURNAL OF EMERGENCY MEDICINE, 2016, 34 (03): : 459 - 463
  • [4] Evaluation of prehospital emergency care in the field and during the ambulance drive to the hospital
    Soysal, S
    Karcioglu, O
    Topacoglu, H
    Yenal, S
    Koparan, H
    Yaman, O
    [J]. ADVANCES IN THERAPY, 2005, 22 (01) : 44 - 48
  • [5] Evaluation of prehospital emergency care in the field and during the ambulance drive to the hospital
    Suna Soysal
    Ozgur Karcioglu
    Hakan Topacoglu
    Sinan Yenal
    Harun Koparan
    Orhan Yaman
    [J]. Advances in Therapy, 2005, 22 : 44 - 48
  • [6] A Comparison of Care Delivered in Hospital-based and Freestanding Emergency Departments
    Pines, Jesse M.
    Zocchi, Mark S.
    Black, Bernard S.
    [J]. ACADEMIC EMERGENCY MEDICINE, 2018, 25 (05) : 538 - 550
  • [7] Accuracy of prehospital point-of-care lactate in early in-hospital mortality
    Martin-Rodriguez, Francisco
    Lopez-Izquierdo, Raul
    Medina-Lozano, Elena
    Ortega Rabbione, Guillermo
    del Pozo Vegas, Carlos
    Carbajosa Rodriguez, Virginia
    Castro Villamor, Miguel Angel
    Sanchez-Soberon, Irene
    Sanz-Garcia, Ancor
    [J]. EUROPEAN JOURNAL OF CLINICAL INVESTIGATION, 2020, 50 (12)
  • [8] HYGROMETRY OF GAS DELIVERED WITH HIGH FLOW OXYGEN HUMIDIFIERS: A BENCH STUDY
    Bouchard, P. -A.
    Bojmehrani, A.
    Bouchard, C.
    Lellouche, F.
    [J]. INTENSIVE CARE MEDICINE, 2011, 37 : S190 - S190
  • [9] EMERGENCY CARE OF THE OBESE AND EXTREMELY OBESE (PREHOSPITAL AND HOSPITAL) THROUGH THE EYE OF THE MEDICAL TEAM
    Koedbaum, Z.
    Tolnai, B.
    Lelovics, Z.
    Mate, A.
    Varge, C.
    [J]. INTENSIVE CARE MEDICINE, 2011, 37 : S69 - S69
  • [10] Concentration verification of ethanol/nitrogen compressed gas cylinders prior to use for periodic determinations of accuracy in California
    Matthias, DJ
    Harvey, DC
    DeFraga, DE
    [J]. JOURNAL OF ANALYTICAL TOXICOLOGY, 2001, 25 (03) : 215 - 218