Real-Time Functional MRI Using Pseudo-Continuous Arterial Spin Labeling

被引:9
|
作者
Hernandez-Garcia, Luis [1 ,2 ]
Jahanian, Hesamoddin [2 ]
Greenwald, Mark K. [3 ]
Zubieta, Jon-Kar [4 ]
Peltier, Scott J. [1 ,2 ]
机构
[1] Univ Michigan, Funct MRI Lab, Ann Arbor, MI 48105 USA
[2] Univ Michigan, Dept Biomed Engn, Ann Arbor, MI 48105 USA
[3] Wayne State Univ, Sch Med, Dept Psychiat & Neurosci, Detroit, MI USA
[4] Univ Michigan, Dept Psychiat, Ann Arbor, MI 48105 USA
关键词
arterial spin labeling; real time; fMRI; PERFUSION-BASED FMRI; CEREBRAL-BLOOD-FLOW; BRAIN ACTIVATION; DESIGN; BOLD; SENSITIVITY; MODEL; QUANTIFICATION; OPTIMIZATION; FEASIBILITY;
D O I
10.1002/mrm.22922
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The first implementation of real-time acquisition and analysis of arterial spin labeling-based functional magnetic resonance imaging time series is presented in this article. The implementation uses a pseudo-continuous labeling scheme followed by a spiral k-space acquisition trajectory. Real-time reconstruction of the images, preprocessing, and regression analysis of the functional magnetic resonance imaging data were implemented on a laptop computer interfaced with the MRI scanner. The method allows the user to track the current raw data, subtraction images, and the cumulative t-statistic map overlaid on a cumulative subtraction image. The user is also able to track the time course of individual time courses and interactively selects a region of interest as a nuisance covariate. The pulse sequence allows the user to adjust acquisition and labeling parameters while observing their effect on the image within two successive pulse repetition times. This method is demonstrated by two functional imaging experiments: a simultaneous finger-tapping and visual stimulation paradigm, and a bimanual finger-tapping task. Magn Reson Med 65:1570-1577, 2011. (C)2011 Wiley-Liss, Inc.
引用
收藏
页码:1570 / 1577
页数:8
相关论文
共 50 条
  • [21] Hypoperfusion in nucleus accumbens in chronic migraine using 3D pseudo-continuous arterial spin labeling imaging MRI
    Mengqi Liu
    Yijie Sun
    Xin Li
    Zhiye Chen
    The Journal of Headache and Pain, 2022, 23
  • [22] Hypoperfusion in nucleus accumbens in chronic migraine using 3D pseudo-continuous arterial spin labeling imaging MRI
    Liu, Mengqi
    Sun, Yijie
    Li, Xin
    Chen, Zhiye
    JOURNAL OF HEADACHE AND PAIN, 2022, 23 (01):
  • [23] Spatial-temporal perfusion patterns of the human liver assessed by pseudo-continuous arterial spin labeling MRI
    Martirosian, Petros
    Pohmann, Rolf
    Schraml, Christina
    Schwartz, Martin
    Kuestner, Thomas
    Schwenzer, Nina Franziska
    Scheffler, Klaus
    Nikolaou, Konstantin
    Schick, Fritz
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2019, 29 (02): : 173 - 183
  • [24] In vivo visualization of the PICA perfusion territory with super-selective pseudo-continuous arterial spin labeling MRI
    Hartkamp, Nolan S.
    De Cocker, Laurens J.
    Helle, Michael
    van Osch, Matthias J. P.
    Kappelle, L. Jaap
    Bokkers, Reinoud P. H.
    Hendrikse, Jeroen
    NEUROIMAGE, 2013, 83 : 58 - 65
  • [25] Selective multivessel labeling approach for perfusion territory imaging in pseudo-continuous arterial spin labeling
    Helle, Michael
    Ruefer, Susanne
    van Osch, Matthias J. P.
    Jansen, Olav
    Norris, David G.
    MAGNETIC RESONANCE IN MEDICINE, 2012, 68 (01) : 214 - 219
  • [26] Measurement of brain perfusion in newborns: Pulsed arterial spin labeling (PASL) versus pseudo-continuous arterial spin labeling (pCASL)
    Boudes, Elodie
    Gilbert, Guillaume
    Leppert, Ilana Ruth
    Tan, Xianming
    Pike, G. Bruce
    Saint-Martin, Christine
    Wintermark, Pia
    NEUROIMAGE-CLINICAL, 2014, 6 : 126 - 133
  • [27] Time-encoded pseudo-continuous arterial spin labeling: Increasing SNR in ASL dynamic angiography
    Woods, Joseph G.
    Schauman, S. Sophie
    Chiew, Mark
    Chappell, Michael A.
    Okell, Thomas W.
    MAGNETIC RESONANCE IN MEDICINE, 2023, 89 (04) : 1323 - 1341
  • [28] On the application of pseudo-continuous arterial spin labeled MRI for pulmonary perfusion imaging
    Greer, Joshua S.
    Wang, Yiming
    Udayakumar, Durga
    Hussain, Tarique
    Madhuranthakam, Ananth J.
    MAGNETIC RESONANCE IMAGING, 2023, 104 : 80 - 87
  • [29] Off-resonance correction for pseudo-continuous arterial spin labeling using the optimized encoding scheme
    Berry, Eleanor S. K.
    Jezzard, Peter
    Okell, Thomas W.
    NEUROIMAGE, 2019, 199 : 304 - 312
  • [30] Reliability and Reproducibility of Hadamard Encoded Pseudo-Continuous Arterial Spin Labeling in Healthy Elderly
    Neumann, Katja
    Schidlowski, Martin
    Guenther, Matthias
    Stoecker, Tony
    Duezel, Emrah
    FRONTIERS IN NEUROSCIENCE, 2021, 15