Electrical impedance scanning for melanoma diagnosis: A validation study

被引:31
|
作者
Har-Shai, Y
Glickman, YA
Siller, G
McLeod, R
Topaz, M
Howe, C
Ginzburg, A
Zamir, B
Filo, O
Kenan, G
Ullmann, Y
机构
[1] Linn Med Ctr, Dept Plast Surg, TransScan Med Ltd, IL-35152 Haifa, Israel
[2] Princess Alexandra Hosp, Dept Dermatol & Surg, Brisbane, Qld, Australia
关键词
D O I
10.1097/01.prs.0000176258.52201.22
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: A multicenter study was conducted to test the ability of electrical impedance scanning to differentiate between benign and malignant skin lesions. The performance of a dual electrical impedance scanning/image analysis device was also assessed. Methods: Electrical impedance scanning measurements of 449 preoperative lesions found on 382 patients and including 53 melanomas from the trunk and extremities were performed. Results were correlated with histopathologic findings. In addition, ABCD parameters for the lesions were automatically calculated by the system. Results: Electrical impedance scanning detected melanomas of the trunk and extremities with 91 percent sensitivity and 64 percent specificity. Moreover, sensitivity of ' electrical impedance scanning was increased to 100 percent for in situ and thin melanomas of smaller size (n = 27). Visual examination identified as malignant only 67 percent of these early tumors (p = 0.002). Clinical examination detected 96 percent of the larger or thicker melanomas (n = 26), whereas electrical impedance scanning detected only 81 percent of them. Combined electrical impedance scanning and image analysis detected 100 percent of the melanomas, independent of their thickness,. and with no significant decrease of specificity. Because of electrical differences between the head/neck and the rest of the body, the assessed electrical impedance scanning parameters were not adequate for the diagnosis of melanomas from the head and neck. Conclusions: A validation study proved the value of electrical impedance scanning as a noninvasive technique for detection of melanoma lesions of the trunk and extremities, specifically, of in situ and thin type. In addition, image analysis was shown to be a valuable, complementary procedure. New parameters should be designed to optimize the performance of electrical impedance scanning for melanomas of the head and neck.
引用
收藏
页码:782 / 790
页数:9
相关论文
共 50 条
  • [11] Electrical impedance scanning of thyroid nodules prior to thyroid surgery: A prospective study
    A. Stojadinovic
    S. I. Fields
    C. D. Shriver
    T. Peretz
    H. R. Freund
    A. Nissan
    Annals of Surgical Oncology, 2004, 11 : S64 - S64
  • [12] Electrical impedance scanning of thyroid nodules before thyroid surgery: A prospective study
    Stojadinovic, A
    Fields, SI
    Shriver, CD
    Lenington, S
    Ginor, R
    Peoples, GE
    Burch, HB
    Peretz, T
    Freund, HR
    Nissan, A
    ANNALS OF SURGICAL ONCOLOGY, 2005, 12 (02) : 152 - 160
  • [13] Electrical Impedance Scanning of Thyroid Nodules Before Thyroid Surgery: A Prospective Study
    Alexander Stojadinovic
    Scott I. Fields
    Craig D. Shriver
    Sarah Lenington
    Ron Ginor
    George E. Peoples
    Henry B. Burch
    Tamar Peretz
    Herbert R. Freund
    Aviram Nissan
    Annals of Surgical Oncology, 2005, 12 : 152 - 160
  • [14] Electrical impedance scanning of thyroid nodules prior to thyroid surgery: A prospective study
    Stojadinovic, A
    Fields, SI
    Shriver, CD
    Peretz, T
    Freund, HR
    Nissan, A
    ANNALS OF SURGICAL ONCOLOGY, 2004, 11 (02) : S64 - S64
  • [15] ELECTRICAL IMPEDANCE MYOGRAPHY IN THE DIAGNOSIS OF RADICULOPATHY
    Spieker, Andrew J.
    Narayanaswami, Pushpa
    Fleming, Laura
    Keel, John C.
    Muzin, Stefan C.
    Rutkove, Seward B.
    MUSCLE & NERVE, 2013, 48 (05) : 800 - 805
  • [16] Electrical impedance spectroscopy and diagnosis of tendinitis
    Yoon, Kisung
    Lee, Kyeong Woo
    Kim, Sang Beom
    Han, Tai Ryoon
    Jung, Dong Keun
    Roh, Mee Sook
    Lee, Jong Hwa
    PHYSIOLOGICAL MEASUREMENT, 2010, 31 (02) : 171 - 182
  • [17] Electrical Impedance Analysis for Lung Cancer: A Prospective, Multicenter, Blind Validation Study
    Yang, Dawei
    Gu, Chuanjia
    Gu, Ye
    Zhang, Xiaodong
    Ge, Di
    Zhang, Yong
    Wang, Ningfang
    Zheng, Xiaoxuan
    Wang, Hao
    Yang, Li
    Chen, Saihua
    Xie, Pengfei
    Chen, Deng
    Yu, Jinming
    Sun, Jiayuan
    Bai, Chunxue
    FRONTIERS IN ONCOLOGY, 2022, 12
  • [18] A validation study of stroke volume measurement by means of electrical impedance tomography.
    Janse, A
    Noordegraaf, AV
    Marcus, JT
    Heethaar, RM
    Postmus, PE
    Faes, TJC
    de Vries, PMJM
    PROCEEDINGS OF THE 18TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 18, PTS 1-5, 1997, 18 : 772 - 773
  • [19] A systematic review of elastography, electrical impedance scanning, and digital infrared thermography for breast cancer screening and diagnosis
    Thomas D. Vreugdenburg
    Cameron D. Willis
    Linda Mundy
    Janet E. Hiller
    Breast Cancer Research and Treatment, 2013, 137 : 665 - 676
  • [20] A systematic review of elastography, electrical impedance scanning, and digital infrared thermography for breast cancer screening and diagnosis
    Vreugdenburg, Thomas D.
    Willis, Cameron D.
    Mundy, Linda
    Hiller, Janet E.
    BREAST CANCER RESEARCH AND TREATMENT, 2013, 137 (03) : 665 - 676