Measurement of the thermal conductivity of polyacrylamide tissue-equivalent material

被引:42
|
作者
Davidson, SRH
Sherar, MD
机构
[1] Princess Margaret Hosp, Ontario Canc Inst, Univ Hlth Network, Div Med Phys, Toronto, ON M5G 2M9, Canada
[2] Univ Toronto, Dept Med Biophys, Toronto, ON, Canada
[3] Univ Toronto, Dept Radiat Oncol, Toronto, ON, Canada
关键词
polyacrylamide; phantoms; thermal conductivity; hyperthermia; thermal therapy;
D O I
10.1080/02656730310001607995
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
The purpose of this work was to measure the thermal conductivity of polyacrylamide (PAG) and compare it with previously reported values. Polyacrylamide phantoms play an important role in the development of hyperthermia and high-temperature thermal therapies based on electromagnetic ( EM) radiation by providing a material that mimics the electrical and thermal properties of human tissue. The thermal properties of PAG have, up until now, not been thoroughly investigated and at least two significantly different values have been published. In this study, the thermal conductivity of polyacrylamide was measured from the steady state temperature drop across samples exposed to a known heat flux. The measured conductivity was 0.56 +/- 0.047W m(-1) degreesC(-1). To validate the correct set of thermal properties for polyacrylamide, simple heating experiments were performed in a PAG phantom and then simulated using a finite element numerical model that incorporated the measured thermal conductivity along with literature values for specific heat and density. Temperature predictions from the model agreed with average temperatures measured in the phantom to within 1 SD of the measured temperatures.
引用
收藏
页码:551 / 562
页数:12
相关论文
共 50 条
  • [41] TISSUE-EQUIVALENT MATERIALS FOR HUMAN PHANTOMS AND DOSIMETRY
    FRIGERIO, NA
    LAVOY, LL
    STOWE, RS
    RADIATION RESEARCH, 1976, 67 (03) : 615 - 615
  • [42] A SIMPLE TISSUE-EQUIVALENT WEDGE FILTRATION TECHNIQUE
    KRAMER, RW
    VETERINARY RADIOLOGY & ULTRASOUND, 1993, 34 (01) : 18 - 19
  • [43] Studies of Positron Range in Tissue-Equivalent Materials
    Alva-Sanchez, Hector
    Quintana-Bautista, Christian
    Martinez-Davalos, Arnulfo
    Angel Avila-Rodriguez, Miguel
    Rodriguez-Villafuerte, Mercedes
    MEDICAL PHYSICS: Fourteenth Mexican Symposium on Medical Physics, 2016, 1747
  • [44] Optimized design of tissue-equivalent proportional counter
    Zhang, Wei-Hua
    Wang, Zhi-Qiang
    Xiao, Xue-Fu
    Liu, Yi-Na
    Li, Chun-Juan
    Luo, Hai-Long
    Yuanzineng Kexue Jishu/Atomic Energy Science and Technology, 2014, 48 (05): : 930 - 933
  • [45] HEAT DEFECT IN TISSUE-EQUIVALENT RADIATION CALORIMETERS
    BEWLEY, DK
    SAKATA, S
    PAGE, BC
    MCCULLOUGH, EC
    PHYSICS IN MEDICINE AND BIOLOGY, 1972, 17 (01): : 95 - +
  • [46] Optimising tissue-equivalent materials for proton therapy
    Cook, H.
    Simard, M.
    Niemann, N.
    Gillies, C.
    Palmans, H.
    Hussein, M.
    Bouchard, H.
    Royle, G.
    Lourenco, A.
    RADIOTHERAPY AND ONCOLOGY, 2022, 170 : S827 - S828
  • [47] Operation and application of tissue-equivalent proportional counters
    Gerdung, S
    Pihet, P
    Grindborg, JE
    Roos, H
    Schrewe, UJ
    Schuhmacher, H
    RADIATION PROTECTION DOSIMETRY, 1995, 61 (04) : 381 - 404
  • [48] TISSUE-EQUIVALENT DOSIMETERS FOR PION THERAPY BEAMS
    SHORTT, KR
    PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (03): : 279 - 281
  • [49] Measurement of of the first Townsend ionization coefficient in a methane-based tissue-equivalent gas
    Petri, A. R.
    Goncalves, J. A. C.
    Mangiarotti, A.
    Botelho, S.
    Bueno, C. C.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 849 : 31 - 40
  • [50] STATISTICAL MODELING OF THE SLOWING DOWN OF ELECTRONS WITH ENERGIES BELOW 30 KEV IN A TISSUE-EQUIVALENT MATERIAL
    PITKEVICH, VA
    VIDENSKII, VG
    DUBA, VV
    SOVIET ATOMIC ENERGY, 1982, 52 (03): : 195 - 199