Dose calculation of dynamic trajectory radiotherapy using Monte Carlo

被引:15
|
作者
Manser, Peter [1 ,2 ,3 ]
Frauchiger, Daniel [1 ,2 ,3 ]
Frei, Daniel [1 ,2 ,3 ]
Volken, Werner [1 ,2 ,3 ]
Terribilini, Dario [1 ,2 ,3 ]
Fix, Michael K. [1 ,2 ,3 ]
机构
[1] Bern Univ Hosp, Inselspital, Div Med Radiat Phys, Freiburgstr 15, CH-3010 Bern, Switzerland
[2] Bern Univ Hosp, Dept Radiat Oncol, Inselspital, Bern, Switzerland
[3] Univ Bern, Bern, Switzerland
来源
ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK | 2019年 / 29卷 / 01期
关键词
Dynamic trajectories; Monte Carlo; Dose calculation; Verification; RADIATION-THERAPY; MODULATED ARC; NECK CANCER; DELIVERY; IMRT; IRRADIATION; VALIDATION; RAPIDARC; QA;
D O I
10.1016/j.zemedi.2018.03.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Using volumetric modulated arc therapy (VMAT) delivery technique gantry position, multi-leaf collimator (MLC) as well as dose rate change dynamically during the application. However, additional components can be dynamically altered throughout the dose delivery such as the collimator or the couch. Thus, the degrees of freedom increase allowing almost arbitrary dynamic trajectories for the beam. While the dose delivery of such dynamic trajectories for linear accelerators is technically possible, there is currently no dose calculation and validation tool available. Thus, the aim of this work is to develop a dose calculation and verification tool for dynamic trajectories using Monte Carlo (MC) methods. Methods: The dose calculation for dynamic trajectories is implemented in the previously developed Swiss Monte Carlo Plan (SMCP). SMCP interfaces the treatment planning system Eclipse with a MC dose calculation algorithm and is already able to handle dynamic MLC and gantry rotations. Hence, the additional dynamic components, namely the collimator and the couch, are described similarly to the dynamic MLC by defining data pairs of positions of the dynamic component and the corresponding MU-fractions. For validation purposes, measurements are performed with the Delta4 phantom and film measurements using the developer mode on a TrueBeam linear accelerator. These measured dose distributions are then compared with the corresponding calculations using SMCP. First, simple academic cases applying one-dimensional movements are investigated and second, more complex dynamic trajectories with several simultaneously moving components are compared considering academic cases as well as a clinically motivated prostate case. Results: The dose calculation for dynamic trajectories is successfully implemented into SMCP. The comparisons between the measured and calculated dose distributions for the simple as well as for the more complex situations show an agreement which is generally within 3% of the maximum dose or 3 mm. The required computation time for the dose calculation remains the same when the additional dynamic moving components are included. Conclusion: The results obtained for the dose comparisons for simple and complex situations suggest that the extended SMCP is an accurate dose calculation and efficient verification tool for dynamic trajectory radiotherapy. This work was supported by Varian Medical Systems.
引用
收藏
页码:31 / 38
页数:8
相关论文
共 50 条
  • [31] Biological dose calculation with Monte Carlo physics simulation for heavy-ion radiotherapy
    Kase, Yuki
    Kanematsu, Nobuyuki
    Kanai, Tatsuaki
    Matsufuji, Naruhiro
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2006, 51 (24): : N467 - N475
  • [32] Development of a GPU-Based Monte Carlo Dose Calculation Package for Cancer Radiotherapy
    Jia, X.
    Gu, X.
    Graves, Y.
    Jiang, S.
    [J]. MEDICAL PHYSICS, 2011, 38 (06)
  • [33] Voxel size effects on Monte Carlo dose calculation for clinical radiotherapy electron beams
    Husson, F
    Guillerminet, C
    [J]. RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S153 - S153
  • [34] Monte Carlo Evaluation of Superposition-Algorithm Skin Dose Calculation for Brain Radiotherapy
    Li, J.
    Xiao, Y.
    Yu, Y.
    Shi, V.
    [J]. MEDICAL PHYSICS, 2011, 38 (06) : 3653 - +
  • [35] Development of a GPU-Based Monte Carlo Dose Calculation Package for Proton Radiotherapy
    Jia, X.
    Schuemann, J.
    Paganetti, H.
    Jiang, S.
    [J]. MEDICAL PHYSICS, 2012, 39 (06) : 3945 - 3945
  • [36] Implementation of Monte Carlo Dose Calculation for a Multileaf Collimator Equipped Robotic Radiotherapy System
    Li, J. S.
    Fan, J.
    Ma, C-M
    [J]. MEDICAL PHYSICS, 2015, 42 (06) : 3442 - 3442
  • [37] DOSE CALCULATION FOR PROTON RADIOTHERAPY BASED-ON MONTE CARLO SIMULATION USING ANATOMICAL DETAILED HUMAN MODEL
    Guo, B.
    Liu, Y.
    Shi, C.
    Mavroidis, P.
    Ha, C.
    Xu, G.
    Papanikolaou, N.
    [J]. RADIOTHERAPY AND ONCOLOGY, 2008, 88 : S113 - S114
  • [38] Monte Carlo simulation of dose distribution of radiotherapy
    Liu, YM
    Xue, DY
    Xu, XH
    Chen, Z
    [J]. System Simulation and Scientific Computing, Vols 1 and 2, Proceedings, 2005, : 422 - 426
  • [39] ORANGE: a Monte Carlo dose engine for radiotherapy
    van der Zee, W
    Hogenbirk, A
    van der Marck, SC
    [J]. PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (04): : 625 - 641
  • [40] Monte Carlo dose calculation on deforming anatomy
    Peterhans, Matthias
    Frei, Daniel
    Manser, Peter
    Aguirre, Mauricio Reyes
    Fix, Michael K.
    [J]. ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2011, 21 (02): : 113 - 123