Comparison of different techniques of hemodialysis vascular access flow evaluation: Blood Temperature Monitoring thermodilution and doppler debimetry

被引:2
|
作者
Sacquepee, Mathieu [1 ]
Tivollier, Jean-Michel [2 ]
Doussy, Yves [1 ]
Quirin, Nicolas [2 ]
Valery, Jean-Claude [1 ]
Cantin, Jean-Francois [2 ]
机构
[1] ATIR NC, Noumea 98846, New Caledonia
[2] Ctr Hosp Terr, Serv Nephrol Hemodialise, Noumea 98849, New Caledonia
来源
NEPHROLOGIE & THERAPEUTIQUE | 2012年 / 8卷 / 02期
关键词
Renal dialysis; Thermodilution; Thrombosis prevention; Doppler; Hemodialysis vascular access; RECIRCULATION; VALIDATION; DILUTION;
D O I
10.1016/j.nephro.2011.09.001
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background. - Access blood flow measurements are considered useful indicators for thrombosis prevention. It was the purpose of this study to compare measurements of access blood flow by two different techniques: duplex doppler and BTM thermodilution. Methods. - Patients included must be on chronic hemodialysis on arterioveinous vascular access. They must be in a unit fit with hemodialysis generator equipped with BTM tool. The measurements of access blood flow were made during the first hour of the hemodialysis session. A measurement with each technique was performed for each patient. Results. - Fifteen patients were included: seven men and eight women, average age 60.8 +/- 9.2 years, average weight 76 16 kg, duration on hemodialysis therapy 6.6 +/- 6.1 years. Access blood flow was native fistula (14 patients) and a prothetic access (one patient). Average access blood flow was 1088 +/- 586 mL/mn (doppler) and 1094 +/- 570 mL/mn (BTM). Comparison of access flows obtained by the BTM and doppler techniques showed a strong linear relationship. The average time to perform a measure was six minutes for the doppler technique and five minutes for the BTM technique. No adverse effect was observed in our study. Conclusion. - Our study shows a strong correlation between the two techniques (doppler and BTM) for the measurement of hemodialysis access blood flow. The BTM access blood flow measurement technique is fast, economic and made during the hemoclialysis session by the nurse. (C) 2011 Association Societe de nephrologie. Published by Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:96 / 100
页数:5
相关论文
共 50 条
  • [21] Variable flow Doppler for hemodialysis access evaluation: Theory and clinical feasibility
    Weitzel, WF
    Rubin, JM
    Swartz, RD
    Woltmann, DJ
    Messana, JM
    [J]. ASAIO JOURNAL, 2000, 46 (01) : 65 - 69
  • [22] Analysis of variable flow Doppler hemodialysis access flow measurements and comparison with ultrasound dilution
    Weitzel, WF
    Rubin, JM
    Leavey, SF
    Swartz, RD
    Dhingra, RK
    Messana, JM
    [J]. AMERICAN JOURNAL OF KIDNEY DISEASES, 2001, 38 (05) : 935 - 940
  • [23] Vascular access blood flow monitoring reduces access morbidity and costs
    McCarley, P
    Wingard, RL
    Shyr, Y
    Pettus, W
    Hakim, RM
    Ikizler, TA
    [J]. KIDNEY INTERNATIONAL, 2001, 60 (03) : 1164 - 1172
  • [24] The relation between vascular access flow and different types of vascular access with systemic hemodynamics in hemodialysis patients
    Wijnen, E
    Keuter, XH
    Planken, NR
    van der Sande, FM
    Tordoir, JH
    Leunissen, KM
    Kooman, JP
    [J]. ARTIFICIAL ORGANS, 2005, 29 (12) : 960 - 964
  • [25] Blood Flow in Idealized Vascular Access for Hemodialysis: A Review of Computational Studies
    Bogdan Ene-Iordache
    Andrea Remuzzi
    [J]. Cardiovascular Engineering and Technology, 2017, 8 : 295 - 312
  • [26] Hemodialysis arteriovenous access: Detection of stenosis and response to treatment by vascular access blood flow
    Schwab, SJ
    Oliver, MJ
    Suhocki, P
    McCann, R
    [J]. KIDNEY INTERNATIONAL, 2001, 59 (01) : 358 - 362
  • [27] Blood Flow in Idealized Vascular Access for Hemodialysis: A Review of Computational Studies
    Ene-Iordache, Bogdan
    Remuzzi, Andrea
    [J]. CARDIOVASCULAR ENGINEERING AND TECHNOLOGY, 2017, 8 (03) : 295 - 312
  • [28] COMPARISON OF HEMODIALYSIS BLOOD ACCESS FLOW RATES USING 3 ON-LINE TECHNIQUES AND ULTRASOUND DILUTION
    Frank, Le Roy
    Melanie, Hanoy
    Dominique, Bertrand
    Michel, Godin
    [J]. NEPHROLOGY DIALYSIS TRANSPLANTATION, 2012, 27 : 262 - 263
  • [29] A comparison of methods for the measurement of hemodialysis access recirculation and access blood flow rate
    Lindsay, RM
    Bradfield, E
    Rothera, C
    Kianfar, C
    Malek, P
    Blake, PG
    [J]. ASAIO JOURNAL, 1998, 44 (01) : 62 - 67
  • [30] Measurement of hemodialysis vascular access flow using extracorporeal temperature gradients
    Wijnen, E.
    van der Sande, F. M.
    Kooman, J. P.
    de Graaf, T.
    Tordoir, J. H. M.
    Leunissen, K. M. L.
    Schneditz, D.
    [J]. KIDNEY INTERNATIONAL, 2007, 72 (06) : 736 - 741