Comparison of different techniques of hemodialysis vascular access flow evaluation: Blood Temperature Monitoring thermodilution and doppler debimetry

被引:2
|
作者
Sacquepee, Mathieu [1 ]
Tivollier, Jean-Michel [2 ]
Doussy, Yves [1 ]
Quirin, Nicolas [2 ]
Valery, Jean-Claude [1 ]
Cantin, Jean-Francois [2 ]
机构
[1] ATIR NC, Noumea 98846, New Caledonia
[2] Ctr Hosp Terr, Serv Nephrol Hemodialise, Noumea 98849, New Caledonia
来源
NEPHROLOGIE & THERAPEUTIQUE | 2012年 / 8卷 / 02期
关键词
Renal dialysis; Thermodilution; Thrombosis prevention; Doppler; Hemodialysis vascular access; RECIRCULATION; VALIDATION; DILUTION;
D O I
10.1016/j.nephro.2011.09.001
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Background. - Access blood flow measurements are considered useful indicators for thrombosis prevention. It was the purpose of this study to compare measurements of access blood flow by two different techniques: duplex doppler and BTM thermodilution. Methods. - Patients included must be on chronic hemodialysis on arterioveinous vascular access. They must be in a unit fit with hemodialysis generator equipped with BTM tool. The measurements of access blood flow were made during the first hour of the hemodialysis session. A measurement with each technique was performed for each patient. Results. - Fifteen patients were included: seven men and eight women, average age 60.8 +/- 9.2 years, average weight 76 16 kg, duration on hemodialysis therapy 6.6 +/- 6.1 years. Access blood flow was native fistula (14 patients) and a prothetic access (one patient). Average access blood flow was 1088 +/- 586 mL/mn (doppler) and 1094 +/- 570 mL/mn (BTM). Comparison of access flows obtained by the BTM and doppler techniques showed a strong linear relationship. The average time to perform a measure was six minutes for the doppler technique and five minutes for the BTM technique. No adverse effect was observed in our study. Conclusion. - Our study shows a strong correlation between the two techniques (doppler and BTM) for the measurement of hemodialysis access blood flow. The BTM access blood flow measurement technique is fast, economic and made during the hemoclialysis session by the nurse. (C) 2011 Association Societe de nephrologie. Published by Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:96 / 100
页数:5
相关论文
共 50 条
  • [1] Comparison of different techniques of hemodialysis vascular access flow evaluation
    Lopot, F
    Nejedly, B
    Sulková, S
    Bláha, J
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2003, 26 (12): : 1056 - 1063
  • [2] Blood flow measurements during hemodialysis vascular access interventions - Catheter-based thermodilution or Doppler ultrasound?
    Heerwagen, Soren T.
    Hansen, Marc A.
    Schroeder, Torben V.
    Ladefoged, Soren D.
    Lonn, Lars
    [J]. JOURNAL OF VASCULAR ACCESS, 2012, 13 (02): : 145 - 151
  • [3] Comparison of urea recirculation and thermodilution for monitoring of vascular access in patients undergoing hemodialysis
    Vega, Almudena
    Abad, Soraya
    Aragoncillo, Ines
    Galan, Isabel
    Macias, Nicolas
    Cedeno, Santiago
    Santos, Alba
    Garcia, Ana
    Linares, Tania
    Martinez-Villaescusa, Maria
    Lopez-Gomez, Juan M.
    [J]. JOURNAL OF VASCULAR ACCESS, 2018, 19 (03): : 283 - 290
  • [4] A COMPARISON OF DIFFERENT ONLINE TECHNIQUES TO MEASURE VASCULAR ACCESS FLOW IN HEMODIALYSIS PATIENTS
    Eloot, S.
    Dhondt, A.
    Hoeben, H.
    Vanholder, R.
    [J]. INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 2009, 32 (07): : 393 - 393
  • [5] Practical utility of thermodilution versus doppler ultrasound to measure hemodialysis blood access flow
    Fontsere, Nestor
    Mestres, Gaspar
    Barrufet, Marta
    Burrel, Marta
    Vera, Manel
    Arias, Marta
    Masso, Elisabeth
    Cases, Aleix
    Maduell, Francisco
    Campistol, Josep M.
    [J]. NEFROLOGIA, 2013, 33 (03): : 325 - 332
  • [6] COMPARISON OF UREA RECIRCULATION AND THERMODILUTION TO MONITOR VASCULAR ACCESS IN HEMODIALYSIS
    Vega, Almudena
    Abad, Soraya
    Aragoncillo, Ines
    Macias, Nicolas
    Galan, Isabel
    Cedeno, Santiago
    Santos, Alba
    Garcia-Prieto, Ana
    Linares, Tania
    Luno, Jose
    [J]. NEPHROLOGY DIALYSIS TRANSPLANTATION, 2017, 32
  • [7] EFFECTS OF VASCULAR REEFING SURGERY ON EXCESSIVE BLOOD FLOW AT HEMODIALYSIS ACCESS VESSELS UNDER DOPPLER ULTRASONOGRAPHY MONITORING
    Miyake, Katsunori
    Isogai, Naoko
    Kawachi, Jun
    Shimoyama, Rai
    Shibata, Kazuhiko
    Ogino, Hidemitsu
    Kobayashi, Shuzo
    [J]. NEPHROLOGY DIALYSIS TRANSPLANTATION, 2017, 32
  • [8] Use of blood temperature monitor with Twister device for the surveillance of vascular access in maintenance hemodialysis: Comparison with Doppler ultrasonography
    Lee, Jun Young
    Park, Eung Joo
    Chang, Jae Hyun
    Han, Byoung-Geun
    Kim, Gheun-Ho
    [J]. JOURNAL OF VASCULAR ACCESS, 2021, 22 (02): : 218 - 224
  • [9] Comparison between two on-line reversed line position hemodialysis vascular access flow measurement techniques: Saline dilution and thermodilution
    Wijnen, Edwin
    Essers, Stig
    van Meijel, Ger
    Kooman, Jeroen P.
    Tordoir, Jan
    Leunissen, Karel M. L.
    van der Sande, Frank M.
    [J]. ASAIO JOURNAL, 2006, 52 (04) : 410 - 415
  • [10] COLOR DOPPLER SONOGRAPHY OF HEMODIALYSIS VASCULAR ACCESS - COMPARISON WITH ANGIOGRAPHY
    MIDDLETON, WD
    PICUS, DD
    MARX, MV
    MELSON, GL