Study of a noninvasive blood glucose detection model using the near-infrared light based on SA-NARX

被引:12
|
作者
Cheng, Jinxiu [1 ]
Ji, Zhong [1 ,2 ]
Li, Mengze [1 ]
Dai, Juan [1 ]
机构
[1] Chongqing Univ, Coll Biol Engn, Chongqing 400044, Peoples R China
[2] Chongqing Med Elect Engn Technol Ctr, Chongqing 400044, Peoples R China
基金
中国国家自然科学基金;
关键词
Near-infrared; Noninvasive blood glucose measurement; Input variable selection; Sensitivity analysis; NARX; REVERSE IONTOPHORESIS; IN-VIVO; SPECTROSCOPY; TEMPERATURE; TECHNOLOGY; SENSORS;
D O I
10.1016/j.bspc.2019.101694
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Accumulated attempts have been made to develop models for near-infrared noninvasive measurement of human blood glucose concentration. Most of them focus on the relationship between near-infrared absorbance and blood glucose concentration, but do not consider the fluctuation regularity of blood glucose concentration and the influence of environmental factors and human physiological state on near-infrared absorption. In order to improve the performance of prediction model, a hybrid method is proposed in this paper. The nonlinear autoregressive model with exogenous input (NARX) was introduced as prediction model. 7 variables, including 1550 nm near-infrared absorbance, ambient temperature, ambient humidity, systolic pressure, diastolic pressure, pulse rate and body temperature were introduced as initial input variables. The sensitivity analysis (SA) method was employed to select the relative important input variables for NARX model. Based on the result of SA, a robust and accurate NARX model with 4 input variables (1550 nm near-infrared absorbance, systolic pressure, pulse rate and body temperature) was derived. Compared with the back propagation neural network (BPNN) with the same selected 4 input variables and the BPNN with initial 7 input variables, the NARX model developed there showed better prediction performance, of which the root mean square error and correlation coefficients were 0.72 mmol/L and 0.85 respectively for the 10-fold cross validation set. The percentages of the 10-fold cross validation set samples falling in region A and B of the Clarke error grid analysis were 90.27% and 9.73% respectively. These results demonstrate the potential of our model for noninvasive measurement of blood glucose concentration. (C) 2019 Elsevier Ltd. All rights reserved.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] Noninvasive Blood Glucose Detection Using an Improved Sparrow Search Algorithm Combined with an Extreme Learning Machine Based on Near-Infrared Spectroscopy
    Li, Qing-bo
    Wang, Yun-hui
    JOURNAL OF APPLIED SPECTROSCOPY, 2023, 90 (03) : 655 - 660
  • [22] Accuracy of Noninvasive Glucose Sensing Based on Near-Infrared Spectroscopy
    Liu, Jin
    Liu, Rong
    Xu, Kexin
    APPLIED SPECTROSCOPY, 2015, 69 (11) : 1313 - 1318
  • [23] Noninvasive Blood Glucose Detection Using an Improved Sparrow Search Algorithm Combined with an Extreme Learning Machine Based on Near-Infrared Spectroscopy
    Qing-bo Li
    Yun-hui Wang
    Journal of Applied Spectroscopy, 2023, 90 : 655 - 660
  • [24] COMPLETELY NONINVASIVE MEASUREMENTS OF HUMAN BLOOD-GLUCOSE INVIVO USING NEAR-INFRARED WAVES
    NODA, M
    TANIGUCHI, N
    KIMURA, M
    KUZUYA, N
    KANAZAWA, Y
    DIABETOLOGIA, 1992, 35 : A204 - A204
  • [25] Telemed noninvasive glucose blood level measuring by near infrared light
    Piyapan, Siriporn
    Krachaechuang, Mathurada
    Samakit, Tanyawat
    Somboonwong, Tanadul
    Srininrat, Kankanit
    Naiyanertr, Phornphop
    2020 8TH INTERNATIONAL ELECTRICAL ENGINEERING CONGRESS (IEECON), 2020,
  • [26] Near-Infrared Spectroscopic Method for Noninvasive Blood Glucose Prediction without Using Multivariate Analyses
    Maruo, Katsuhiko
    Yamada, Yukio
    DIABETES, 2018, 67
  • [27] Principle and research analysis of noninvasive blood glucose detection based on near infrared spectroscopy
    Liu, H. M.
    Lin, J.
    Liu, Y. J.
    BASIC & CLINICAL PHARMACOLOGY & TOXICOLOGY, 2018, 123 : 110 - 110
  • [28] A nonlinear model for calibration of blood glucose noninvasive measurement using near infrared spectroscopy
    Li, Qing-Bo
    Li, Li-Na
    Zhang, Guang-Jun
    INFRARED PHYSICS & TECHNOLOGY, 2010, 53 (05) : 410 - 417
  • [29] Noninvasive blood glucose monitoring by newly developed near-infrared spectroscopic system
    Maruo, K
    Tsurugi, M
    Tamura, M
    Ishii, M
    DIABETES, 2002, 51 : A124 - A124
  • [30] Near-Infrared Spectroscopy for Noninvasive Measurement of Blood Glucose: Problems, Progress, Tasks
    K. V. Pozhar
    M. O. Mikhailov
    E. L. Litinskaia
    E. A. Polyakova
    Biomedical Engineering, 2022, 56 : 64 - 68