The new EYE-D™ dosemeter for measurements of HP(3) for medical staff

被引:64
|
作者
Bilski, P. [1 ]
Bordy, J-M. [4 ]
Daures, J. [4 ]
Denoziere, M. [4 ]
Fantuzzi, E. [3 ]
Ferrari, P. [3 ]
Gualdrini, G. [3 ]
Kopec, M. [5 ]
Mariotti, F. [3 ]
Monteventi, F. [3 ]
Wach, S. [2 ]
机构
[1] Polish Acad Sci, Inst Nucl Phys, PL-31342 Krakow, Poland
[2] RADCARD, PL-30003 Krakow, Poland
[3] ENEA IRP Radiat Protect Inst, I-40136 Bologna, BO, Italy
[4] CEA, LIST, Lab Natl Henri Becquerel LNE LNHB, F-91191 Gif Sur Yvette, France
[5] AGH Univ Sci & Technol, Krakow, Poland
关键词
Interventional radiology; Dosimetry; Thermo luminescence; RADIATION; CATARACT; RISK;
D O I
10.1016/j.radmeas.2011.04.031
中图分类号
TL [原子能技术]; O571 [原子核物理学];
学科分类号
0827 ; 082701 ;
摘要
In recent years an increased interest in assessment of eye lens doses in interventional radiology has been observed, due to the fact that it seems that cataracts may be induced by radiation doses at levels lower than so far expected. One of the tasks of the EU FP7 ORAMED project was to develop the first dosemeter specially dedicated to measurements of Hp(3). This goal was achieved by designing and testing of the EYE-D (TM), dosemeter. This dosemeter, comprising an MCP-N (LiF:Mg,CU,P) thermoluminescent detector and an optimized polyamide capsule, was developed by the RADCARD company. The dosemeter is designed for an indefinite use and enables cold sterilization. The test measurements and Monte Carlo calculations of the photon energy response and angular response produced very satisfactory results: all obtained values are within about 20% around unity (with respect to Cs-137). The dosemeter fulfills all requirements for its application in dosimetry in interventional radiology. (C) 2011 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1239 / 1242
页数:4
相关论文
共 50 条
  • [1] The β-ray energy and angular response of the EYE-D™ eye-lens dosemeter
    Szumska, A.
    Budzanowski, M.
    Kopec, R.
    Olko, P.
    Ciupek, K.
    RADIATION MEASUREMENTS, 2013, 56 : 374 - 379
  • [2] Calibration of an eye lens dosemeter in terms of Hp(3) to be used in interventional radiology
    Borges, F. L. S.
    Guimaraes, M. C.
    Da Silva, T. A.
    Nogueira Tavares, M. S.
    RADIATION PHYSICS AND CHEMISTRY, 2014, 104 : 45 - 47
  • [3] INFLUENCE OF THE PHANTOM SHAPE (SLAB, CYLINDER OR ALDERSON) ON THE PERFORMANCE OF AN Hp(3) EYE DOSEMETER
    Behrens, R.
    Hupe, O.
    RADIATION PROTECTION DOSIMETRY, 2016, 168 (04) : 441 - 449
  • [4] Medical staff monitoring in interventional cardiology: over apron dosemeter placement based on measurements and simulations
    Ferrari, Paolo
    Venturi, Giovanna
    Campani, Lorenzo
    Mariotti, Francesca
    Becker, Frank
    Jansen, Jan
    Jovanovic, Zoran
    Krstic, Dragana
    Teles, Pedro
    RADIATION PROTECTION DOSIMETRY, 2024, 200 (08) : 802 - 807
  • [5] Type testing of a head band dosemeter for measuring eye lens dose in terms of Hp(3)
    Gilvin, P. J.
    Baker, S. T.
    Gibbens, N. J.
    Roberts, G. H.
    Tanner, R. J.
    Eakins, J. S.
    Hager, L. G.
    Daniels, T. J.
    RADIATION PROTECTION DOSIMETRY, 2013, 157 (03) : 430 - 436
  • [6] A NEW TL-DOS NEUTRON DOSEMETER FOR MEASUREMENTS OF THE PERSONAL DOSE EQUIVALENT HP(10)
    Heiny, M.
    Busch, F.
    Kroeninger, K.
    Theinert, R.
    Walbersloh, J.
    RADIATION PROTECTION DOSIMETRY, 2020, 188 (01) : 8 - 12
  • [7] Occupational Radiation Dose, Especially for Eye Lens: Hp(3), in Medical Staff Members Involved in Computed Tomography Examinations
    Osanai, Minoru
    Sato, Hidenori
    Sato, Kana
    Kudo, Kohsei
    Hosoda, Masahiro
    Hosokawa, Shota
    Kitajima, Maiko
    Tsushima, Megumi
    Fujita, Akira
    Hosokawa, Yoichiro
    Saito, Yoko
    APPLIED SCIENCES-BASEL, 2021, 11 (10):
  • [8] Measurement of Hp(10), Hp(3) and Hp(0.07) to medical staff in endoscopic retrograde cholangiopancreatography, using thermoluminescence dosimetry
    Zarifsanayei, Ava
    Sina, Sedigheh
    RADIATION PROTECTION DOSIMETRY, 2024, 200 (05) : 473 - 480
  • [9] 3D-printed eye lens dosemeter holder for use in interventional radiology and interventional cardiology
    Mihic, Marija Suric
    Pavelic, Luka
    Kortmis, Maja Vojnic
    Sisko, Jerko
    Maltar-Strmecki, Nadica
    Prlic, Ivica
    RADIATION MEASUREMENTS, 2020, 135
  • [10] Medical application of 3-D polymer gel dosemeter evaluated by nuclear magnetic resonance
    Novotny, J
    Dvorák, P
    Spevácek, V
    Tintera, J
    Novotny, J
    Cechák, T
    RADIATION PROTECTION DOSIMETRY, 2002, 101 (1-4) : 399 - 402