Image quality phantom and parameters for high spatial resolution small-animal SPECT

被引:8
|
作者
Visser, Eric P. [1 ]
Harteveld, Anita A. [2 ]
Meeuwis, Antoi P. W. [1 ]
Disselhorst, Jonathan A. [1 ]
Beekman, Freek J. [3 ,4 ]
Oyen, Wim J. G. [1 ]
Boerman, Otto C. [1 ]
机构
[1] Radboud Univ Nijmegen, Med Ctr, Dept Nucl Med, NL-6500 HB Nijmegen, Netherlands
[2] Univ Twente, Inst Tech Med, NL-7500 AE Enschede, Netherlands
[3] Delft Univ Technol, Sect Radiat Detect & Med Imaging, Delft, Netherlands
[4] Mol Imaging Labs, Utrecht, Netherlands
关键词
SPECT; Small animal; Image quality; NEMA; Phantom; PINHOLE SPECT; PET SCANNER; RECONSTRUCTION; COLLIMATION; TOMOGRAPHY; DEVICE; SYSTEM; CT;
D O I
10.1016/j.nima.2011.06.036
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
At present, generally accepted standards to characterize small-animal single photon emission tomographs (SPECT) do not exist. Whereas for small-animal positron emission tomography (PET), the NEMA NU 4-2008 guidelines are available, such standards are still lacking for small-animal SPECT. More specifically, a dedicated image quality (IQ) phantom and corresponding IQ parameters are absent. The structures of the existing PET IQ phantom are too large to fully characterize the sub-millimeter spatial resolution of modern multi-pinhole SPECT scanners, and its diameter will not fit into all scanners when operating in high spatial resolution mode. We therefore designed and constructed an adapted IQ phantom with smaller internal structures and external diameter, and a facility to guarantee complete filling of the smallest rods. The associated IQ parameters were adapted from NEMA NU 4. An additional parameter, effective whole-body sensitivity, was defined since this was considered relevant in view of the variable size of the field of view and the use of multiple bed positions as encountered in modern small-animal SPECT scanners. The usefulness of the phantom was demonstrated for (99m)Tc in a USPECT-II scanner operated in whole-body scanning mode using a multi-pinhole mouse collimator with 0.6 mm pinhole diameter. (C) 2011 Elsevier B.V. All rights reserved.
引用
下载
收藏
页码:539 / 545
页数:7
相关论文
共 50 条
  • [31] Evaluation of Quantitative Accuracy of Tc-99m Small-Animal SPECT Imaging in Phantom Studies
    Gerdekoohi, S. Khorasani
    Vosoughi, N.
    Tanha, K.
    Assadi, M.
    Ay, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2016, 43 : S475 - S475
  • [32] A multipinhole small animal SPECT system with submillimeter spatial resolution
    Funk, Tobias
    Despres, Philippe
    Barber, William C.
    Shah, Kanai S.
    Hasegawa, Bruce H.
    MEDICAL PHYSICS, 2006, 33 (05) : 1259 - 1268
  • [33] Performance Characterization of a High-Purity Germanium Detector for Small-Animal SPECT
    Johnson, Lindsay C.
    Campbell, Desmond L.
    Ovchinnikov, Oleg S.
    Peterson, Todd E.
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 607 - 612
  • [34] Small-Animal SPECT and SPECT/CT: Important Tools for Preclinical Investigation
    Franc, Benjamin L.
    Acton, Paul D.
    Mari, Carina
    Hasegawa, Bruce H.
    JOURNAL OF NUCLEAR MEDICINE, 2008, 49 (10) : 1651 - 1663
  • [35] High performance of an integrated PET/SPECT/CT system for small-animal imaging
    Magota, Keiichi
    Kubo, Naoki
    Kuge, Yuji
    Nishijima, Ken-ichi
    Zhao, Songji
    Tamaki, Nagara
    JOURNAL OF NUCLEAR MEDICINE, 2010, 51
  • [36] Compact CT/SPECT small-animal imaging system
    Kastis, GA
    Furenlid, LR
    Wilson, DW
    Peterson, TE
    Barber, HB
    Barrett, HH
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2004, 51 (01) : 63 - 67
  • [37] Spatial resolution of the Inveon small-animal PET scanner for the entire field of view
    Disselhorst, J. A.
    Boerman, O. C.
    Oyen, W. J. G.
    Slump, C. H.
    Visser, E. P.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 615 (02): : 245 - 248
  • [38] Detectors for small-animal SPECT I - Overview of technologies
    Barrett, HH
    Hunter, WCJ
    SMALL ANIMAL SPECT IMAGING, 2005, : 9 - 48
  • [39] A Skew-Slit Collimator for Small-Animal SPECT
    Zeng, Gengsheng L.
    JOURNAL OF NUCLEAR MEDICINE TECHNOLOGY, 2008, 36 (04) : 207 - 212
  • [40] Sampling considerations for high resolution small animal SPECT
    Damiani, C
    Di Domenico, G
    Moretti, E
    Sabba, N
    Zavattini, G
    Del Guerra, A
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1770 - 1774