An algorithm for noise correction of dual-energy computed tomography material density images

被引:7
|
作者
Maia, Rafael Simon [1 ]
Jacob, Christian [1 ]
Hara, Amy K. [2 ]
Silva, Alvin C. [2 ]
Pavlicek, William [2 ]
Ross, Mitchell J. [2 ]
机构
[1] Univ Calgary, Dept Comp Sci, Calgary, AB T2N 1N4, Canada
[2] Mayo Clin, Dept Radiol, Scottsdale, AZ 85259 USA
关键词
Material density; Dual-energy computed tomography; Noise reduction; Anisotropic diffusion; CT; REDUCTION; SIMULATION; DIFFUSION; NUMBERS;
D O I
10.1007/s11548-014-1006-z
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Dual-energy computed tomography (DECT) images can undergo a two-material decomposition process which results in two images containing material density information. Material density images obtained by that process result in images with increased pixel noise. Noise reduction in those images is desirable in order to improve image quality. A noise reduction algorithm for material density images was developed and tested. A three-level wavelet approach combined with the application of an anisotropic diffusion filter was used. During each level, the resulting noise maps are further processed, until the original resolution is reached and the final noise maps obtained. Our method works in image space and, therefore, can be applied to any type of material density images obtained from any DECT vendor. A quantitative evaluation of the noise-reduced images using the signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and 2D noise power spectrum was done to quantify the improvements. The noise reduction algorithm was applied to a set of images resulting in images with higher SNR and CNR than the raw density images obtained by the decomposition process. The average improvement in terms of SNR gain was about 49 % while CNR gain was about 52 %. The difference between the raw and filtered regions of interest mean values was far from reaching statistical significance (minimum , average ). We have demonstrated through a series of quantitative analyses that our novel noise reduction algorithm improves the image quality of DECT material density images.
引用
收藏
页码:87 / 100
页数:14
相关论文
共 50 条
  • [31] Dual-energy Computed Tomography Applications in Uroradiology
    Jong Park
    Hersh Chandarana
    Michael Macari
    Alec J. Megibow
    [J]. Current Urology Reports, 2012, 13 : 55 - 62
  • [32] Dual-Energy Computed Tomography in Body Imaging
    Toia, Giuseppe V.
    Kim, Sooah
    Dighe, Manjiri K.
    Mileto, Achille
    [J]. SEMINARS IN ROENTGENOLOGY, 2018, 53 (02) : 132 - 146
  • [33] Dual-Energy Computed Tomography Technology and Challenges
    Krauss, Bernhard
    [J]. RADIOLOGIC CLINICS OF NORTH AMERICA, 2018, 56 (04) : 497 - +
  • [34] Pediatric Applications of Dual-Energy Computed Tomography
    Pena-Trujillo, Valeria
    Gallo-Bernal, Sebastian
    Tung, Erik L.
    Gee, Michael S.
    [J]. RADIOLOGIC CLINICS OF NORTH AMERICA, 2023, 61 (06) : 1069 - 1083
  • [35] Dual-Energy Computed Tomography : Technological Considerations
    Chung, Ryan
    Dane, Bari
    Yeh, Benjamin M.
    Morgan, Desiree E.
    Sahani, Dushyant V.
    Kambadakone, Avinash
    [J]. RADIOLOGIC CLINICS OF NORTH AMERICA, 2023, 61 (06) : 945 - 961
  • [36] Dual-energy Computed Tomography Applications in Uroradiology
    Park, Jong
    Chandarana, Hersh
    Macari, Michael
    Megibow, Alec J.
    [J]. CURRENT UROLOGY REPORTS, 2012, 13 (01) : 55 - 62
  • [37] Dual-energy computed tomography: what is it useful for?
    Delgado Sanchez-Gracian, C.
    Martinez Rodriguez, C.
    Trinidad Lopez, C.
    [J]. RADIOLOGIA, 2013, 55 (04): : 346 - 352
  • [38] Dual-Energy Computed Tomography Applications in Rheumatology
    Park, Eun Hae
    O'Donnell, Thomas
    Fritz, Jan
    [J]. RADIOLOGIC CLINICS OF NORTH AMERICA, 2024, 62 (05) : 849 - 863
  • [39] Vascular Applications of Dual-Energy Computed Tomography
    Rajiah, Prabhakar S.
    Kambadakone, Avinash
    Ananthakrishnan, Lakshmi
    Sutphin, Patrick
    Kalva, Sanjeeva P.
    [J]. RADIOLOGIC CLINICS OF NORTH AMERICA, 2023, 61 (06) : 1011 - 1029
  • [40] Dual-Energy Computed Tomography Applications in Neurointervention
    Wolman, Dylan N.
    Patel, Bhavik P.
    Wintermark, Max
    Heit, Jeremy J.
    [J]. JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2018, 42 (06) : 831 - 839